Skip to content

ENGAGE 1-10012-550Recalls & reported problems

Smith & Nephew, Inc. · ENGAGE POROUS TIBIAL TRAY SIZE 5-RIGHT MEDIAL · Class 2 · Implant

All models of ENGAGE (80)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00885556873205), received between May 8, 2022 and Aug 1, 2024 — classified as 5 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K190439, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 2, 2023Class IIOpen, Classified

    Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.

    Z-1672-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Off-Label Use1

Patient problems

  • Arthralgia2
  • Deformity/ Disfigurement1
  • Failure of Implant1
  • Inadequate Osseointegration1
  • Insufficient Information1
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 1, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT, AFTER A RIGHT UNICOMPARTMENTAL KNEE ARTHROPLASTY, THE PATIENT EXPERIENCED A LOOSE TIBIAL TRAY. THIS ADVERSE EVENT WAS SOLVED BY REVISION SURGERY ON (B)(6) 2023. CURRENT HEALTH STATUS OF PATIENT IN UNKNOWN.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19888816 ↗
  • InjuryJun 29, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A TKA SURGERY WAS PERFORMED ON (B)(6) 2022, THE PATIENT EXPERIENCED PAIN. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2023, IN WHICH THE ENGAGE POROUS TIBIAL TRAY SZ 5-RT MED WAS EXCHANGED. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Arthralgia

    MAUDE report 17225487 ↗
  • InjuryMay 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A PARTIAL KNEE REPLACEMENT SURGERY PERFORMED ON (B)(6) 2021, THE PATIENT REPORTED WEIGHT BEARING PAIN AT THE JOINT LINE 6 MONTHS POST-OP. SURGEON RECOMMENDED PATIENT TO WAIT AS PRIOR EXPERIENCE AND LITERATURE SUGGESTS THAT PAIN WOULD RESOLVE AS IMPLANTS SETTLED IN OVER TIME. PATIENT WAS REPORTEDLY NOT WILLING TO WAIT AND OPTED FOR A REVISION TO A TKA. SURGEON NOTED THAT HER LONG LIMB ALIGNMENT POST-SURGERY WAS ¿NOT THAT GREAT¿. PATIENT WAS SUCCESSFULLY CONVERTED TO A CEMENTED PRIMARY TKA ON (B)(6) 2021 WITH NO AUGMENTS AND MINIMAL BONE LOSS DUE TO THE REVISION PROCEDURE. ALSO, SURGEON NOTED THAT PATIENT MAY NOT HAVE BEEN A GOOD CANDIDATE FOR CEMENTLESS UKA AS SHE HAD A TIBIA VARA DEFORMITY IN HER LEFT TIBIA "THAT" MAY HAVE CONTRIBUTED TO HER PAIN.

    Patient problems: Pain

    MAUDE report 17014971 ↗
  • InjuryMay 9, 2022Off-Label Use; Off-Label Use

    IT WAS REPORTED THAT, AFTER A RIGHT UNICOMPARTMENTAL KNEE ARTHROPLASTY COMBINED WITH AN ACL RECONSTRUCTION HAD BEEN PERFORMED ON (B)(6) 2020, THE PATIENT EXPERIENCED THE RUPTURE OF THE ACL GRAFT. TO TREAT THIS ADVERSE EVENT, AN ADDITIONAL SURGERY AND CONVERSION TO A TKA WAS PERFORMED ON (B)(6) 2022 DUE TO GLOBAL KNEE INSTABILITY. DURING THIS PROCEDURE, ALL THE UKA DEVICES FROM THE CONSTRUCT WERE EXPLANTED AND REPORTED TO BE WELL-FIXED. THE PATIENT'S OUTCOME IS UNKNOWN.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 14342009 ↗
  • InjuryMay 8, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A RIGHT UNICOMPARTMENTAL KNEE ARTHROPLASTY HAD BEEN PERFORMED ON (B)(6) 2021, THE PATIENT EXPERIENCED WEIGHT BEARING PAIN AT THE JOINT LINE 5 MONTHS POSTOPERATIVELY. A SUBSTANTIAL AMOUNT OF RESIDUAL VARUS IN THE LONG LIMB ALIGNMENT WAS REVEALED BY POSTOPERATIVE ANTEROPOSTERIOR X-RAYS OF THE RIGHT TIBIA. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 TO TREAT THIS ADVERSE EVENT. DURING THIS PROCEDURE, ALL COMPONENTS WERE EXPLANTED AND REPLACED WITH A FULL TKA CONSTRUCT. INTRAOPERATIVELY, ALL IMPLANTS WERE WELL FIXED AND WITH A GOOD BONE INGROWTH IF POROUS COATED. THE PATIENT'S OUTCOME IS UNKNOWN.

    Patient problems: Arthralgia; Deformity/ Disfigurement

    MAUDE report 14332483 ↗

FDA clearance and classification

Cleared through 510(k) K230653 on May 4, 2023 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Smith & Nephew, Inc. ANTHEM Total Knee System, Genesis Uni Knee System, JOURNEY BCS and II Knee Systems, JOURNEY II CR Knee System, JOURNEY II Uni Knee System, Journey II UK Knee System, JOURNEY II XR Bi-Cruciate Retaining Knee System, JOURNEY PFJ System, JOURNEY Uni Knee System, Legion Hinge Total Knee System, Legion Total Knee System, ZUK Unicompartmental Knee System, ENGAGE Partial Knee System and Porous Patella and Porous Tibial Baseplates”. The FDA review took 56 days. FDA record ↗

Cleared through 510(k) K190439 on Nov 21, 2019 (decision: Substantially Equivalent). Submitted by Engage Uni, LLC as “Engage™ Partial Knee System”. The FDA review took 269 days. FDA record ↗

The FDA regulates product code HSX (prosthesis, knee, femorotibial, non-constrained, cemented, metal/polymer) under 21 CFR 888.3520 as a Class II device. 21 CFR 888.3520 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K190439, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1671-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6
  • Z-1670-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial);
  • Z-1673-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c)

Common questions about the ENGAGE 1-10012-550

Has the ENGAGE 1-10012-550 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1672-2023 (Class II), began on May 2, 2023: Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the ENGAGE 1-10012-550?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between May 8, 2022 and Aug 1, 2024. In those reports, the most common specific device problems are “Loosening of Implant Not Related to Bone-Ingrowth” (1) and “Off-Label Use” (1); the most common patient problems are “Arthralgia” (2), “Deformity/ Disfigurement” (1) and “Failure of Implant” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ENGAGE 1-10012-550?

None of the reports are classified as deaths, 5 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the ENGAGE 1-10012-550 FDA approved?

It was cleared through 510(k) K190439 on Nov 21, 2019; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the ENGAGE 1-10012-550 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the ENGAGE 1-10012-550, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand ENGAGE, model 1-10012-550). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates