REAL INTELLIGENCE recallZ-0708-2024 · Blue Belt Technologies, Inc · initiated Nov 13, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Blue Belt Technologies, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Possibility that the user is unable to resolve drill disconnection error messages.

FDA root cause category: Device Design

What the company is doing

The recalling firm sent an Urgent Medical Device Recall/Correction notice, dated 11/13/23, by mail to consignees. The notice informed consignees that they may continue to use affected devices as is provided that they ensure the backup Robotics Instrument Tray is available during surgery and to review the CORI Surgical System User Manual, which provides instructions for solving error messages. If consignees are unable to resolve the error messages, they should be prepared to convert to manual surgical procedure following the instructions in Appendix D: Recovery Procedure Guidelines of the CORI Surgical System User Manual. In accordance with routine servicing, consignees should return the CORI drill after 75 uses for servicing, during which the recalling firm will make the hardware update required to correct this recall event. Consignees are asked to contact their Account Representative to report pervasive drill disconnection error messages. Completed acknowledgement forms are to be returned to the recalling firm by email at [email protected]. Questions are to be directed by email to [email protected] or to your Account Representative.

Recall details

Recall number
Z-0708-2024
Status
Open, Classified
Initiated by the company
Nov 13, 2023
Posted by the FDA
Jan 11, 2024
Recalling firm
Blue Belt Technologies, Inc
Quantity
1859 units
Product code
OLO
FDA submissions
K231963
Distribution
Domestic: Nationwide Distribution.

Product description

CORI REAL INTELLIGENCE Robotic Drill, Product No. ROB10013 and ROB10013S

Code information (models, lots, serial numbers)

UDI-DI: 00885556757321 and 00885556780886; Serial No. SN00001 through SN001202, SN500001 through SN502583.

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates