ENGAGE 1-10003-610Recalls & reported problems
Smith & Nephew, Inc. · POROUS COATED FEMORAL COMP SIZE 6-LEFT MEDIAL · Class 2 · Implant
All models of ENGAGE (80)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556873038), received between Apr 27, 2022 and May 26, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K190439, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 2, 2023Class IIOpen, Classified
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
Z-1670-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Failure to Osseointegrate1
- Migration1
Patient problems
- Pain3
- Ambulation or Postural Difficulties1
- Arthralgia1
- Inadequate Osseointegration1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 26, 2023Migration; Migration
IT WAS REPORTED THAT, AFTER A PARTIAL KNEE REPLACEMENT SURGERY PERFORMED ON (B)(6) 2021, THE PATIENT EXPERIENCED A TRAUMATIC FALL 4 MONTHS POST-OP. THE PATIENT REPORTEDLY KNOCKED HIMSELF UNCONSCIOUS AND EXPERIENCED INCREASED KNEE PAIN AFTER THE FALL. DUE TO THE TRAUMATIC FALL, THE FEMORAL COMPONENT, SIZE 6 - LEFT MEDIAL WAS COMPRESSED INTO THE BONE SURFACE AND WAS NO LONGER SITTING ON TOP OF THE BONE. THE FEMUR REPORTEDLY APPEARED TO BE INSET DUE TO FORCE APPLIED DURING THE FALL. PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2021 AND THE FEMORAL COMPONENT, SIZE 6 - LEFT MEDIAL, THE TIBIAL INSERT, SIZE 5, LEFT MEDIAL, 10MM AND THE TIBIAL ANCHOR STEM, SIZE 5-6 WERE SUCCESSFULLY REMOVED AND REPLACED. SURGEON DECIDED AFFINIUM 3D COATED TIBIAL TRAY, LEFT MEDIAL, SIZE 5 WAS NOT LOOSE AND ELECTED TO KEEP IT. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Pain
MAUDE report 17011274 ↗InjuryMay 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER PARTIAL RIGHT KNEE ARTHROPLASTY WAS PERFORMED ON (B)(6) 2020 (DOCUMENTED UNDER (B)(4) THE PATIENT EXPERIENCED UNEXPLAINED PAIN AROUND 5-MONTHS POST-OPERATION. X-RAY AND CT IMAGING SHOWED NO EVIDENCE OF PATIENT INJURY OR DEVICE MALFUNCTION. IMPLANTS WERE INTACT, WELL PLACED, AND THE SURROUNDING BONE HAD NO EVIDENCE OF DAMAGE. CT SHOWED GOOD BONE APPROXIMATION OF THE IMPLANTS. EVEN THOUGH, DUE TO THE PAIN, REVISION SURGERY WAS CONDUCTED ON (B)(6) 2021. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Pain
MAUDE report 17010966 ↗InjuryApr 27, 2022Failure to Osseointegrate; Failure to Osseointegrate
IT WAS REPORTED THAT, AFTER AN UNICOMPARTMENTAL KNEE ARTHROPLASTY HAD BEEN PERFORMED AROUND (B)(6) 2021, THE PATIENT EXPERIENCED PERSISTENT PAIN. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2022 TO TREAT THIS ADVERSE EVENT. DURING THIS PROCEDURE, THE TIBIAL AND FEMORAL COMPONENT WERE REVISED TO A NEW CEMENTED ENGAGE TIBIAL AND FEMORAL COMPONENT. INTRAOPERATIVELY, THE TIBIAL COMPONENT WAS NOTICED TO BE WELL FIXED TO THE BONE. THE FEMORAL COMPONENT WAS NOT WELL FIXED AND HAD ONLY FIBROUS TISSUE ON THE POROUS SURFACE OF THE IMPLANT. THE PATIENT'S OUTCOME IS UNKNOWN.
Patient problems: Ambulation or Postural Difficulties; Arthralgia; Inadequate Osseointegration; Pain
MAUDE report 14227390 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K190439, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
- Z-1672-2023 · May 2, 2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-1
- Z-1671-2023 · May 2, 2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6
- Z-1673-2023 · May 2, 2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c)
Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- LinkSymphoKnee 880-307/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- Scorpio 73-3710Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls1 report
- Scorpio 73-3110Howmedica Osteonics Corp. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls0 reports
- Arthrex® AR-503-TTTEArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall14 reports
- Arthrex® AR-503-TTTJArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall9 reports
- ENGAGE 1-10011-300Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- Arthrex® AR-503-TTTGArthrex, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10003-400Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page