BioPince™ Full Core Biopsy Instrument 370-1580-01Recalls & reported problems

Argon Medical Devices, Inc. · BioPince™ Full Core Biopsy Instrument 16G x 15cm · Class 2

All models of BioPince Full Core Biopsy Instrument (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00886333004362), received between Aug 6, 2020 and Dec 12, 2024 — classified as 3 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K904987); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 2, 2024Class IIOpen, Classified

    Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.

    Z-3192-2024 · Root cause: Packaging change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Therapeutic or Diagnostic Output Failure3
  • Defective Device2
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Detachment of Device or Device Component1
  • Difficult to Remove1
  • Mechanical Problem1
  • Scratched Material1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding3
  • Insufficient Information3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Injury1
  • Necrosis1
  • Needle Stick/Puncture1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THERE WAS A SMALL AMOUNT OF BLEEDING DURING THE OPERATION, AND SYMPTOMATIC HEMOSTATIC TREATMENT WAS GIVEN. AFTER RE EXAMINATION OF CT, NO CLEAR BLEEDING WAS FOUND. DURING THE PUNCTURE PROCESS, A FOLLOW-UP CT SCAN SHOWED A SMALL AMOUNT OF PNEUMOTHORAX ON THE LEFT SIDE. THE HOSPITAL REPORTED THIS CASE TO NMPA.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Pneumothorax

    MAUDE report 20915346 ↗
  • MalfunctionNov 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AFTER TAKING LIVER TISSUE, IT WAS FOUND THAT THE TISSUE WAS NOT COMPLETELY REMOVED POSTOPERATIVE OBSERVATION SHOWS THAT THE DETACHMENT FRAGMENT AT THE PUNCTURE NEEDLE SITE IS BROKEN, AND THE PATIENT EXPERIENCES SIGNIFICANT PAIN DURING AND AFTER THE OPERATION.

    Patient problems: Pain

    MAUDE report 20765542 ↗
  • InjuryDec 5, 2022Defective Device

    WHEN THE PATIENT WAS PERFORMING LIVER PUNCTURE UNDER THE GUIDANCE OF B-ULTRASOUND, THE BIOPSY NEEDLE WAS USED. NO TISSUE WAS TAKEN FROM THE PUNCTURE TWICE, ONLY BLOOD. LATER, THE PUNCTURE NEEDLE OF OTHER PRODUCTS WAS REPLACED SUCCESSFULLY. THE DOCTOR BELIEVED THAT THE CONDITION PROLONGED THE OPERATION TIME AND CAUSED SECONDARY INJURY TO THE PATIENT, SO HE REPORTED THE INCIDENT TO THE COMPETENT AUTHORITY.

    Patient problems: Insufficient Information

    MAUDE report 15916614 ↗
  • MalfunctionNov 17, 2022Difficult to Remove; Therapeutic or Diagnostic Output Failure

    EMPTY SAMPLE. NO SAMPLE WHEN THE BIOTWEEZER IS REMOVED. DIFFICULTY REMOVING THE BIOWEEZER FROM THE COAXIAL SHEATH. NEED TO REPEAT A NEW BIOPSY INCREASING THE RISK OF BLEEDING.

    Patient problems: Needle Stick/Puncture

    MAUDE report 15812137 ↗
  • MalfunctionJun 9, 2022Therapeutic or Diagnostic Output Failure

    THE TISSUE COULD NOT BE TAKEN OUT; THIS INCIDENT WAS USED WITH A COAXIAL GUIDE NEEDLE. THE GUN WAS NOT TESTED BEFORE USE. WHEN THE INSTRUMENT WAS TAKEN OUT, IT WAS ROTATED 360 DEGREES. THE RANGE OF USE WAS 23MM. THE EMPTY GUN OCCURRED AT THE FIRST NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14643373 ↗
  • MalfunctionMay 6, 2022Mechanical Problem

    THE PATIENT UNDERWENT A LUNG BIOPSY WITH 5 PUNCTURES WITH A BIOPSY NEEDLE, BUT NO TISSUE WAS REMOVED. CHANGED A GUN AND DIDN'T REMOVE THE TISSUE. WHEN THE THIRD GUN WAS USED, THE TISSUE WAS SUCCESSFULLY REMOVED. ; THIS INCIDENT IS USED TOGETHER WITH THE COAXIAL GUIDE NEEDLE. THE GUN HAS BEEN TESTED BEFORE USE. WHEN THE INSTRUMENT IS TAKEN OUT, IT IS ROTATED 360 DEGREES. THE RANGE OF USE IS 23MM. THE EMPTY GUN OCCURS AT THE FIRST NEEDLE.

    Patient problems: Insufficient Information

    MAUDE report 14304711 ↗
  • InjuryJul 21, 2021Break

    RENAL PUNCTURE WAS PERFORMED IN THREE PATIENTS, AND BLEEDING OCCURRED IN TWO PATIENTS; ANOTHER PATIENT WAS UNABLE TO TAKE OUT THE TISSUE AT THE FIRST PUNCTURE, AND THE NEEDLE IS DAMAGED. FOUR BP WERE REPLACED CONTINUOUSLY. AT THE FIFTH PUNCTURE, A LARGE AMOUNT OF BLEEDING AND NECROSIS OF THE RIGHT KIDNEY OCCURRED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Necrosis

    MAUDE report 12205020 ↗
  • MalfunctionFeb 24, 2021Defective Device

    DURING THE BIOPSY ON MASS FRAGMENTS OF THE RECTUS MUSCLES OF A PATIENT NO SAMPLE WAS TAKEN AFTER 3 ATTEMPTS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11376791 ↗
  • MalfunctionFeb 17, 2021Therapeutic or Diagnostic Output Failure

    BIOPINCE: 16GX15CM WAS USED FOR LUNG BIOPSY AND THE DEVICE DIDN¿T FETCH ANY SAMPLE EVEN AFTER MULTIPLE FIRING. DOCTOR USED BARD BIOPSY GUN AND IT FETCHED SAMPLE .

    Patient problems: Insufficient Information

    MAUDE report 11340862 ↗
  • MalfunctionAug 6, 2020Scratched Material

    FOLLOW UP.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Injury

    MAUDE report 10376905 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K904987. One submission often covers many models, so check the product and lot details in each record.

  • Z-2427-2020 · Dec 2, 2019 · Terminated
    BioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS - Product Usage: The disposable, automatic BioPince Full Core Biopsy Ins

Other Instrument, Biopsy devices

All Instrument, Biopsy devices →