Skip to content

BioPince™ Full Core Biopsy Instrument 370-1080-01Recalls & reported problems

Argon Medical Devices, Inc. · BioPince™ Full Core Biopsy Instrument 16G x 10cm · Class 2

All models of BioPince Full Core Biopsy Instrument (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00886333004386), received between Feb 3, 2021 and Jul 6, 2024 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Failure to Obtain Sample” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K904987); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 2, 2024Class IIOpen, Classified

    Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.

    Z-3192-2024 · Root cause: Packaging change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Obtain Sample3
  • Crack1
  • Improper or Incorrect Procedure or Method1
  • Material Twisted/Bent1
  • Mechanics Altered1
  • Therapeutic or Diagnostic Output Failure1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information2
  • Discomfort1
  • Pain1
  • Unspecified Mental, Emotional or Behavioral Problem1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 6, 2024Failure to Obtain Sample; Failure to Obtain Sample

    DEVICE USED FOR A LUNG BIOPSY. DESPITE APPARENTLY NORMAL OPERATION, THAT IS TO SAY THAT THE DEVICE RETRACTED WHEN THE USER PRESSED THE ¿RETRACTION BUTTON¿ BUT NO TISSUE WAS TAKEN.AFTER FOUR UNSUCCESSFUL ATTEMPTS WITH THIS DEVICE, ANOTHER BIOTWEEZER FROM ANOTHER BATCH WAS USED WITHOUT ANY PROBLEMS. MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: Insufficient Information

    MAUDE report 19685845 ↗
  • MalfunctionOct 16, 2023Failure to Obtain Sample; Material Twisted/Bent; Failure to Obtain Sample; Material Twisted/Bent

    THE SURGEON USED THE NEEDLE FOR THE FIRST CORE BIOPSY SPECIMEN. THE PATIENT EXPERIENCED A LOT OF PAIN AND DISCOMFORT. UPON OBTAINING THE FIRST SPECIMEN HE EXAMINED THE NEEDLE AND SAW IT WAS BENT. HE THEN CHANGED TO ANOTHER NEEDLE, THE SAME TYPE AND SIZE. TWO MORE SPECIMENS WERE OBTAINED. THE PATIENT DID NOT EXPERIENCED PAIN AND DISCOMFORT WITH THE 2ND NEEDLE.

    Patient problems: Discomfort; Pain

    MAUDE report 17941841 ↗
  • MalfunctionAug 18, 2022Use of Device Problem; Mechanics Altered

    ON JULY 8, LYMPH NODE PUNCTURE, THE TISSUE COULD NOT BE REMOVED DURING THE OPERATION. AFTER TAKING OUT THE BIOPSY NEEDLE AND TRYING SEVERAL TIMES, IT WAS FOUND THAT THE SPRING FELL OFF. THE DIRECTOR SAID THAT IT HAD CAUSED PSYCHOLOGICAL AND PHYSICAL DAMAGE TO DOCTORS AND PATIENTS.; THIS EVENT WAS NOT USED WITH A COAXIAL GUIDE NEEDLE. THE GUN WAS NOT TESTED BEFORE USE. THE INSTRUMENT WAS NOT ROTATED 360 DEGREES WHEN IT WAS TAKEN OUT. THE RANGE OF USE WAS 23MM. THE EMPTY GUN OCCURRED AT THE FIRST NEEDLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Mental, Emotional or Behavioral Problem; Unspecified Tissue Injury

    MAUDE report 15248655 ↗
  • InjuryMay 16, 2022Crack; Improper or Incorrect Procedure or Method

    THE NEEDLE WAS CRACKED AND SAMPLING WAS NOT POSSIBLE; THIS INCIDENT WAS NOT USED WITH A COAXIAL GUIDE NEEDLE, THE GUN WAS NOT TESTED BEFORE USE, THE INSTRUMENT WAS NOT ROTATED 360 DEGREES WHEN IT WAS TAKEN OUT, THE RANGE OF USE WAS 23MM, AND THE EMPTY GUN OCCURRED AT THE FIRST NEEDLE.

    Patient problems: Insufficient Information

    MAUDE report 14407569 ↗
  • MalfunctionApr 21, 2022Therapeutic or Diagnostic Output Failure

    PHYSICIAN REPORTS THAT ONCE THE DEVICE IS USED, NO SAMPLE IS TAKEN. AND IN ONE OF THE THREE USED UNITS, THE DOCTOR SAW RECOIL OF THE TAP THAT HOLDS THE SAMPLE WITH THE DANGER OF TEARING FOR THE PATIENT. HOSPITAL NEEDS A LETTER EXPLAINING THE ROOT CAUSE OF THE INCIDENTS AND WACREES NEEDS CREDIT NOTE PER 3 UNITS OF THE SAME REFERENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14165230 ↗
  • MalfunctionFeb 3, 2021Failure to Obtain Sample

    PHYSICIAN REPORTS THAT DURING PROCEDURE 3 UNITS OF THE DEVICE, THEY DO NOT TAKE SPECIMEN AND IN ONE OF THE UNITS, THE RECEDING OF THE TAB THAT HOLDS THE SAMPLE IS OBSERVED, WHICH IS A RISK OF TEARING FOR THE PATIENT. IT WAS A FAILURE WHEN PATIENT WAS PUNCTURED WITH THE NEEDLE AND NO SAMPLE WAS CAPTURED. NO DAMAGE OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11276604 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K904987. One submission often covers many models, so check the product and lot details in each record.

  • Z-2427-2020 · Dec 2, 2019 · Terminated
    BioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS - Product Usage: The disposable, automatic BioPince Full Core Biopsy Ins

Other Instrument, Biopsy devices

All Instrument, Biopsy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page