ENGAGE 1-10012-600Recalls & reported problems
Smith & Nephew, Inc. · ENGAGE POROUS TIBIAL TRAY SIZE 6-LEFT MEDIAL · Class 2 · Implant
All models of ENGAGE (80)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556873212), received between May 8, 2022 and May 26, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Failure to Osseointegrate” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K190439, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 2, 2023Class IIOpen, Classified
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
Z-1672-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Failure to Osseointegrate1
Patient problems
- Pain2
- Arthralgia1
- Inadequate Osseointegration1
- Joint Dislocation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER UKA SURGERY WAS PERFORMED ON (B)(6) 2021, AP X-RAYS DEMONSTRATED TIBIAL COLLAPSE WHERE TIBIAL IMPLANT WAS ROTATED IN VALGUS. THIS ADVERSE EVENT WAS TREATED WITH REVISION SURGERY ON (B)(6) 2021 (5 WEEKS POST-OP) FROM A UKA TO TKA. DURING THIS PROCEDURE, IT WAS OBSERVED THAT THE BONE QUALITY WAS VERY POOR. IT WAS DETERMINED THAT A CEMENTED TOTAL KNEE ARTHROPLASTY SHOULD HAVE BEEN PERFORMED INSTEAD AS INITIAL SURGERY, SINCE THE PATIENT WAS NOT A GOOD CANDIDATE FOR A UKA. THE SURGEON WAS ABLE TO LIFT THE TIBIAL TRAY & ANCHOR OUT OF THE PATIENT WITH A RONGEUR BY HAND. REVISION PROCEDURE FROM UKA TO TKA WAS SUCCESSFULLY COMPLETED. BASED ON THE INFORMATION PROVIDED BY 3 INDEPENDENT INDIVIDUALS, INCLUDING 2 TRAINED ORTHOPEDIC MEDICAL DOCTORS, THE CAUSE OF THE INCIDENT WAS POOR BONE QUALITY OF THE PATIENT, AND THE DEVICE DID NOT CAUSE OR CONTRIBUTE TO THE INCIDENT.
Patient problems: Joint Dislocation; Pain
MAUDE report 17014958 ↗InjuryMay 8, 2022Failure to Osseointegrate; Failure to Osseointegrate
IT WAS REPORTED THAT, AFTER A LEFT UNICOMPARTMENTAL KNEE ARTHROPLASTY HAD BEEN PERFORMED ON (B)(6) 2021, THE PATIENT EXPERIENCED PAIN. DURING A FOLLOW UP CONDUCTED SIX (6) WEEKS POSTOPERATIVELY, X-RAYS APPEAR TO REVEAL THAT THE TIBIAL TRAY WAS NOT FULLY SEATED ON THE TIBIAL BONE. 8 MONTHS POSTOPERATIVELY, PATHOLOGIC RADIOLUCENT LINES WITH NON-UNIFORM THICKNESS UNDER THE TIBIAL TRAY WERE PRESENT AS REVEALED BY X-RAYS. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 TO TREAT THIS ADVERSE EVENT. DURING THIS PROCEDURE, ALL THE UKA DEVICES WERE EXPLANTED AND EXCHANGED WITH A FULL TKA SYSTEM. INTRAOPERATIVELY, NO EVIDENT GAPS WERE NOTICED AT THE IMPLANT INTERFACES, BUT LIMITED BONE IN-GROWTH WAS NOTED ON THE POROUS SURFACE OF THE IMPLANTS. THE PATIENT'S OUTCOME IS UNKNOWN.
Patient problems: Arthralgia; Inadequate Osseointegration; Pain
MAUDE report 14332493 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K190439, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
- Z-1670-2023 · May 2, 2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial);
- Z-1671-2023 · May 2, 2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6
- Z-1673-2023 · May 2, 2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c)
Other Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer devices
- ENGAGE 1-10003-410Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall10 reports
- ENGAGE 1-10012-500Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall6 reports
- ENGAGE 1-10003-600Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall4 reports
- ENGAGE 1-10003-500Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall3 reports
- ENGAGE 1-10011-200Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall3 reports
- ENGAGE 1-10012-350Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall3 reports
- ENGAGE 1-10003-310Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall2 reports
- ENGAGE 1-10012-200Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page