Nellcor MAX-ARecalls & reported problems
Hygia Health Services, Inc. · Nellcor OxiMax Adult Pulse Oximeter Sensor · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10817357020011), received between Nov 30, 2023 and Dec 15, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Communication or Transmission Problem1
- Failure to Analyze Signal1
- Incorrect Measurement1
- Patient Device Interaction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Ulcer1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 15, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
ACCORDING TO THE REPORTER, DURING USE, THE PATIENT HAD PRESSURE ULCER WHILE USING THE SENSOR. THE PATIENT'S HAND WERE SWOLLEN AFTER USING THE DEVICE.
Patient problems: Ulcer
MAUDE report 18339204 ↗MalfunctionNov 30, 2023Incorrect Measurement; Failure to Analyze Signal; Communication or Transmission Problem; Incorrect Measurement; Failure to Analyze Signal; Communication or Transmission Problem
ACCORDING TO THE REPORTER, THE SENSOR HAD NO SPO2 READINGS OR SIGNAL ACQUISITION. WHEN USED ON A PATIENT, A READING OF 100% WAS COMING AND GOING AND IT HAD A NOTICEABLE INTERFERENCE IN TRACE. A WRAP WAS USED TO SECURE IT ON PATIENT'S FINGER AND SITE WAS CHECKED EVERY 4 HOURS. THE SAME SENSOR WAS PLUGGED IN A DIFFERENT MONITOR AND GOT SOLID READING OF 90%, HOWEVER WAVEFORM WAS NOT VERY CLEAN. ANOTHER SENSOR WAS PLUGGED INTO DIFFERENT MODULE AND HAD SAME RESULT AS BEFORE OF NOISY TRACE AND INTERMITTENT INCORRECT READING. THE MODULE WAS REMOVED FROM THE MAIN MONITOR AND PLUGGED INTO A PORTABLE MONITOR RUNNING IN BATTERY WITH SAME RESULT. THE REPORTED SENSOR WAS COVERED WITH TOWEL AND THE WAVEFORM IMPROVED AND CORRECT SPO2 READING WAS OBTAINED. THE TOWEL COVERING THE SENSOR WAS THEN REMOVED AND WAVEFORM HAD GOOD QUALITY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18239154 ↗
Other Oximeter, Reprocessed devices
- N/AStryker Sustainability Solutions, Inc. · OXIMETER, REPROCESSED4 reports
- Nellcor MAXNRCovidien LP · OXIMETER, REPROCESSED3 reports
- N/AStryker Sustainability Solutions, Inc. · OXIMETER, REPROCESSED2 reports
- MEDLINE RENEWAL MAXARMedline Industries, Inc. · Oximeter, reprocessed2 reports
- MEDLINE RENEWAL 1860RMedline Industries, Inc. · Oximeter, reprocessed1 report
- N/AStryker Sustainability Solutions, Inc. · OXIMETER, REPROCESSED1 report
- Nellcor MAXARCovidien LP · OXIMETER, REPROCESSED1 report
- Nellcor MAX-ALHygia Health Services, Inc. · Oximeter, Reprocessed1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.