Skip to content

N/A (REF 5921-030-235)Recalls & reported problems

Stryker Sustainability Solutions, Inc. · Tourniquet Color Cuff Dual Port, Single Bladder (Royal Blue) Quick Connect · Class 1

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Stryker Color Cuff dual-port tourniquet cuff is a disposable inflatable band used to limit blood flow during surgery. The 38 reports all describe the cuff leaking air or losing pressure during use, sometimes from loose O-rings or tearing fabric at the connection; no deaths or injuries were reported. One open recall, started in March 2023, cites cuffs that fail to reach or hold the desired pressure because of fabric tears and pinched or unseated O-rings at the connections.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 00885825016166), received between Jun 30, 2021 and Oct 30, 2023 — classified as 38 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (38 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 1, 2023Open, Classified

    Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications.

    Z-1315-2023 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash38

Patient problems

  • No Clinical Signs, Symptoms or Conditions38

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 30, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE DEVICE WAS DEFECTIVE AS IT HAD AN AIR LEAK. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18031366 ↗
  • MalfunctionAug 4, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE DEVICE STARTED LEAKING AT THE CONNECTORS WHILE ON A PATIENT DURING A CASE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17465248 ↗
  • MalfunctionJun 22, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE DEVICE WAS LEAKING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. THE COMPLAINANT IS NOT AWARE OF ANY EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17187672 ↗
  • MalfunctionJun 22, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE DEVICE WAS LEAKING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. THE COMPLAINANT IS NOT AWARE OF ANY EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17187630 ↗
  • MalfunctionApr 13, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE DEVICE WAS LEAKING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS LESS THAN 30 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16738769 ↗
  • MalfunctionApr 13, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE DEVICE WAS LEAKING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS LESS THAN 30 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16738764 ↗
  • MalfunctionApr 6, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE RUBBER O-RINGS WERE LOOSE, AIR WAS AUDIBLY LEAKING FROM THE CONNECTION POINT, AND THE DEVICE PUMP WAS BEEPING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION REGARDING EXTENDED PROCEDURE TIME. NO INFORMATION WAS PROVIDED, HOWEVER, THERE WAS NO EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16689609 ↗
  • MalfunctionApr 6, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE RUBBER O-RINGS WERE LOOSE, AIR WAS AUDIBLY LEAKING FROM THE CONNECTION POINT, AND THE DEVICE PUMP WAS BEEPING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION REGARDING EXTENDED PROCEDURE TIME. NO INFORMATION WAS PROVIDED, HOWEVER, THERE WAS NO EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16689603 ↗
  • MalfunctionApr 5, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE DEVICE LEAKED AND COULD NOT MAINTAIN PRESSURE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16686288 ↗
  • MalfunctionMar 31, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE DEVICE WAS NOT INFLATE DUE TO A LARGE TEAR. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND THE EXTENDED PROCEDURE TIME REPORTED WAS THREE MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16659970 ↗
  • MalfunctionMar 30, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED CUFF NOT HOLDING PRESSURE AND LEAKING DURING CASE WHILE LIMB WAS MANIPULATED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16652098 ↗
  • MalfunctionMar 29, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFF LEAKED DURING A CASE. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16640707 ↗
  • MalfunctionMar 29, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFFS WOULDN'T INFLATE WHEN CONNECTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16640706 ↗
  • MalfunctionMar 29, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFF WILL NOT HOLD INFLATION. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16640705 ↗
  • MalfunctionMar 29, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFF WILL NOT HOLD INFLATION. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16640703 ↗
  • MalfunctionMar 29, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE DEVICE WAS LEAKING THROUGHOUT THE ENTIRE CASE AND THE NURSE WAS QUIETING NOISE ON THE PUMP MACHINE DURING THE 35 MINUTES OF DEVICE INFLATION TIME. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16640271 ↗
  • MalfunctionMar 29, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFFS WOULDN'T INFLATE WHEN CONNECTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16640110 ↗
  • MalfunctionMar 29, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFFS WOULDN'T INFLATE WHEN CONNECTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16640102 ↗
  • MalfunctionMar 29, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFF WILL NOT HOLD INFLATION. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16640033 ↗
  • MalfunctionMar 28, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED VARIOUS SIZES OF TOURNIQUET CUFFS DEVICES WERE LEAKING AIR. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16630650 ↗
  • MalfunctionMar 28, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED VARIOUS SIZES OF TOURNIQUET CUFFS DEVICES WERE LEAKING AIR. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16629948 ↗
  • MalfunctionMar 27, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFF WOULD NOT HOLD INFLATION OR PRESSURE DURING THE PROCEDURE, FLUCTUATING BY 5-10 MM UP AND DOWN. THE O-RING IS SLIPPING OUT OF PLACE ONCE CONNECTED TO THE SMART PUMP'S BLACK TUBING. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16625027 ↗
  • MalfunctionMar 27, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFF WOULD NOT HOLD INFLATION OR PRESSURE DURING THE PROCEDURE, FLUCTUATING BY 5-10 MM UP AND DOWN. THE O-RING IS SLIPPING OUT OF PLACE ONCE CONNECTED TO THE SMART PUMP'S BLACK TUBING. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16625025 ↗
  • MalfunctionMar 27, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED TOURNIQUET CUFF WOULD NOT HOLD INFLATION OR PRESSURE DURING THE PROCEDURE, FLUCTUATING BY 5-10 MM UP AND DOWN. THE O-RING IS SLIPPING OUT OF PLACE ONCE CONNECTED TO THE SMART PUMP'S BLACK TUBING. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16624993 ↗
  • MalfunctionMar 27, 2023Leak/Splash; Leak/Splash

    IT WAS REPORTED THE TOURNIQUET CUFF LEAK WOULD NOT FILL WITH AIR. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16624938 ↗

Other TOURNIQUET, PNEUMATIC devices

All TOURNIQUET, PNEUMATIC devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.