BARD Dynamic Deca Steerable, Product Number 201101 recallZ-2174-2026 · Stryker Sustainability Solutions · initiated Apr 10, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Sustainability Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Incomplete seals on sterile product
FDA root cause category: Process control
What the company is doing
Stryker issued a Medical Device Recall notice to its consignees on 4/10/2026 via USPS certified mail. The notice explained the product issue, potential risk, and requested the following: "Customer Actions Requested: Our records indicate that you may have distributed, received and/or used the above affected products. 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product. Segregate the affected products, quarantine and discontinue use. 3. Please complete the Business Reply Form and return the completed via email to [email protected], or use the QR Code on Attachment B. Response is required. a. If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility." "For any additional information or questions please reach out to [email protected] or Stryker Customer Service at 800-624-4422."
Recall details
- Recall number
- Z-2174-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 10, 2026
- Posted by the FDA
- May 13, 2026
- Recalling firm
- Stryker Sustainability Solutions
- Quantity
- 3063 units
- Product code
- NLH
- FDA submissions
- K012708, K030187, K052603, K100254
- Distribution
- US Nationwide and the countries of Israel and Canada.
Product description
BARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Code information (models, lots, serial numbers)
Show the full code information (24k characters)
UDI 00885825003845, Lot Numbers: 3021656, 4158368, 4305347, 4457640, 4560383, 4660045, 4815497, 4881594, 4987017, 5032942, 3107869, 4160047, 4305348, 4457904, 4562726, 4660052, 4815499, 4881600, 4987018, 5033611, 3171909, 4160048, 4305500, 4460044, 4565142, 4662242, 4816776, 4881613, 4987355, 5033612, 3197667, 4160049, 4305998, 4461505, 4565897, 4662243, 4816890, 4881615, 4989058, 5033626, 3231845, 4160051, 4307998, 4461555, 4565899, 4662546, 4816891, 4881618, 4989060, 5033627, 3241416, 4160053, 4308004, 4461558, 4565901, 4662547, 4816898, 4882585, 4990629, 5033628, 3248595, 4161235, 4309667, 4461560, 4565902, 4662549, 4817898, 4883073, 4990630, 5033673, 3254499, 4161239, 4311144, 4461580, 4565903, 4662550, 4818896, 4883093, 4990647, 5033679, 3259151, 4162340, 4311159, 4461628, 4565904, 4662551, 4818897, 4883094, 4990648, 5034212, 3272697, 4163622, 4311161, 4461639, 4565905, 4662552, 4818901, 4883096, 4990649, 5034213, 3304525, 4163719, 4312042, 4461640, 4565906, 4662560, 4820191, 4883336, 4990650, 5034214, 3350827, 4164530, 4312217, 4461893, 4565907, 4663275, 4820282, 4883337, 4990653, 5034215, 3370014, 4166100, 4312218, 4461894, 4565908, 4663469, 4820283, 4883338, 4990796, 5034796, 3371667, 4167414, 4312220, 4462110, 4565909, 4664191, 4820499, 4883339, 4990798, 5035334, 3377338, 4167415, 4312223, 4462871, 4565910, 4664193, 4820501, 4883807, 4990799, 5035489, 3403757, 4167923, 4314048, 4462872, 4565911, 4664196, 4820502, 4883808, 4990856, 5037613, 3404142, 4167925, 4314189, 4462873, 4566255, 4665171, 4820597, 4883809, 4990857, 5037831, 3406269, 4167928, 4314374, 4462874, 4566256, 4670446, 4822107, 4883810, 4990858, 5038073, 3439491, 4167929, 4317696, 4462875, 4566257, 4670448, 4822139, 4884751, 4990859, 5038074, 3468445, 4167937, 4318005, 4462876, 4566258, 4672131, 4822143, 4890818, 4990860, 5038075, 3556073, 4168753, 4319656, 4462878, 4566259, 4673453, 4822144, 4890819, 4992052, 5038076, 3591743, 4169893, 4319681, 4463037, 4566260, 4676972, 4822145, 4890820, 