JOURNEY 74016414Recalls & reported problems

Smith & Nephew, Inc. · JOURNEY II UK RESECTION ANTERIOR/POSTERIOR BLOCK RIGHT MEDIAL/LEFT LATERAL SZ 4 · Class 2

All models of JOURNEY (566)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556690055), received between Jun 18, 2020 and Jun 11, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Device Damaged by Another Device” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K190085); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Damaged by Another Device1
  • Flaked1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2024Peeled/Delaminated; Peeled/Delaminated

    IT WAS REPORTED THAT, DURING UKA SURGERY, A JRNY II UNI RESECT AP BLOCK RM/LL SZ 4 HAD CONSIDERABLE METAL ABRASION, IT OCCURRED DURING FEMORAL PREPARATION, ALL THE PIECES WERE ACCOUNTED AND REMOVED WITH SUCTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19512656 ↗
  • InjuryJun 18, 2020Flaked; Device Damaged by Another Device; Flaked; Device Damaged by Another Device

    IT WAS REPORTED THAT DURING USE WHEN THE INSTRUMENT WAS INSIDE THE PATIENT, PRODUCED METAL SHAVING WHEN CONTACTED BY THE DRILL IN THE FEMUR PREP PROCESS. THERE WAS A DELAY BETWEEN 0-30 MIN. THE SURGERY WAS FINISHED WITH THE SAME DEVICE.

    Patient problems: Device Embedded In Tissue or Plaque; Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10172122 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190085. One submission often covers many models, so check the product and lot details in each record.

  • Z-1393-2023 · Mar 27, 2023 · Open, Classified
    JOURNEY II UNI XLPE TIBIA INSERT MEDIAL SZ 7-8 8MM
  • Z-2132-2020 · Jan 28, 2020 · Terminated
    JOURNEY II UNI RESECTION PREP TRIAL, LEFT MEDIAL/ RIGHT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of moti
  • Z-2133-2020 · Jan 28, 2020 · Terminated
    JOURNEY II UNI RESECTION PREP TRIAL, RIGHT MEDIAL/ LEFT LATERAL - Product Usage: The JOURNEY II Uni Resection Prep Trial is a reusable surgical instrument used during range of moti

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →