Skip to content

ENGAGE 1-10003-210Recalls & reported problems

Smith & Nephew, Inc. · POROUS COATED FEMORAL COMP SIZE 2-LEFT MEDIAL · Class 2 · Implant

All models of ENGAGE (80)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556872956), received between Mar 13, 2023 and Mar 13, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Failure to Osseointegrate” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K190439, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 2, 2023Class IIOpen, Classified

    Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.

    Z-1670-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Osseointegrate1

Patient problems

  • Failure of Implant1
  • Inadequate Osseointegration1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 13, 2023Failure to Osseointegrate; Failure to Osseointegrate

    IT WAS REPORTED THAT, AFTER A TKA SURGERY HAD BEEN PERFORMED ON (B)(6) 2022, THE PATIENT EXPERIENCED CHRONIC PAIN. THIS ADVERSE EVENT WAS SOLVED BY A REVISION SURGERY ON (B)(6) 2023, IN WHICH ALL THE IMPLANTS WERE EXCHANGED WITH COMPETITOR DEVICES. DURING THE PROCEDURE THE SURGEON NOTICED THAT THE FEMORAL COMPONENT HAD VERY LITTLE BONE INGROWTH. THA PATIENT OUTCOME IS UNKNOWN.

    Patient problems: Failure of Implant; Inadequate Osseointegration; Pain

    MAUDE report 16527336 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190439, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1671-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6
  • Z-1672-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-1
  • Z-1673-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c)

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page