4992053, 5038550, 3591997, 4169894, 4319682, 4463107, 4566529, 4677300, 4822264, 4890822, 4992267, 5038553, 3596167, 4170928, 4320208, 4463108, 4566530, 4677302, 4824646, 4890823, 4992268, 5038627, 3626212, 4171482, 4320215, 4463109, 4566531, 4680082, 4825070, 4890824, 4992269, 5038630, 3630102, 4171492, 4322931, 4463110, 4569049, 4680083, 4825072, 4890825, 4992468, 5038632, 3652910, 4174826, 4322932, 4463111, 4569597, 4680575, 4825074, 4891811, 4992469, 5038633, 3671843, 4176099, 4323006, 4463112, 4570065, 4682470, 4825075, 4892192, 4992470, 5038634, 3682233, 4177260, 4323007, 4463113, 4570066, 4682613, 4825076, 4892194, 4992471, 5038635, 3709599, 4177560, 4323008, 4463115, 4570071, 4682615, 4825077, 4892196, 4992472, 5038779, 3737804, 4178120, 4323596, 4463116, 4570072, 4682616, 4825083, 4892197, 4992473, 5038780, 3746230, 4179655, 4323636, 4463117, 4570109, 4682806, 4825180, 4892198, 4992474, 5038870, 3749681, 4179903, 4323637, 4463118, 4570110, 4682807, 4825599, 4895074, 4992475, 5038930, 3762325, 4182227, 4324335, 4463119, 4570168, 4685502, 4825973, 4896259, 4992480, 5039102, 3764156, 4183258, 4324402, 4463121, 4570169, 4685550, 4826435, 4896338, 4992481, 5039431, 3766673, 4183831, 4324411, 4463122, 4570627, 4685553, 4827202, 4896339, 4992482, 5039635, 3789896, 4185345, 4324415, 4463123, 4570629, 4685554, 4827204, 4896340, 4993202, 5039638, 3792944, 4185753, 4324418, 4463124, 4570630, 4685555, 4827558, 4896366, 4993221, 5039981, 3798457, 4186272, 4324508, 4463125, 4570631, 4686035, 4827753, 4896367, 4993319, 5039982, 3799155, 4186734, 4325665, 4463126, 4571251, 4686558, 4828246, 4896368, 4993584, 5040067, 3800244, 4186758, 4325779, 4463127, 4571253, 4687144, 4828250, 4896369, 4993585, 5043129, 3801425, 4189779, 4326537, 4463518, 4571254, 4687551, 4828848, 4896370, 4993586, 5043338, 3807481, 4190389, 4326644, 4463519, 4571255, 4687553, 4828849, 4896371, 4993587, 5043340, 3808053, 4190392, 4327481, 4463775, 4571259, 4687650, 4828852, 4896372, 4994032, 5043341, 3813862, 4192118, 4327487, 4465825, 4571771, 4687659, 4828854, 4896373, 4994034, 5043343, 3814243, 4194586, 4331536, 4465827, 4571803, 4692401, 4829569, 4896374, 4994035, 5043558, 3815024, 4195259, 4332341, 4465829, 4571804, 4692406, 4829570, 4897426, 4994036, 5043674, 3815653, 4195820, 4332342, 4465830, 4571805, 4693182, 4829571, 4897427, 4994037, 5044227, 3817471, 4195881, 4332349, 4467660, 4571806, 4693183, 4830423, 4939140, 4994136, 5044240, 3826027, 4196488, 4332351, 4467663, 4571840, 4693426, 4830428, 4939141, 4994137, 5044241, 3826029, 4196913, 4332368, 4467665, 4571869, 4693427, 4830435, 4939626, 4994174, 5044264, 3826368, 4197172, 4334034, 4467667, 4571870, 4693428, 4831027, 4939805, 4994253, 5044265, 3829648, 4197180, 4334035, 4467668, 4573043, 4693429, 4831544, 4939806, 4994254, 5044266, 3830309, 4197381, 4334037, 4468544, 4573047, 4693430, 4831654, 4940758, 4994255, 5044275, 3836033, 4197471, 4334206, 4468629, 4573050, 4698208, 4831885, 4940853, 4994887, 5044276, 3836616, 4197476, 4334214, 4468630, 4573051, 4698209, 4831887, 4941081, 4996427, 5044277, 3837306, 4199913, 4334225, 4468728, 4573454, 4702614, 4832285, 4941082, 4997893, 5044279, 3838506, 4200917, 4334593, 4468733, 4573457, 4702615, 4833139, 4941084, 4997900, 5044280, 3839168, 4202053, 4335089, 4469020, 4573517, 4712565, 4833146, 4941085, 4998038, 5044316, 3839618, 4202448, 4335760, 4469028, 4573518, 4712678, 4833152, 4941086, 4998039, 5044318, 3841993, 4203481, 4336153, 4470134, 4573521, 4717757, 4833155, 4941088, 4998041, 5044319, 3843114, 4203772, 4336410, 4470319, 4573524, 4717759, 4833483, 4941176, 4998068, 5044396, 3843961, 4206085, 4336424, 4470320, 4574469, 4719867, 4833484, 4941177, 4998073, 5044397, 3843966, 4206086, 4337185, 4470321, 4574470, 4720184, 4833485, 4941181, 4998075, 5044403, 3845368, 4206780, 4337187, 4470322, 4574471, 4720688, 4834264, 4941188, 4998076, 5044404, 3845386, 4206782, 4337395, 4470323, 4574472, 4720976, 4834267, 4941191, 4998089, 5044405, 3845388, 4206857, 4337399, 4470324, 4574473, 4722022, 4834268, 4941192, 4998092, 5044447, 3848384, 4209665, 4338188, 4470325, 4574474, 4722023, 4834269, 4941199, 4998230, 5044618, 3850841, 4209675, 4340808, 4470326, 4574475, 4724268, 4834414, 4941200, 4998475, 5044619, 3862298, 4210196, 4343330, 4471183, 4574476, 4724354, 4834588, 4945962, 4998761, 5044621, 3870810, 4210737, 4343452, 4471186, 4574477, 4725481, 4834589, 4945963, 4998762, 5044622, 3871027, 4211337, 4345975, 4471189, 4574478, 4725483, 4834591, 4945964, 4998829, 5044656, 3892407, 4212632, 4345976, 4471190, 4574635, 4725495, 4835226, 4945965, 4998832, 5044985, 3893099, 4212714, 4346068, 4471191, 4574640, 4725793, 4837445, 4945966, 4998833, 5045172, 3903290, 4213107, 4346238, 4471192, 4574887, 4725798, 4837666, 4945967, 4999042, 5047323, 3908835, 4213868, 4347754, 4471194, 4575016, 4726625, 4837780, 4945968, 4999043, 5047674, 3908836, 4214314, 4348509, 4471195, 4575345, 4727196, 4837782, 4945969, 4999044, 5050562, 3908837, 4217560, 4348591, 4471196, 4575347, 4727805, 4837783, 4945971, 4999552, 5050685, 3909140, 4218429, 4348650, 4471197, 4575348, 4727892, 4837785, 4946262, 4999553, 5052307, 3909145, 4218786, 4350017, 4471201, 4575350, 4728813, 4838058, 4946263, 4999555, 5054865, 3910881, 4218964, 4351307, 4471202, 4575351, 4728867, 4838059, 4947194, 4999730, 5055253, 3918683, 4219059, 4351796, 4471204, 4576479, 4728869, 4838122, 4948058, 5001610, 5055439, 3920095, 4219068, 4351799, 4471208, 4576480, 4728909, 4839004, 4948890, 5001611, 5055440, 3923729, 4220699, 4352142, 4471278, 4576760, 4731097, 4839005, 4948891, 5001613, 5056172, 3923760, 4220702, 4352290, 4471279, 4576769, 4731113, 4839007, 4949647, 5001969, 5056301, 3931283, 4221246, 4352291, 4471281, 4576847, 4731124, 4840367, 4949648, 5001974, 5056302, 3936190, 4222274, 4352337, 4471282, 4576899, 4731159, 4840373, 4949649, 5001977, 5056622, 3940240, 4223649, 4352340, 4471283, 4577011, 4731216, 4841837, 4949650, 5001978, 5056623, 3941951, 4226101, 4352541, 4471284, 4577012, 4732046, 4842036, 4949652, 5001979, 5056624, 3942107, 4226177, 4352543, 4471285, 4577104, 4732671, 4842037, 4949653, 5001980, 5057124, 3943606, 4226184, 4352736, 4472334, 4577120, 4732800, 4842267, 4949654, 5002478, 5057125, 3943902, 4226245, 4352739, 4472612, 4577122, 4732906, 4844789, 4949655, 5002479, 5057126, 3945472, 4227605, 4355105, 4474146, 4577124, 4732908, 4844790, 4949656, 5002480, 5057516, 3946386, 4227618, 4357470, 4474194, 4577125, 4733236, 4845514, 4949657, 5002481, 5057517, 3949174, 4227812, 4357574, 4474995, 4577126, 4733237, 4845760, 4949659, 5002482, 5058223, 3951841, 4227815, 4357601, 4475053, 4577132, 4733242, 4848977, 4949801, 5002483, 5058224, 3952280, 4227849, 4360620, 4475109, 4577135, 4733244, 4850122, 4949802, 5002485, 5058225, 3956113, 4227880, 4361998, 4476388, 4577140, 4733246, 4850265, 4952926, 5002676, 5058226, 3956116, 4228025, 4362006, 4477901, 4577345, 4733247, 4850612, 4952927, 5002711, 5058227, 3956158, 4228181, 4362009, 4478200, 4577407, 4733696, 4850613, 4952929, 5002883, 5058228, 3956159, 4228262, 4362186, 4478824, 4577447, 4734064, 4851139, 4953046, 5002884, 5058954, 3956160, 4228493, 4362193, 4479128, 4579167, 4734893, 4852389, 4953068, 5002885, 5059048, 3956164, 4229507, 4362194, 4479535, 4580369, 4736198, 4852390, 4953071, 5002901, 5059055, 3958638, 4229514, 4362240, 4479727, 