SmartPump 5920011000Recalls & reported problems

Stryker Corporation · Dual Channel Pump · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker SmartPump dual channel pump is used to control pressure in a pneumatic tourniquet, which limits bleeding during surgery. Most of the 72 reports describe the pump failing to hold the intended pressure — either dropping or rising — and many of these were caught before the device was used on a patient. One injury was reported, a bruise or contusion, alongside numbness, and there were no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
72
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 72 adverse event reports that list this exact device (DI 07613154589647), received between Jun 14, 2018 and Apr 22, 2026 — classified as 1 injury, 71 malfunctions.

The most frequently reported specific device problem is “Decrease in Pressure” (49 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Decrease in Pressure49
  • Increase in Pressure16
  • Pressure Problem10
  • Gas/Air Leak2
  • Adverse Event Without Identified Device or Use Problem1
  • Excess Flow or Over-Infusion1
  • Inappropriate Audible Prompt/Feedback1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions38
  • No Consequences Or Impact To Patient29
  • Insufficient Information12
  • No Patient Involvement3
  • Bruise/Contusion1
  • Erythema1
  • Hemorrhage/Blood Loss/Bleeding1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 72 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 22, 2026Decrease in Pressure

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2026. THIS REPORT SUMMARIZES 0 EVENTS FOR THE FAILURE: DECREASE IN PRESSURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24954388 ↗
  • MalfunctionJan 23, 2026Decrease in Pressure

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1 ¿ DECEMBER 31, 2025. THIS REPORT SUMMARIZES 16 EVENTS FOR THE FAILURE: DECREASE IN PRESSURE. - 6 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. - 5 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. - 3 EVENTS HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD INSUFFICIENT INFORMATION. - 1 EVENT HAD MINOR INJURY/ILLNESS/IMPAIRMENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24163096 ↗
  • MalfunctionOct 28, 2025Decrease in Pressure; Decrease in Pressure

    THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2025 ¿ (B)(6) 2025. THIS REPORT SUMMARIZES 11 EVENTS FOR THE FAILURE: DECREASE IN PRESSURE. - 9 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 1 EVENT HAD INSUFFICIENT INFORMATION.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23399775 ↗
  • MalfunctionOct 23, 2025Increase in Pressure; Increase in Pressure

    NO CHANGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23369356 ↗
  • MalfunctionJul 29, 2025Decrease in Pressure; Decrease in Pressure

    THIS REPORT SUMMARIZES 11 EVENTS FOR THE FAILURE: DECREASE IN PRESSURE. 1 EVENT HAD PROBLEM IDENTIFIED DURING NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 8 EVENTS HAD PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. 2 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22647052 ↗
  • MalfunctionJul 29, 2025Decrease in Pressure; Increase in Pressure; Decrease in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURES: DECREASE IN PRESSURE; INCREASE IN PRESSURE. 1 EVENT HAD INSUFFICIENT INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22647006 ↗
  • MalfunctionApr 21, 2025Decrease in Pressure; Increase in Pressure; Decrease in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE AND THEN AN INCREASE IN PRESSURE. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21876279 ↗
  • MalfunctionFeb 20, 2025Decrease in Pressure; Decrease in Pressure

    IT WAS REPORTED THAT THE DEVICE DOES NOT HOLD PRESSURE. THERE WAS NO REPORT OF PATIENT INVOLVEMENT OR ADVERSE CONSEQUENCES.

    Patient problems: Insufficient Information

    MAUDE report 21420622 ↗
  • MalfunctionJan 22, 2025Decrease in Pressure; Decrease in Pressure

    THIS REPORT SUMMARIZES 24 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE. 3 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 16 EVENTS HAD THE PROBLEM IDENTIFIED BEFORE CLINICAL USE/EXPOSURE; THERE WAS NO PATIENT INVOLVEMENT. 4 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED REGARDING HEALTH IMPACT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21207550 ↗
  • MalfunctionJan 22, 2025Increase in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCREASE IN PRESSURE. - 3 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21207493 ↗
  • MalfunctionOct 21, 2024Increase in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCREASE IN PRESSURE. - 2 EVENTS HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20498532 ↗
  • MalfunctionOct 21, 2024Decrease in Pressure; Decrease in Pressure

    THIS REPORT SUMMARIZES 19 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE. 10 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. 5 EVENTS HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT. 4 EVENTS HAD INSUFFICIENT INFORMATION RECEIVED REGARDING HEALTH IMPACT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20498498 ↗
  • MalfunctionOct 21, 2024Decrease in Pressure; Increase in Pressure; Decrease in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A PRESSURE IRREGULARITY. 3 EVENTS WHICH HAD A MILD INJURY, ILLNESS OR IMPAIRMENT WHICH CAN BE TREATED WITH MINIMAL OR NO INTERVENTION.

