REAL INTELLIGENCE ROB10036Recalls & reported problems
Smith & Nephew, Inc. · REAL INTELLIGENCE 6 MM CYLINDRICAL BUR · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00885556757512), received between Oct 24, 2022 and Apr 29, 2024 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Break2
- Material Deformation1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Laceration(s)1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2024Break; Break
IT WAS REPORT THAT. DURING A CORI ASSISTED SURGERY. THE REAL INTELLIGENCE 6 MM CYLINDRICAL BUR WAS NOTICED TO BE BROKEN INSIDE A REAL INTELLIGENCE ROBOTIC DRILL DURING THE PROCEDURE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED TO REMOVE THE BONE ADDITIONALLY BY SURGICAL INSTRUMENTS, WITHOUT ANY DELAY. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19210036 ↗MalfunctionJan 15, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING A CORI ASSISTED UKA SURGERY, THE SURGEON REMOVED THE PATIENT BONE USING THE REAL INTELLIGENCE 6 MM CYLINDRICAL BUR, BUT THE BUR WAR BROKE INSIDE THE REAL INTELLIGENCE ROBOTIC DRILL. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A BACK-UP DRILL. NO PATIENT INJURIES AND COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18511189 ↗MalfunctionDec 20, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A CORI ASSISTED UKR, WHILE REMOVING THE PATIENT'S BONE USING EXPOSURE MODE WITH A REAL INTELLIGENCE 6 MM CYLINDRICAL BUR, THE PATIENT'S BONE SUDDENLY WOULD NOT BE REMOVED; THE DRILL WAS CHECKED AND THE BUR WAS BROKEN INSIDE THE DRILL. NO PIECES OF THE DEVICE FELL INTO THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A BACKUP ROBOTIC DRILL. NO PATIENT INJURIES OR COMPLICATIONS WERE REPORTED DUE TO THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18370327 ↗MalfunctionDec 13, 2023Break; Break
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY, WHILE CUTTING THE BONE, THE BURR 6 MM WAS BROKEN. TO CONTINUE THE CASE THEY REBOOTED THE CONTROL UNIT AND USED A BACK-UP HANDPIECE, STARTED AGAIN TO COMPLETE THE CASE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18322732 ↗InjuryOct 26, 2023Material Deformation; Unintended Movement; Material Deformation; Unintended Movement
IT WAS REPORTED THAT, DURING A CORI-ASSISTED TKA, THE TARGET BONE WAS OVER-CUT AND BONE REMOVAL SCREEN SHOWED THE BONE SURFACE IN RED WHEN REMOVING THE BONE. THE PROCEDURE WAS COMPLETED BY FILLING WITH BONE CEMENT THE OVER-CUT. THERE WAS A TWENTY-MINUTE PROCEDURAL DELAY. THEY DID NOT FEEL SATISFIED WITH SURGERY¿S OUTCOME. IN ADDITION, AFTER THE PROCEDURE, THE TIP OF THE REAL INTELLIGENCE 6 MM CYLINDRICAL BUR WAS PROTRUDED ABOUT 1-2 MM FROM THE DRILL GUARD. IT WAS FOUND THAT THE CONNECTION PART OF THE BURR WHICH WAS USED IN PROCEDURE, WAS DEFORMED. SURGERY WAS PERFORMED, DOING A MEDICAL INTERVENTION. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Laceration(s)
MAUDE report 18009518 ↗MalfunctionJul 20, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A CORI-ASSISTED TKA, WHILE TIBIAL BURR, THE REAL INTELLIGENCE 6 MM CYLINDRICAL BUR STOPPED SPINNING AND GOT FREE FROM THE REAL INTELLIGENCE ROBOTIC DRILL. THE FEMUR CUT WAS FINE. THE AFFECTED DRILL DID NOT ACCEPT A NEW BURR, AND VISUALLY THE INITIAL BURR END TIP CONNECTION FRACTURED GOT STUCK INTO IT. SURGERY WAS PERFORMED AFTER A NON-SIGNIFICANT DELAY, WITH A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17368150 ↗MalfunctionMay 23, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A CORI-ASSISTED TKA, WHEN BURRING THE REAL INTELLIGENCE 6 MM CYLINDRICAL BUR FELL OUT. THE BUR BROKE AT END. A SMALL TIP WAS STUCK IN DRILL COLLET, RETRIEVE AND REMOVED FROM FIELD. SURGERY WAS PERFORMED AFTER A NON-SIGNIFICANT DELAY, WITH A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16986589 ↗MalfunctionOct 24, 2022Fracture; Fracture
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY, A REAL INTELLIGENCE 6 MM CYLINDRICAL BUR BROKE OFF AT THE CONNECTION. THE PIECE OF THE BROKEN BUR WAS STUCK INTO THE HANDPIECE. NO PIECE FELL INTO THE PATIENT. SURGERY WAS FINISHED AFTER A NON-SIGNIFICAND DELAY WITH A SN BACK-UP DEVICE. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15664165 ↗
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page