4583474, 4736201, 4852391, 4953869, 5002902, 5059056, 3958646, 4230244, 4362570, 4480982, 4583475, 4736202, 4852392, 4953870, 5002903, 5059059, 3958738, 4233949, 4362572, 4480983, 4583476, 4736203, 4852428, 4953871, 5002904, 5059110, 3960124, 4234099, 4362574, 4482232, 4583479, 4736205, 4853333, 4953872, 5002905, 5059119, 3961165, 4234703, 4363393, 4483258, 4583480, 4741662, 4853338, 4953873, 5002907, 5059121, 3961266, 4234785, 4363395, 4483334, 4583481, 4741663, 4853339, 4953874, 5002908, 5059125, 3962752, 4234849, 4363396, 4483444, 4583500, 4741664, 4853340, 4953875, 5002909, 5059220, 3965810, 4235829, 4363491, 4483920, 4583502, 4741911, 4853970, 4953876, 5002911, 5060535, 3966178, 4238705, 4364043, 4483923, 4583503, 4742982, 4855210, 4953877, 5002912, 5060536, 3968665, 4239473, 4364671, 4484931, 4583505, 4742992, 4855533, 4953884, 5002913, 5061888, 3970107, 4241041, 4365785, 4485044, 4583507, 4744337, 4855534, 4953885, 5002914, 5062736, 3970166, 4241146, 4366870, 4485045, 4583530, 4744367, 4855535, 4960643, 5002915, 5062737, 3970167, 4242079, 4367022, 4485047, 4584043, 4744370, 4855538, 4960644, 5003103, 5062738, 3972443, 4242833, 4367025, 4485048, 4584045, 4745000, 4855539, 4960645, 5003104, 5062739, 3973352, 4243664, 4367129, 4485050, 4584336, 4745735, 4855587, 4960646, 5003689, 5062740, 3973904, 4244027, 4367785, 4485052, 4584350, 4746596, 4856198, 4961419, 5003710, 5062741, 3973909, 4244029, 4367786, 4486190, 4584363, 4747021, 4856491, 4961632, 5003714, 5062742, 3976972, 4244157, 4367927, 4486204, 4584368, 4747279, 4856492, 4961633, 5003716, 5062743, 3979562, 4244602, 4369593, 4486211, 4584369, 4747885, 4856506, 4961634, 5005267, 5062744, 3982515, 4244606, 4369751, 4486212, 4584370, 4748350, 4856507, 4961635, 5005274, 5062866, 3983954, 4246335, 4369875, 4486213, 4584382, 4748533, 4856508, 4961636, 5005279, 5062867, 3984215, 4246692, 4370244, 4486214, 4584553, 4749621, 4856509, 4961637, 5005599, 5062868, 3984244, 4246694, 4373252, 4486215, 4584556, 4751571, 4856511, 4961638, 5005600, 5062869, 3984293, 4246697, 4373805, 4486216, 4584557, 4751572, 4856874, 4961639, 5005601, 5062870, 3985031, 4246698, 4373806, 4486581, 4584571, 4751574, 4856876, 4962152, 5005603, 5062871, 3985255, 4246699, 4374951, 4486608, 4587168, 4751578, 4856877, 4962157, 5005605, 5062872, 3986714, 4246700, 4376799, 4486609, 4587556, 4751579, 4856878, 4962416, 5005606, 5062873, 3987347, 4248039, 4376800, 4486610, 4587611, 4755454, 4857169, 4962580, 5005607, 5062874, 3988964, 4248040, 4376801, 4486837, 4587616, 4755976, 4857922, 4962583, 5005608, 5062905, 3994237, 4248268, 4376848, 4488935, 4587784, 4756430, 4858257, 4962586, 5005610, 5064154, 3997512, 4249745, 4376854, 4489113, 4587788, 4756735, 4858258, 4963018, 5009366, 5064180, 4000163, 4249884, 4376875, 4489116, 4587950, 4761514, 4859369, 4963019, 5009690, 5064181, 4000919, 4251382, 4376878, 4491068, 4587951, 4762568, 4859370, 4963096, 5009691, 5064182, 4002116, 4251400, 4376910, 4491069, 4588537, 4762569, 4859618, 4963098, 5010104, 5066528, 4002117, 4251524, 4376920, 4491070, 4589470, 4762570, 4860167, 4963099, 5010106, 5066529, 4006480, 4252419, 4377550, 4491236, 4589592, 4762571, 4860175, 4963100, 5011247, 5066530, 4014582, 4252422, 4377551, 4495760, 4589926, 4762572, 4860329, 4963102, 5011251, 5067230, 4015006, 4252680, 4377554, 4495761, 4590062, 4762573, 4861125, 4963103, 5011321, 5067388, 4016905, 4253916, 4377559, 4496101, 4591455, 4762574, 4861916, 4963104, 5011857, 5067389, 4018032, 4254519, 4377566, 4496646, 4592171, 4762576, 4862610, 4963410, 5011860, 5067391, 4018367, 4254521, 4377572, 4496648, 4592997, 4762577, 4862611, 4963411, 5011861, 5067392, 4018371, 4254527, 