    Patient problems: Bruise/Contusion

    MAUDE report 20493932 ↗
  • InjurySep 9, 2024Increase in Pressure; Increase in Pressure

    IT WAS REPORTED THAT AFTER REMOVING THE TOURNIQUET IN THE OPERATING ROOM, IT WAS NOTED THAT THE PATIENT'S SKIN WAS WARM, DRY, AND BLANCHABLE. WHEN THE SKIN ASSESSMENT WAS COMPLETED IN THE POST ANESTHESIA CARE UNIT, THE REGISTERED NURSE INSPECTED THE RIGHT LOWER EXTREMITY AND NOTED THAT THE PATIENT HAD CIRCUMFERENTIAL SWELLING IN THE RIGHT THIGH AT THE TOURNIQUET SITE WITH REDNESS AND BLANCHABLE. IT WAS FURTHER REPORTED THAT THE PATIENT SUFFERED ERYTHEMA AND SWELLING/ EDEMA AS WELL AS DEVELOPING COMPARTMENT SYNDROME, WHICH REQUIRED A FASCIOTOMY. THERE WAS NO DELAY IN THE PROCEDURE AS THE DEVICE WAS USED IN THE PRIMARY CASE, AND THE COMPLICATION OF THE COMPARTMENT SYNDROME WAS DIAGNOSED POST-OPERATIVELY. APPROPRIATE INTERVENTIONS WERE PERFORMED WITHOUT DELAY.

    Patient problems: Erythema; Swelling/ Edema

    MAUDE report 20175793 ↗
  • MalfunctionJul 24, 2024Decrease in Pressure; Increase in Pressure; Decrease in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE AND THEN AN INCREASE IN PRESSURE. - 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823246 ↗
  • MalfunctionJul 24, 2024Decrease in Pressure; Decrease in Pressure

    THIS REPORT SUMMARIZES 32 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE. - 18 EVENTS HAD THE PROBLEM IDENTIFIED DURING A NON-CLINICAL PROCEDURE; THERE WAS NO PATIENT INVOLVEMENT. - 11 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 3 EVENTS HAD INSUFFICIENT INFORMATION REGARDING HEALTH IMPACT RECEIVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19823144 ↗
  • MalfunctionJul 24, 2024Increase in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY HAD AN INCREASE IN PRESSURE. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19818513 ↗
  • MalfunctionApr 17, 2024Decrease in Pressure; Increase in Pressure; Decrease in Pressure; Increase in Pressure

    IT WAS REPORTED THAT THE DEVICE IS ALARMING WHEN INFLATED PAST NINETY; THE PRESSURE THEN BEGINS TO SWING WILDLY. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING CLEANING AT THE END USER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19129005 ↗
  • MalfunctionJan 22, 2024Decrease in Pressure; Decrease in Pressure

    THIS REPORT SUMMARIZES 30 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE. 20 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 10 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18557608 ↗
  • MalfunctionJan 22, 2024Increase in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY HAD AN INCREASE IN PRESSURE. - 1 EVENT HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18555913 ↗
  • MalfunctionOct 23, 2023Decrease in Pressure; Decrease in Pressure

    THIS REPORT SUMMARIZES 31 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE. 24 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. 6 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17993144 ↗
  • MalfunctionOct 23, 2023Decrease in Pressure; Increase in Pressure; Decrease in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE AND THEN AN INCREASE IN PRESSURE. 1 EVENT HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17990689 ↗
  • MalfunctionJul 26, 2023Decrease in Pressure; Decrease in Pressure

    THIS REPORT SUMMARIZES 44 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE. - 34 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 9 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT HAD INSUFFICIENT INFORMATION RECEIVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17401587 ↗
  • MalfunctionJul 26, 2023Increase in Pressure; Increase in Pressure

    THIS REPORT SUMMARIZES 4 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCREASE IN PRESSURE. - 4 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17397014 ↗
  • MalfunctionApr 25, 2023Decrease in Pressure; Decrease in Pressure

    THIS REPORT SUMMARIZES 16 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD A DECREASE IN PRESSURE. - 2 EVENTS HAD NO PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 14 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16807849 ↗

FDA clearance and classification

The FDA regulates product code KCY (tourniquet, pneumatic) under 21 CFR 878.5910 as a Class I device. 21 CFR 878.5910 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SmartPump 5920011000

Has the SmartPump 5920011000 been recalled?

No FDA recall record in our data names this exact device identifier (DI 07613154589647), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SmartPump 5920011000?

The FDA's MAUDE database lists 72 adverse event reports for this device, received between Jun 14, 2018 and Apr 22, 2026. In those reports, the most common specific device problems are “Decrease in Pressure” (49), “Increase in Pressure” (16) and “Pressure Problem” (10); the most common patient problems are “Bruise/Contusion” (1), “Erythema” (1) and “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SmartPump 5920011000?

None of the reports are classified as deaths, 1 as an injury and 71 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SmartPump 5920011000 FDA approved?

Devices of this type (product code KCY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker Corporation.

Who makes the SmartPump 5920011000, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand SmartPump, model 5920011000). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other TOURNIQUET, PNEUMATIC devices

All TOURNIQUET, PNEUMATIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page