Real Intelligence ROB10086Recalls & reported problems
Smith & Nephew, Inc. · REAL INTELLIGENCE PLANE VISUALIZER · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556774878), received between Jun 10, 2022 and Oct 21, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Application Program Problem: Parameter Calculation Error” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Problem: Parameter Calculation Error1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2024Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY, IT WAS NOTICED THAT THE REAL INTELLIGENCE PLANE VISUALIZER IS OFF. IT SHOWS THAT THE CUT GUIDE IS AT EXTERNAL ROTATION OF 2.5 MM. REPLACED WITH A DIFFERENT PLANE CHECKER AND SHOWED THAT THE CUT GUIDE DID NOT HAVE ANY EXTERNAL ROTATION. THE SURGERY WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20493579 ↗MalfunctionJun 10, 2022Unintended Movement; Unintended Movement
IT WAS REPORTED THAT, DURING PREVENTATIVE MAINTENANCE, IT WAS NOTICED THAT THE REAL INTELLIGENCE PLANE VISUALIZER WON'T SLIDE ALL THE WAY ONTO THE POINT PROBE. IT WAS TESTED WITH DIFFERENT POINT PROBES TO CONFIRM AND IT WOULDN'T SLIDE ON EITHER. IT APPEARS THERE MIGHT BE A PIECE OF METAL THAT WASN¿T PROPERLY CUT WHEN YOU LOOK THROUGH THE OPENING OF THE PLANE VISUALIZER. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14659102 ↗
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.