4377607, 4496772, 4593332, 4762578, 4862879, 4963412, 5011863, 5067573, 4018506, 4255525, 4377612, 4496773, 4593441, 4762579, 4862899, 4964382, 5011864, 5067575, 4018512, 4255526, 4377613, 4496774, 4595763, 4762580, 4862900, 4964383, 5012158, 5067576, 4019399, 4258278, 4377614, 4496775, 4596865, 4762581, 4864419, 4966249, 5012665, 5067577, 4019486, 4263035, 4377615, 4496776, 4597950, 4762582, 4864420, 4966283, 5012666, 5067578, 4019487, 4263072, 4377617, 4496777, 4598142, 4762583, 4864423, 4966530, 5012667, 5067697, 4019488, 4263077, 4377671, 4496779, 4598222, 4762584, 4866141, 4968941, 5012982, 5067699, 4019489, 4263080, 4377673, 4496836, 4599666, 4762585, 4866205, 4969022, 5013137, 5067700, 4021497, 4263084, 4378212, 4496837, 4599667, 4762586, 4866206, 4969039, 5016581, 5069477, 4022208, 4263086, 4378299, 4496840, 4599673, 4762587, 4867137, 4969040, 5016605, 5069480, 4022291, 4263460, 4378300, 4496841, 4599675, 4762588, 4867345, 4969130, 5016611, 5070819, 4022293, 4263463, 4378302, 4496844, 4600363, 4762589, 4867382, 4969134, 5016827, 5070830, 4023149, 4263468, 4380367, 4496909, 4601576, 4762590, 4867710, 4969137, 5016828, 5070845, 4026748, 4263766, 4380592, 4498428, 4601577, 4762591, 4868401, 4970389, 5016829, 5070996, 4026749, 4265175, 4380706, 4498429, 4601894, 4762592, 4869290, 4970391, 5016830, 5070997, 4026750, 4265178, 4383514, 4498430, 4601902, 4762593, 4869291, 4970392, 5016832, 5070998, 4026751, 4265179, 4383518, 4498431, 4603430, 4762595, 4869316, 4970425, 5016889, 5071000, 4027042, 4265464, 4384743, 4500605, 4603449, 4762596, 4869319, 4970523, 5017222, 5071001, 4027046, 4265471, 4384753, 4501232, 4603464, 4762597, 4869330, 4971143, 5017223, 5071003, 4030015, 4265796, 4385355, 4501236, 4604203, 4764693, 4869333, 4971210, 5017224, 5071004, 4033674, 4266427, 4385356, 4501240, 4609365, 4765197, 4869334, 4971929, 5017225, 5071010, 4034872, 4266429, 4385358, 4501243, 4609571, 4765204, 4869335, 4971930, 5017226, 5071392, 4034949, 4266431, 4385866, 4501253, 4609668, 4765215, 4869344, 4971931, 5017227, 5071435, 4035099, 4266436, 4386907, 4502524, 4611329, 4766445, 4869354, 4971932, 5017326, 5071493, 4036526, 4266621, 4386908, 4502525, 4612216, 4766449, 4869363, 4971933, 5017342, 5071494, 4041446, 4266622, 4387303, 4503789, 4612234, 4766450, 4869378, 4971934, 5017343, 5071677, 4041659, 4266628, 4387305, 4504341, 4612480, 4766451, 4869379, 4971935, 5017549, 5071678, 4041965, 4267534, 4387306, 4504794, 4612487, 4766454, 4869385, 4971936, 5017589, 5071679, 4043201, 4267536, 4387460, 4505290, 4612751, 4766456, 4869386, 4971940, 5017590, 5071680, 4044025, 4267537, 4389173, 4507863, 4612770, 4766457, 4869388, 4971941, 5017591, 5072189, 4044331, 4267539, 4389596, 4509205, 4612771, 4766458, 4869389, 4971942, 5017593, 5072190, 4045774, 4267541, 4389625, 4510535, 4613337, 4768963, 4869392, 4971943, 5017594, 5072227, 4046886, 4267542, 4389741, 4515714, 4613926, 4768964, 4869402, 4972069, 5017596, 5072228, 4047898, 4267544, 4390050, 4518897.4614279, 4768965, 4869403, 4972073, 5018644, 5072229, 4049071, 4267742, 4391287, 4519350, 4614412, 4768966, 4869406, 4972075, 5019470, 5072230, 4055039, 4267768, 4391288, 4521754, 4615324, 4769573, 4869410, 4972076, 5020472, 5072249, 4058659, 4268205, 4391291, 4522756, 4616725, 4774326, 4869412, 4972077, 5020894, 5072251, 4059926, 4268207, 4391293, 4522813, 4616726, 4774950, 4869413, 4972456, 5021877, 5073507, 4059934, 4271745, 4391295, 4523028, 4616732, 4774951, 4869416, 4972457, 5021980, 5073508, 4061921, 4272128, 4391296, 4523343, 4617007, 4775225, 4869423, 4972912, 5022497, 5076613, 4062655, 4272969, 4391297, 4529564, 4618021, 4775919, 4870259, 4972915, 5022499, 5076615, 