REAL INTELLIGENCE ROB10035Recalls & reported problems

Smith & Nephew, Inc. · REAL INTELLIGENCE 5 MM CYLINDRICAL BUR · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00885556757505), received between Nov 13, 2020 and Apr 28, 2026 — classified as 3 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Insufficient Device Problem Information2
  • Difficult to Open or Remove Packaging Material1
  • Failure to Calibrate1
  • Mechanical Jam1
  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Injury2
  • Unspecified Tissue Injury1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 28, 2026Mechanical Jam

    IT WAS REPORTED THAT, DURING CORI ASSISTED TKA SURGERY, WHILE DRILL CONNECTING, THE REAL INTELLIGENCE 5 MM CYLINDRICAL BU COULD NOT BE REMOVED FROM THE REAL INTELLIGENCE ROBOTIC DRILL. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, WITH CHANGE IN SURGICAL TECHNIQUE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25016593 ↗
  • MalfunctionNov 1, 2024Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY, ONE (1) REAL INTELLIGENCE 5 MM CYLINDRICAL BU POPPED OUT RIGHT BEFORE BURRING AND WOULD NOT SEAT PROPERLY. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACKUP DEVICE. NO INJURY WAS REPORTED AS A RESULT OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20588586 ↗
  • InjuryApr 25, 2023Failure to Calibrate; Failure to Calibrate

    IT WAS REPORTED THAT, WHILE SETTING-UP FOR A CORI-ASSISTED TKA, UPON DRILL CALIBRATION, THE FOLLOWING MESSAGE KEPT APPEARING: "CALIBRATION VERIFICATION ERROR: FAILED TO VERIFY THE ROBOTIC DRILL CALIBRATION". A COMBINATION OF THREE DIFFERENT DRILLS, POINT PROBES, DRILL ATTACHMENTS AND LONG ATTACHMENTS WERE TRIED FROM ALL THREE CORI INSTRUMENT SETS, AS WELL AS DELETING AND ENTERING NEW CASE. IN ADDITION, THE CORI CONSOLE WAS SHUT DOWN AND TURNED BACK ON. HOWEVER, THE ERROR PERSISTED. SURGERY WAS PERFORMED, WITHOUT DELAY, BY MANUAL PROCEDURE INSTEAD. AFTER THE PROCEDURE WAS COMPLETED, CALIBRATION WAS SUCCESSFUL BY USING A DIFFERENT BURR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16808185 ↗
  • MalfunctionOct 1, 2021Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT, DURING SET UP FOR A CORI ASSISTED SURGERY, THEY WERE OPENING THE REAL INTELLIGENCE 5 MM CYLINDRICAL BUR, WHEN THEY TOOK IT OUT OF THE STERILE PACK THE LID OF THE TUBE SLID OFF AND THE BURR FELL ON THE GROUND. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITHOUT SIGNIFICANT DELAY. THE PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12565275 ↗
  • InjuryNov 13, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING CORI TKA PROCEDURE WHILE BURRING ALL UNDER EXPOSURE MODE ON THE TIBIA IN AN XR CASE, SURGEON COMPLAINED THAT THE BUR "SKIPPED INTO SOFT TISSUE. IT IS UNKNOWN HOW WAS THE PROCEDURE COMPLETED.

    Patient problems: Injury; Unspecified Tissue Injury

    MAUDE report 10835428 ↗
  • InjuryNov 13, 2020Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING CORI TKA PROCEDURE WHILE BURRING ALL UNDER EXPOSURE MODE ON THE TIBIA IN AN XR CASE, SURGEON COMPLAINED THAT THE BUR "SKIPPED" INTO THE FINISHED FEMORAL CONDYLE, TAKING OUT A CHUNK OF BONE. THE SURGEON DID NOT NOTE IF THIS CAUSED INJURY, OR IF IT WAS OUTSIDE THE CHAMFER CUT. THE FEMUR IMPLANT WAS SUCCESSFULLY CEMENTED.

    Patient problems: Injury; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10835420 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →