Titan Modular Shoulder System 2.5 IMP-0926-030-00ARecalls & reported problems
Smith & Nephew, Inc. · COMBO INSERTER/IMPACTOR. XS · Class 1
All models of Titan Modular Shoulder System 2.5 (13)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885556860724), received between May 12, 2023 and Oct 30, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 30, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A TSA SURGERY, WHEN SURGEON WAS INSERTING THE FINLOCK IMPLANT USING A COMBO INSERTER/IMPACTOR. XS, ONE OF THE CLIPS ON THE INSTRUMENT BROKE OFF. SUCTION AND FORCEPS WERE USED TO REMOVE THE BROKEN PIECES. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18029453 ↗MalfunctionJun 14, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING TOTAL SHOULDER ARTHROPLASTY SURGERY, WHEN INSERTING THE FINLOCK IMPLANT INTO THE PATIENT, ONE OF THE ARMS OF THE COMBO INSERTER/IMPACTOR. XS BROKE OFF. THE BROKEN PART WAS REMOVED FROM THE PATIENT WITH FORCEPS AND SUCTION, AND THE SURGEON USED THE SECONDARY IMPACTOR TO ENSURE THE IMPLANT HAD GOOD CONTACT. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SAME DEVICE. SURGERY WAS NOT DELAYED. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17132239 ↗MalfunctionMay 20, 2023Fracture; Fracture
IT WAS REPORTED THAT WHILE INSERTING THE FINLOCK IMPLANT INTO THE PATIENT DURING A TSA, THE PLASTIC INSERTER OF THE COMBO INSERTER/IMPACTOR. XS BROKE. THE BROKEN PIECES WERE REMOVED WITH FORCEPS AND SUCTION FROM THE PATIENT. IT IS UNKNOWN HOW WAS THE SURGERY COMPLETED AND IF A DELAY WAS PRESENTED. NO INJURY TO PATIENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16970551 ↗MalfunctionMay 12, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TSA SURGERY, WHILE INSERTING THE FIN-LOCK IMPLANT INTO THE PATIENT, TWO COMBO INSERTER/IMPACTOR. XS BROKE OFF. THE BROKEN PIECES WERE REMOVED WITH FORCEPS AND SUCTION FROM THE PATIENT. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16924684 ↗
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,295 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,539 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,498 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,456 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall875 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR716 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR704 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR337 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page