4066543, 4272972, 4391298, 4529664, 4618022, 4776776, 4871287, 4972916, 5022971, 5076616, 4066544, 4273081, 4391299, 4529667, 4618028, 4777201, 4871292, 4972918, 5022972, 5076678, 4066664, 4273094, 4392411, 4530387, 4618031, 4777208, 4871294, 4972919, 5022976, 5076681, 4066665, 4273506, 4392455, 4531012, 4618032, 4777265, 4871619, 4972921, 5023249, 5076928, 4066875, 4274072, 4392759, 4536235, 4618034, 4777557, 4871724, 4972924, 5023251, 5077725, 4066895, 4274175, 4392961, 4537815, 4618038, 4777872, 4871745, 4972936, 5023252, 5080178, 4066897, 4275121, 4392962, 4538569, 4618043, 4780122, 4873198, 4972937, 5023253, 5080179, 4066898, 4275409, 4393026, 4538629, 4618063, 4781657, 4873200, 4972938, 5023257, 5080180, 4066899, 4275411, 4393547, 4538630, 4619052, 4781680, 4873201, 4972939, 5023258, 5080181, 4066900, 4275560, 4393830, 4538635, 4619094, 4784899, 4873202, 4973654, 5023259, 5080182, 4066901, 4275628, 4395013, 4545279, 4619753, 4785760, 4873203, 4973689, 5023260, 5080268, 4066902, 4275637, 4395474, 4545291, 4619813, 4785902, 4873204, 4973690, 5023261, 5080269, 4066903, 4275639, 4396681, 4546478, 4619883, 4786811, 4873205, 4973694, 5023262, 5080270, 4069700, 4275640, 4396683, 4548505, 4619884, 4787289, 4873208, 4973800, 5023268, 5080317, 4071120, 4275646, 4396768, 4548900, 4621767, 4789907, 4873210, 4973803, 5023269, 5080318, 4071122, 4275977, 4396773, 4548924, 4621768, 4789908, 4873238, 4973843, 5023270, 5080319, 4071206, 4276090, 4396778, 4548925, 4624635, 4789912, 4873241, 4973905, 5023271, 5080320, 4071224, 4276091, 4397180, 4548928, 4624694, 4789913, 4873244, 4974063, 5023272, 5080321, 4071238, 4276096, 4397576, 4548930, 4624697, 4789916, 4873245, 4974090, 5023274, 5080793, 4071239, 4276629, 4397985, 4548932, 4624704, 4789917, 4873246, 4974102, 5023275, 5080867, 4074807, 4276705, 4399560, 4548934, 4625039, 4792633, 4873248, 4974405, 5023277, 5080868, 4074808, 4278355, 4400379, 4548936, 4625040, 4792642, 4873249, 4974423, 5023347, 5080869, 4077436, 4278356, 4401429, 4548937, 4625086, 4792781, 4873251, 4974514, 5023917, 5080870, 4077448, 4278485, 4401561, 4548958, 4625088, 4792996, 4873252, 4974528, 5024343, 5080871, 4077449, 4278489, 4401750, 4549015, 4628269, 4793053, 4873255, 4974741, 5025462, 5080931, 4085793, 4278491, 4402535, 4549023, 4628509, 4793227, 4873258, 4975447, 5025470, 5080951, 4085796, 4278494, 4403414, 4549031, 4628531, 4793669, 4873259, 4975508, 5025701, 5081390, 4085855, 4278537, 4405288, 4549032, 4628553, 4793670, 4873261, 4975509, 5025702, 5082234, 4085908, 4278538, 4405292, 4549035, 4629102, 4793681, 4873262, 4975526, 5026065, 5082235, 4085910, 4278684, 4405448, 4549036, 4629103, 4794120, 4873263, 4976224, 5026066, 5082236, 4088864, 4278687, 4407183, 4549040, 4630515, 4795263, 4873264, 4976573, 5029154, 5082288, 4088868, 4279463, 4408885, 4549045, 4632167, 4796282, 4873266, 4976576, 5029155, 5082289, 4090869, 4279660, 4408887, 4549340, 4632592, 4798039, 4873268, 4976768, 5029156, 5082290, 4092529, 4281214, 4409092, 4549341, 4635227, 4798040, 4873275, 4978077, 5029157, 5082603, 4095331, 4281414, 4409238, 4549391, 4636021, 4798041, 4873380, 4979490, 5029158, 5082604, 4097283, 4281418, 4409779, 4549393, 4636988, 4799880, 4873381, 4979491, 5029159, 5082605, 4097285, 4281484, 4409787, 4549394, 4637061, 4799881, 4873383, 4979524, 5029160, 5082606, 4097420, 4281486, 4414755, 4549395, 4639897, 4800785, 4873384, 4980342, 5029161, 5082607, 4097432, 4281487, 4416110, 4549396, 4640280, 4800786, 4873385, 4980395, 5029332, 5082608, 4097925, 4281686, 4416112, 4549401, 4640281, 4801194, 4873389, 4980396, 5029344, 5082609, 4097954, 4281687, 4416201, 4549402, 4640286, 4801431, 4873390, 4980488, 