Skip to content

REAL INTELLIGENCE ROB10019Recalls & reported problems

Smith & Nephew, Inc. · REAL INTELLIGENCE TIBIA TRACKER · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556757383), received between Dec 11, 2023 and Dec 29, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Degraded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded1
  • Material Deformation1
  • Positioning Problem1
  • Unintended Movement1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 29, 2025Unstable; Material Deformation; Unstable; Material Deformation

    IT WAS REPORTED THAT BEFORE A TKA, A REAL INTELLIGENCE TIBIA TRACKER WAS NOTICED TO BE MOVING A BIT AND LOOKING PRETTY ROUNDED OFF. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23905748 ↗
  • MalfunctionJun 6, 2025Degraded; Unintended Movement; Degraded; Unintended Movement

    IT WAS REPORTED THAT DURING A CORI-ASSISTED TKA, THEY NOTICED THE REAL INTELLIGENCE TIBIA TRACKER HAVE WEAR AROUND WHERE THE CLAMPS TIGHTEN ON THE TRACKER FRAME. THEY WERE UNABLE TO COMPLETE TIBIA BONE REMOVAL BECAUSE OF THE MOVEMENT. SURGERY WAS PERFORMED AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICES. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22159170 ↗
  • MalfunctionDec 11, 2023Positioning Problem; Positioning Problem

    IT WAS REPORTED THAT THE REAL INTELLIGENCE TIBIA TRACKER IS LOOSE AND NOT PROPERLY HELD BY THE TRACKER CLAMP. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18300639 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →

Sources and updates