Skip to content

ENGAGE 1-10011-100Recalls & reported problems

Smith & Nephew, Inc. · TIBIAL ANCHOR STEM SIZE 1-2 · Class 2 · Implant

All models of ENGAGE (80)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556873083), received between May 9, 2022 and May 26, 2023 — classified as 2 injuries.

The most frequently reported specific device problem is “Crack” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K190439, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 2, 2023Class IIOpen, Classified

    Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.

    Z-1671-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Crack1

Patient problems

  • Arthralgia1
  • Failure of Implant1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, SIX MONTHS AFTER UNICOMPARTMENTAL KNEE ARTHROPLASTY WAS PERFORMED ON AN UNKNOWN DATE, THE PATIENT EXPERIENCED PAIN. THIS ADVERSE EVENT WAS TREATED WITH A REVISION SURGERY TO TKA ON (B)(6) 2021. AN ABNORMALLY HIGH POSTERIOR SLOPE OF 19 DEGREES (THE STANDARD IS 5 DEGREES) WAS PRODUCED UPON TIBIAL RESECTION, WHICH IS THE PROBABLE ROOT CAUSE THE REPORTED PAIN. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Pain

    MAUDE report 17013176 ↗
  • InjuryMay 9, 2022Crack; Crack

    IT WAS REPORTED THAT, AFTER A CEMENTLESS UNICOMPARTMENTAL LEFT KNEE ARTHROPLASTY HAD BEEN PERFORMED ON (B)(6) 2021, THE PATIENT START EXPERIENCING PAIN AT THE KNEE. LATERAL X-RAYS PERFORMED TWO (2) MONTHS POSTOPERATIVELY APPEAR TO DEMONSTRATE A MAL-POSITIONED TIBIAL TRAY. LATERAL X-RAYS PERFORMED FIVE AND HALF (5.5) MONTHS POSTOPERATIVELY REVEALED A TIBIAL TRAY SUBSIDENCE AND THE PARTIAL FRACTURE OF THE ANCHOR COMPONENT. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2022 TO TREAT THIS ADVERSE EVENT. DURING THIS PROCEDURE, ALL THE UKA COMPONENTS WERE EXPLANTED AND REPLACED WITH A FULL TKA SYSTEM. THE PATIENT'S OUTCOME IS UNKNOWN.

    Patient problems: Arthralgia; Failure of Implant

    MAUDE report 14333089 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190439, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1672-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-1
  • Z-1670-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial);
  • Z-1673-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c)

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page