5029345, 5082700, 4101672, 4281690, 4416325, 4549410, 4642583, 4801433, 4873394, 4980547, 5029346, 5082736, 4104592, 4281691, 4417573, 4549567, 4643000, 4801545, 4873395, 4980548, 5029347, 5082737, 4107161, 4282195, 4419066, 4549572, 4643004, 4801546, 4874870, 4980549, 5029348, 5082738, 4114293, 4282507, 4420950, 4549580, 4643009, 4801547, 4874875, 4980550, 5029350, 5082829, 4117262, 4282508, 4421186, 4549638, 4643215, 4802583, 4874876, 4980551, 5029385, 5082830, 4118063, 4282509, 4422970, 4549639, 4643234, 4803716, 4875131, 4980552, 5029386, 5082831, 4119112, 4283949, 4423404, 4549949, 4643772, 4803941, 4877580, 4980553, 5029594, 5083010, 4119145, 4284418, 4429526, 4549968, 4643794, 4803942, 4877630, 4980667, 5029596, 5084997, 4120326, 4284419, 4430848, 4550019, 4644117, 4803957, 4877715, 4980924, 5029696, 5084998, 4120919, 4284563, 4431139, 4550020, 4644118, 4803962, 4881060, 4980927, 5029697, 5085000, 4122413, 4284661, 4431886, 4552311, 4644126, 4804063, 4881061, 4981085, 5030129, 5085002, 4129794, 4285698, 4432706, 4552656, 4644683, 4804066, 4881062, 4981114, 5030133, 5085196, 4129851, 4287101, 4435856, 4556997, 4645718, 4804067, 4881063, 4981115, 5030323, 5085364, 4130695, 4287203, 4437858, 4559158, 4645724, 4806913, 4881066, 4981357, 5030324, 5085365, 4134313, 4287575, 4438456, 4559160, 4646683, 4808345, 4881067, 4981358, 5030394, 5085412, 4134411, 4290332, 4441303, 4559161, 4646686, 4808346, 4881068, 4981392, 5030971, 5085917, 4137980, 4290334, 4441306, 4559162, 4648006, 4808347, 4881099, 4981393, 5030974, 5085918, 4138643, 4293260, 4446555, 4559163, 4649196, 4808704, 4881100, 4981394, 5030975, 5085919, 4139564, 4294899, 4446698, 4559164, 4649400, 4808705, 4881101, 4981395, 5032083, 5085920, 4140948, 4294971, 4446891, 4559165, 4649402, 4808728, 4881102, 4981396, 5032084, 5085921, 4143012, 4297980, 4448121, 4559166, 4650553, 4808729, 4881103, 4981397, 5032085, 5085999, 4144717, 4298560, 4450634, 4559168, 4651678, 4808859, 4881104, 4981398, 5032086, 5086005, 4145567, 4299631, 4451314, 4559171, 4652041, 4810414, 4881105, 4981399, 5032087, 5086006, 4147788, 4300369, 4452132, 4559566, 4652047, 4812023, 4881106, 4983497, 5032088, 5086509, 4147883, 4300370, 4454475, 4559567, 4652050, 4813664, 4881107, 4983776, 5032103, 5086510, 4147887, 4300701, 4455033, 4559568, 4655092, 4813667, 4881108, 4984641, 5032104, 5086511, 4148802, 4300726, 4455039, 4559835, 4655625, 4813669, 4881109, 4984797, 5032163, 5086512, 4149227, 4302668, 4455040, 4559836, 4656012, 4813671, 4881254, 4986193, 5032164, 5086711, 4149969, 4303896, 4455168, 4559837, 4656013, 4813770, 4881440, 4986194, 5032187, 5086717, 4150815, 4303903, 4455169, 4559890, 4656020, 4814782, 4881442, 4986195, 5032188, 5086720, 4150816, 4304007, 4456588, 4559892, 4656023, 4814791, 4881453, 4986546, 5032369, 4152114, 4304318, 4456589, 4559941, 4656112, 4814850, 4881454, 4986547, 5032370, 4153497, 4304319, 4456591, 4559942, 4656116, 4815412, 4881455, 4986548, 5032387, 4153504, 4304355, 4457636, 4559943, 4656119, 4815413, 4881457, 4986565, 5032388, 4154957, 4304924, 4457638, 4559944, 4658880, 4815457, 4881458, 4986566, 5032389, 4155772, 4305035, 4457639, 4559946, 4658881, 4815460, 4881593, 4987016, 5032934.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall18 reports
Other devices under the same FDA submission
39 devices cleared under K012708, K030187, K052603, K100254 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED2 recalls0 reports
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall21 reports
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall8 reports
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall3 reports
