CORIOGRAPH ROB20030Recalls & reported problems
Smith & Nephew, Inc. · CORIOGRAPH PRE-OP PLANNING AND MODELING SERVICE · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556917442), received between Aug 22, 2024 and Dec 9, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Application Program Problem: Parameter Calculation Error” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Application Program Problem: Parameter Calculation Error2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 9, 2025Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error
IT WAS REPORTED THAT, DURING A REAL INTELLIGENCE CORI TKA SURGERY, THE CORIOGRAPH PRE-OP SERVICES PLAN INDICATED A SIZE 5 J2 BCS FEMUR, WITH PLANNED POSTERIOR RESECTIONS OF 9.0 MM MEDIALLY AND 7.5 MM LATERALLY, SET AT 0° REFERENCING THE PCA. HOWEVER, AFTER REGISTRATION, THE INITIAL POSTERIOR CUTS DISPLAYED WERE 12.5 MM MEDIALLY AND 7.5 MM LATERALLY, ALSO AT 0° REFERENCING THE PCA. DUE TO THIS DISCREPANCY, IT WAS VERIFY USING THE MANUAL SIZING GUIDE. THE GUIDE CONFIRMED A SIZE 5 FEMUR. THE PROCEDURE WAS RESUMED, AS NORMAL, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23756572 ↗MalfunctionAug 22, 2024Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error
IT WAS REPORTED THAT, DURING A CORI ASSISTED TKA SURGERY WITH CORIOGRAPH, THE FEMUR MODEL WAS UNABLE TO POPULATE WHILE COLLECTING DATA POINTS. THE MODEL WOULD JUMP BETWEEN LOOKING LIKE A LOT IS FILLED IN, BACK TO MISSING MULTIPLE PLACES WITH CRITICAL BLUE AREAS. MORE THEN ENOUGH DATA SHOULD HAVE BEEN PROVIDED FOR THE MODEL TO GENERATE. AFTER THE SURGEON DECIDED TO CONVERT TO AN IMAGE LESS PROCEDURE THE MODEL APPEARED TO GENERATE WITH ALL THE DATA COLLECTED, SHOWING A NORMAL FEMUR MODEL. THE SURGERY WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME REPORTED DEVICE. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20031656 ↗
FDA clearance and classification
Cleared through 510(k) K240113 on Mar 18, 2024 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “CORIOGRAPH Knee Pre-Op Plan”. The FDA review took 62 days. FDA record ↗
The FDA regulates product code PBF (orthopaedic surgical planning and instrument guides) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: As a surgical instrument to assist in pre-operative planning and/or guiding the marking of bone and/or guide surgical instruments for orthopedic surgical procedures.
Common questions about the CORIOGRAPH ROB20030
Has the CORIOGRAPH ROB20030 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00885556917442), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CORIOGRAPH ROB20030?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Aug 22, 2024 and Dec 9, 2025. In those reports, the most common specific device problem is “Application Program Problem: Parameter Calculation Error” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CORIOGRAPH ROB20030?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CORIOGRAPH ROB20030 FDA approved?
It was cleared through 510(k) K240113 on Mar 18, 2024. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the CORIOGRAPH ROB20030, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand CORIOGRAPH, model ROB20030). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Orthopaedic surgical planning and instrument guides devices
- Trabecular Metal 20-8075-004-01Orthosoft Inc · Orthopaedic Surgical Planning And Instrument Guides2 reports
- SurgiCase UEA-MA01-00-01Materialise NV · Orthopaedic Surgical Planning And Instrument Guides2 reports
- SurgiCase O-01-0-101000-0011-010Materialise NV · Orthopaedic Surgical Planning And Instrument Guides1 report
- SurgiCase O-01-0-105000-0011-010Materialise NV · Orthopaedic Surgical Planning And Instrument Guides1 report
- VSP Orthopedics Allograft Tibia Cutting Guide3d Systems, Inc. · Orthopaedic Surgical Planning And Instrument Guides1 report
- Materialise Personalized Solutions O-01-0-101000-0011-010Materialise NV · Orthopaedic Surgical Planning And Instrument Guides1 report
- VSP Orthopedics Allograft Orthopedic Model3d Systems, Inc. · Orthopaedic Surgical Planning And Instrument Guides1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page