Skip to content

Nellcor MAX-ARecalls & reported problems

Hygia Health Services, Inc. · Nellcor OxiMax Adult Pulse Oximeter Sensor · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10817357020011), received between Nov 30, 2023 and Dec 15, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem1
  • Failure to Analyze Signal1
  • Incorrect Measurement1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Ulcer1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 15, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, DURING USE, THE PATIENT HAD PRESSURE ULCER WHILE USING THE SENSOR. THE PATIENT'S HAND WERE SWOLLEN AFTER USING THE DEVICE.

    Patient problems: Ulcer

    MAUDE report 18339204 ↗
  • MalfunctionNov 30, 2023Incorrect Measurement; Failure to Analyze Signal; Communication or Transmission Problem; Incorrect Measurement; Failure to Analyze Signal; Communication or Transmission Problem

    ACCORDING TO THE REPORTER, THE SENSOR HAD NO SPO2 READINGS OR SIGNAL ACQUISITION. WHEN USED ON A PATIENT, A READING OF 100% WAS COMING AND GOING AND IT HAD A NOTICEABLE INTERFERENCE IN TRACE. A WRAP WAS USED TO SECURE IT ON PATIENT'S FINGER AND SITE WAS CHECKED EVERY 4 HOURS. THE SAME SENSOR WAS PLUGGED IN A DIFFERENT MONITOR AND GOT SOLID READING OF 90%, HOWEVER WAVEFORM WAS NOT VERY CLEAN. ANOTHER SENSOR WAS PLUGGED INTO DIFFERENT MODULE AND HAD SAME RESULT AS BEFORE OF NOISY TRACE AND INTERMITTENT INCORRECT READING. THE MODULE WAS REMOVED FROM THE MAIN MONITOR AND PLUGGED INTO A PORTABLE MONITOR RUNNING IN BATTERY WITH SAME RESULT. THE REPORTED SENSOR WAS COVERED WITH TOWEL AND THE WAVEFORM IMPROVED AND CORRECT SPO2 READING WAS OBTAINED. THE TOWEL COVERING THE SENSOR WAS THEN REMOVED AND WAVEFORM HAD GOOD QUALITY. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18239154 ↗

FDA clearance and classification

Cleared through 510(k) K041867 on Sep 10, 2004 (decision: Substantially Equivalent). Submitted by Hygia Health Services, Inc. as “Hygia Health Services Reprocessed Oximax Sensors Model # Hhs-Max-A, Hhs-Max-Al, Hhs-Max-N”. The FDA review took 63 days. FDA record ↗

The FDA regulates product code NLF (oximeter, reprocessed) under 21 CFR 870.2700 as a Class II device. 21 CFR 870.2700 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Nellcor MAX-A

Has the Nellcor MAX-A been recalled?

No FDA recall record in our data names this exact device identifier (DI 10817357020011), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Nellcor MAX-A?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Nov 30, 2023 and Dec 15, 2023. In those reports, the most common specific device problems are “Communication or Transmission Problem” (1), “Failure to Analyze Signal” (1) and “Incorrect Measurement” (1); the most common patient problem is “Ulcer” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Nellcor MAX-A?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Nellcor MAX-A FDA approved?

It was cleared through 510(k) K041867 on Sep 10, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the Nellcor MAX-A MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Nellcor MAX-A, and is it still sold?

The FDA's device database lists Hygia Health Services, Inc. as the company responsible for it (brand Nellcor, model MAX-A). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.