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall3 reports
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall1 report
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall1 report
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall1 report
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall1 report
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall1 report
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall0 reports
- N/AStryker Sustainability Solutions, Inc. · CATHETER, RECORDING, ELECTRODE, REPROCESSED1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98708).
- Z-2175-2026 · Open, ClassifiedBARD Dynamic Tip Steerable, Product Number 200131; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2176-2026 · Open, ClassifiedBARD Dynamic Tip Steerable, Product Number 200344; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2177-2026 · Open, ClassifiedBARD Dynamic Tip Steerable, Product Number 6DYNTP001;
- Z-2178-2026 · Open, ClassifiedBARD Dynamic XT Deca Steerable, Product Number 201102, REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2179-2026 · Open, ClassifiedBARD Dynamic XT Steerable, Product Number 201103, REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2180-2026 · Open, ClassifiedBARD Dynamic XT Steerable, Product Number 201104; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2181-2026 · Open, ClassifiedBARD EP XT Decapolar Steerable, Product Number 201007; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2182-2026 · Open, ClassifiedBARD EP XT Steerable, Product Number 200794; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2183-2026 · Open, ClassifiedBW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2184-2026 · Open, ClassifiedCS Uni-Directional w Auto ID, Product Number D135303; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2185-2026 · Open, ClassifiedCS Uni-Directional w Auto ID, Product Number D135304; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2186-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401575; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2187-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401581; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2188-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401582; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2189-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401603, REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2190-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401652; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2191-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401904; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2192-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401905; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2193-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401908; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2194-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401914; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2195-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401915; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2196-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401923; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2197-2026 · Open, ClassifiedDaig Livewire Steerable, Product Number 401932; REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2198-2026 · Open, ClassifiedDaig Livewire Steerable BDB, Product Number 401918, REPROCESSED ELECTROPHYSIOLOGY CATHETER
- Z-2199-2026 · Open, ClassifiedMedtronic Marinr Steerable, Product Number 72402; REPROCESSED ELECTROPHYSIOLOGY CATHETER
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page