N/A (REF 201101)Recalls & reported problems

Stryker Sustainability Solutions, Inc. · Steerable Diagnostic EP Catheter , LARGE 4.0, 10 electrodes, 2-5-2mm spacing, 6F · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00885825003845), received between Nov 6, 2018 and Mar 23, 2026 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

27 other recalls cover devices cleared under the same FDA submission (K030187); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 10, 2026Class IIOpen, Classified

    Incomplete seals on sterile product

    Z-2174-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanical Problem8
  • Use of Device Problem8
  • Material Deformation4
  • Positioning Problem2
  • Defective Device1
  • Fracture1
  • Inadequacy of Device Shape and/or Size1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • No Consequences Or Impact To Patient2
  • Insufficient Information1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2026Use of Device Problem

    IT WAS REPORTED CATHETER BENDING MECHANISM IS NOT FUNCTIONING PROPERLY AND DOES NOT OPERATE AS INTENDED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND THE EXTENDED PROCEDURE TIME REPORTED WAS A FEW MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24670655 ↗
  • MalfunctionSep 12, 2025Use of Device Problem; Material Deformation; Use of Device Problem; Material Deformation

    IT WAS REPORTED THAT THE DYNAMIC DECA CATHETER LOST ITS CURVE MID-CASE. AS REPORTED, THE DEVICE WAS REPLACED WITH OEM. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23042934 ↗
  • MalfunctionAug 11, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE PLUNGER WAS DIFFICULT TO MANEUVER. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22779348 ↗
  • MalfunctionAug 11, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE PLUNGER WAS DIFFICULT TO MANEUVER. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22779345 ↗
  • MalfunctionJun 25, 2025Use of Device Problem; Material Deformation; Use of Device Problem; Material Deformation

    IT WAS REPORTED THAT A DUODECA CATHETER WAS USED IN A CASE AND THE DOCTOR BELIEVED THERE WAS SOMETHING WRONG WITH THE CURVE OF THE CATHETER AND HAD ISSUES MANIPULATING IT. THE COMPLAINANT IS NOT AWARE OF ANY PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 22332942 ↗
  • MalfunctionJan 15, 2025Fracture; Mechanical Problem; Use of Device Problem; Fracture; Mechanical Problem; Use of Device Problem

    IT WAS REPORTED THAT CATHETER TIP WAS NOT DEFLECTING AND THAT IT WAS POSSIBLY BENT. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21162078 ↗
  • MalfunctionAug 21, 2023Use of Device Problem; Material Deformation; Use of Device Problem; Material Deformation

    IT WAS REPORTED THE DEVICE WAS BROKEN AS THE CURL HAD A GOOSE NECK SHAPE. HOWEVER, IT WAS ALSO REPORTED THAT THE DEVICE WAS NEVER PUT IN-SITU AND THERE WAS NO HARM TO PATIENT. THE COMPLAINANT IS NOT AWARE OF ANY MEDICAL INTERVENTION OR EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17592717 ↗
  • MalfunctionOct 4, 2021Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THE CATHETER WAS, "WORN OUT, DOES NOT STEER." THE DEVICE WAS REPLACED WITH ANOTHER. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12573654 ↗
  • MalfunctionJul 1, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE CATHETER DID NOT DEFLECT PROPERLY. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12104637 ↗
  • MalfunctionJun 21, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED PRIOR TO THE ABLATION PROCEDURE, THE DOCTOR TRIED TO DEFLECT THE CATHETER AND THE EP DID NOT RESPOND AS EXPECTED. THE CURVE HE GOT WAS NOT AS HE EXPECTED AS WELL. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. EXTENDED PROCEDURE TIME WAS REPORTED TO BE A FEW MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12037845 ↗
  • MalfunctionJun 21, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED PRIOR TO THE ABLATION PROCEDURE, THE DOCTOR TRIED TO DEFLECT THE CATHETER AND THE EP DID NOT RESPOND AS EXPECTED. THE CURVE HE GOT WAS NOT AS HE EXPECTED AS WELL. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. EXTENDED PROCEDURE TIME WAS REPORTED TO BE A FEW MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12037824 ↗
  • MalfunctionApr 19, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE DEVICE IS SOFT. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11690946 ↗
  • MalfunctionMar 25, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE CATHETER IS MIS-SHAPED. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11569584 ↗
  • MalfunctionJan 11, 2021Positioning Problem; Positioning Problem

    IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER WAS UNABLE TO BE KEPT IN POSITION. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11151736 ↗
  • MalfunctionNov 11, 2020Mechanical Problem; Inadequacy of Device Shape and/or Size; Mechanical Problem; Inadequacy of Device Shape and/or Size

    IT WAS REPORTED THE CATHETER DOES NOT MAKE PROPER CURVE WHEN TURNING. THERE WAS NO MEDICAL INTERVENTION. THE COMPLAINANT IS NOT AWARE OF ANY PATIENT INJURY, OR EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10823096 ↗
  • MalfunctionNov 2, 2020Positioning Problem; Positioning Problem

    IT WAS REPORTED THE CATHETER HAS POOR STEERING. THE COMPLAINANT IS NOT AWARE OF ANY PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10771537 ↗
  • MalfunctionMay 23, 2019Mechanical Problem; Material Deformation; Mechanical Problem; Material Deformation

    IT WAS REPORTED THAT DURING THE PROCEDURE THE ELECTROPHYSIOLOGY CATHETER DID NOT CURVE PROPERLY. INSTEAD OF A C CURVE IT LOOKED LIKE IT WAS BENT OUT OF SHAPE. THE ELECTROPHYSIOLOGY CATHETER WAS NOT DEFLECTING ENOUGH. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8637435 ↗
  • MalfunctionNov 6, 2018Defective Device; Defective Device

    IT WAS REPORTED THAT THE ELECTROPHYSIOLOGY CATHETER WAS NON-FUNCTIONAL. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8042422 ↗

FDA clearance and classification

Cleared through 510(k) K030187 on Sep 24, 2003 (decision: Substantially Equivalent). Submitted by Alliance Medical, Inc. as “Reprocessed Electrophysiology Catheter”. The FDA review took 246 days. FDA record ↗

The FDA regulates product code NLH (catheter, recording, electrode, reprocessed) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K030187. One submission often covers many models, so check the product and lot details in each record.

  • Z-0085-2027 · Aug 18, 2026 · Open, Classified
    BARD EP XT DECAPOLAR STEERABLE, DIAGNOSTIC EP CATHETER, 6F, LARGE 4.0, 10, 2-5-2MM, REF: 201007
  • Z-2179-2026 · Apr 10, 2026 · Open, Classified
    BARD Dynamic XT Steerable, Product Number 201103, REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2186-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401575; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2188-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401582; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2181-2026 · Apr 10, 2026 · Open, Classified
    BARD EP XT Decapolar Steerable, Product Number 201007; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2175-2026 · Apr 10, 2026 · Open, Classified
    BARD Dynamic Tip Steerable, Product Number 200131; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2198-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable BDB, Product Number 401918, REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2197-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401932; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2191-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401904; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2189-2026 · Apr 10, 2026 · Open, Classified
    Daig Livewire Steerable, Product Number 401603, REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2199-2026 · Apr 10, 2026 · Open, Classified
    Medtronic Marinr Steerable, Product Number 72402; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2182-2026 · Apr 10, 2026 · Open, Classified
    BARD EP XT Steerable, Product Number 200794; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2185-2026 · Apr 10, 2026 · Open, Classified
    CS Uni-Directional w Auto ID, Product Number D135304; REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2178-2026 · Apr 10, 2026 · Open, Classified
    BARD Dynamic XT Deca Steerable, Product Number 201102, REPROCESSED ELECTROPHYSIOLOGY CATHETER
  • Z-2176-2026 · Apr 10, 2026 · Open, Classified
    BARD Dynamic Tip Steerable, Product Number 200344; REPROCESSED ELECTROPHYSIOLOGY CATHETER

…and 12 more.

Common questions about the N/A

Has the N/A been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2174-2026 (Class II), began on Apr 10, 2026: Incomplete seals on sterile product. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the N/A?

The FDA's MAUDE database lists 18 adverse event reports for this device, received between Nov 6, 2018 and Mar 23, 2026. In those reports, the most common specific device problems are “Mechanical Problem” (8), “Use of Device Problem” (8) and “Material Deformation” (4). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the N/A?

None of the reports are classified as deaths, 0 as injuries and 18 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the N/A FDA approved?

It was cleared through 510(k) K030187 on Sep 24, 2003. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the N/A MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the N/A, and is it still sold?

The FDA's device database lists Stryker Sustainability Solutions, Inc. as the company responsible for it (brand N/A, model N/A). Its record shows it as in commercial distribution.