REAL INTELLIGENCE ROB10035Recalls & reported problems
Smith & Nephew, Inc. · REAL INTELLIGENCE 5 MM CYLINDRICAL BUR · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00885556757505), received between Nov 13, 2020 and Apr 28, 2026 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Insufficient Device Problem Information2
- Difficult to Open or Remove Packaging Material1
- Failure to Calibrate1
- Mechanical Jam1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Injury2
- Unspecified Tissue Injury1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2026Mechanical Jam
IT WAS REPORTED THAT, DURING CORI ASSISTED TKA SURGERY, WHILE DRILL CONNECTING, THE REAL INTELLIGENCE 5 MM CYLINDRICAL BU COULD NOT BE REMOVED FROM THE REAL INTELLIGENCE ROBOTIC DRILL. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, WITH CHANGE IN SURGICAL TECHNIQUE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25016593 ↗MalfunctionNov 1, 2024Unintended Movement; Unintended Movement
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY, ONE (1) REAL INTELLIGENCE 5 MM CYLINDRICAL BU POPPED OUT RIGHT BEFORE BURRING AND WOULD NOT SEAT PROPERLY. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACKUP DEVICE. NO INJURY WAS REPORTED AS A RESULT OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20588586 ↗InjuryApr 25, 2023Failure to Calibrate; Failure to Calibrate
IT WAS REPORTED THAT, WHILE SETTING-UP FOR A CORI-ASSISTED TKA, UPON DRILL CALIBRATION, THE FOLLOWING MESSAGE KEPT APPEARING: "CALIBRATION VERIFICATION ERROR: FAILED TO VERIFY THE ROBOTIC DRILL CALIBRATION". A COMBINATION OF THREE DIFFERENT DRILLS, POINT PROBES, DRILL ATTACHMENTS AND LONG ATTACHMENTS WERE TRIED FROM ALL THREE CORI INSTRUMENT SETS, AS WELL AS DELETING AND ENTERING NEW CASE. IN ADDITION, THE CORI CONSOLE WAS SHUT DOWN AND TURNED BACK ON. HOWEVER, THE ERROR PERSISTED. SURGERY WAS PERFORMED, WITHOUT DELAY, BY MANUAL PROCEDURE INSTEAD. AFTER THE PROCEDURE WAS COMPLETED, CALIBRATION WAS SUCCESSFUL BY USING A DIFFERENT BURR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16808185 ↗MalfunctionOct 1, 2021Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material
IT WAS REPORTED THAT, DURING SET UP FOR A CORI ASSISTED SURGERY, THEY WERE OPENING THE REAL INTELLIGENCE 5 MM CYLINDRICAL BUR, WHEN THEY TOOK IT OUT OF THE STERILE PACK THE LID OF THE TUBE SLID OFF AND THE BURR FELL ON THE GROUND. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITHOUT SIGNIFICANT DELAY. THE PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12565275 ↗InjuryNov 13, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING CORI TKA PROCEDURE WHILE BURRING ALL UNDER EXPOSURE MODE ON THE TIBIA IN AN XR CASE, SURGEON COMPLAINED THAT THE BUR "SKIPPED INTO SOFT TISSUE. IT IS UNKNOWN HOW WAS THE PROCEDURE COMPLETED.
Patient problems: Injury; Unspecified Tissue Injury
MAUDE report 10835428 ↗InjuryNov 13, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING CORI TKA PROCEDURE WHILE BURRING ALL UNDER EXPOSURE MODE ON THE TIBIA IN AN XR CASE, SURGEON COMPLAINED THAT THE BUR "SKIPPED" INTO THE FINISHED FEMORAL CONDYLE, TAKING OUT A CHUNK OF BONE. THE SURGEON DID NOT NOTE IF THIS CAUSED INJURY, OR IF IT WAS OUTSIDE THE CHAMFER CUT. THE FEMUR IMPLANT WAS SUCCESSFULLY CEMENTED.
Patient problems: Injury; No Clinical Signs, Symptoms or Conditions
MAUDE report 10835420 ↗
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026