The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00886333216215), received between Oct 11, 2023 and Oct 25, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Too Rigid or Stiff” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Too Rigid or Stiff2
  • Device Misassembled During Manufacturing /Shipping1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 25, 2023Material Too Rigid or Stiff; Material Too Rigid or Stiff

    DURING PROCEDURE, A PHYSICIAN STOPPED USING THE GUIDEWIRE AS THE GUIDEWIRE SOFT TIP(DISTAL TIP) WAS ABNORMALLY STIFFER THAN USUAL. THE PROCEDURE WAS COMPLETED BY USING AN ADDITIONALLY OPENED REPLACEMENT (A395231).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18003495 ↗
  • MalfunctionOct 11, 2023Material Too Rigid or Stiff; Device Misassembled During Manufacturing /Shipping; Material Too Rigid or Stiff; Device Misassembled During Manufacturing /Shipping

    DURING PROCEDURE, A PHYSICIAN STOPPED USING THE GUIDEWIRE AS THE GUIDEWIRE SOFT TIP(DISTAL TIP) WAS ABNORMALLY STIFFER THAN USUAL. A PHYSICIAN IS VERY CONCERN ABOUT THE ABNORMALLY STIFF TIP(DISTAL TIP) WHICH WAS SIMILAR STIFFNESS TO THE PROXIMAL STIFF TIP. -THE ABNORMALLY STIFF TIP(DISTAL TIP) CAN CAUSE SERIOUS INCIDENT SUCH AS CARDIAC PERFORATION ON CLINICAL USE. THE PROCEDURE WAS COMPLETED BY USING AN ADDITIONALLY OPENED REPLACEMENT((B)(6)).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17915064 ↗