Tru-Core™ II Biopsy Instrument 763120160XRecalls & reported problems
Argon Medical Devices, Inc. · Tru-Core™ II Biopsy Instrument 20ga x 16cm · Class 2
All models of Tru-Core II Biopsy Instrument (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00886333006571), received between Jul 21, 2022 and Oct 23, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jul 2, 2024Class IIOpen, Classified
Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.
Z-3192-2024 · Root cause: Packaging change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Deformation2
- Break1
- Entrapment of Device1
- Therapeutic or Diagnostic Output Failure1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 23, 2022Break; Entrapment of Device; Material Deformation
MOVED TO A POSTERIOR APPROACH FOR THE NEXT ATTEMPT. AFTER IMAGING THEY NOTICED A SMALL METALLIC OBJECT FOR THE FIRST TIME AND WHEN HE LOOKED AT THE BIOPSY DEVICE HE NOTICED THE TIP MISSING AND SLIGHT DEFORMED LOOK. THE DEVICE ONLY HAD ONE PASS WHEN TIP BROKE OFF. THEY USED ANOTHER TRU CORE DEVICE SUCCESSFULLY. THE NODULE WAS NEAR RIB CAGE BONE AND IT IS A POSSIBILITY CONTACT TO BONE WAS MADE, BUT THE OPERATOR DID NOT FEEL ANY TACTILE FEEDBACK THAT SUGGESTED THIS. THE PATIENT SAW FURTHER EVALUATION TO SEE IF SURGICAL REMOVAL WAS NEEDED AND THE SURGEON ADVISED AGAINST IT. THE TIP IS IN THE PLURAL SPACE OF THE PATIENT. NOTE- UPON RECEIVING THE DEVICE, THE TIP IS BENT AT NEARLY A 90 DEGREE ANGLE. THE OPERATING DOCTOR SAID THIS TYPE OF BEND DID NOT OCCUR DURING THE PROCEDURE AND MUST HAVE HAPPENED AFTER HANDLING FOR THE PER. THE AREA OF FOCUS IS THE DISTAL TIP. ARGON 19G COAXIAL INTRODUCER MCXS2016TY.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 15656408 ↗MalfunctionJul 21, 2022Material Deformation; Therapeutic or Diagnostic Output Failure
THE TIP OF THE NEEDLE IS DAMAGED AND THE TISSUE CANNOT BE REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15075838 ↗
Other Instrument, Biopsy devices
- MAX-CORE MC1616Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall4,258 reports
- ENCOR® ECP017GSenorx, Inc. · INSTRUMENT, BIOPSY1 recall4,186 reports
- MAX-CORE MC1820Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall2,404 reports
- MAX-CORE MC1816Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall1 death848 reports
- MAX-CORE MC1410Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall399 reports
- MAX-CORE MC1610Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall394 reports
- ENCOR® ECP0110GVSenorx, Inc. · INSTRUMENT, BIOPSY392 reports
- ENCOR® ECP0110GSenorx, Inc. · INSTRUMENT, BIOPSY390 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page