Skip to content

N/A (REF 1859)Recalls & reported problems

Stryker Sustainability Solutions, Inc. · Pulse Oximeter, Adult Sensor · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885825015831), received between Feb 21, 2018 and Feb 21, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Incorrect Measurement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Incorrect Measurement1

Patient problems

  • No Code Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 21, 2018Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE PULSE OXIMETER SENSOR GAVE INACCURATE READINGS. THE FACILITY REPORTED THAT ON A RELATIVELY SICK PATIENT, A PROCEDURE WAS ABORTED TO RESUSCITATE THE PATIENT FROM HYPOXIA WITH A TREMENDOUS A-A GRADIENT, INCLUDING THE USE OF ADVANCED TESTING LIKE BEDSIDE ECHO, INTRACARDIAC ECHO AND THE FACILITY WAS CONSIDERING ECHMO. IT WAS DETERMINED THE PATIENT WASN'T REALLY HYPOXIC. AFTER GETTING ABG RESULTS THAT CONFLICTED WITH THE DATA SEEN IN THE LAB, THE FACILITY SWITCHED THE PROBE, AND THE PATIENT¿S SATURATION WENT FROM 68% TO 100% IMMEDIATELY. THE PATIENT ENDED UP GOING INTUBATED TO THE ICU AFTER AN ABORTED PROCEDURE. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN CLARITY AND ADDITIONAL INFORMATION FROM THE FACILITY. NO INFORMATION WAS PROVIDED.

    Patient problems: No Code Available

    MAUDE report 7287829 ↗

Other OXIMETER, REPROCESSED devices

All OXIMETER, REPROCESSED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.