BioPince Ultra Full Core Biopsy Instrument 360-1580-03Recalls & reported problems
Argon Medical Devices, Inc. · BioPince™ Ultra Full Core Biopsy Instrument with Co-Axial Introducer Needle (Biopsy Instrument: 18ga x 15cm)(Introducer Needle: 16ga x 11.8cm) · Class 2
All models of BioPince Ultra Full Core Biopsy Instrument (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00886333224036), received between Nov 2, 2021 and Jan 29, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Failure to Obtain Sample” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K904987); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 2, 2024Class IIOpen, Classified
Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.
Z-3192-2024 · Root cause: Packaging change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Obtain Sample2
- Misfire2
- Failure to Fire1
- Material Too Rigid or Stiff1
- Material Twisted/Bent1
- Sharp Edges1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Hemorrhage/Blood Loss/Bleeding2
- Insufficient Information2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 29, 2026Failure to Obtain Sample
A CUSTOMER REPORTED A PHYSICIAN WAS PERFORMING RENAL BIOPSY AND THE DEVICE WAS FIRING BUT NOT COLLECTING SAMPLES AFTER THE FIRST PASS. 1 OF THE 10 PASSES ATTEMPTED ONLY COLLECTED A SAMPLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24218109 ↗MalfunctionMay 19, 2025Misfire; Misfire
A RADIOLOGIST ATTEMPTED TO USE THE SIDE BUTTON TO FIRE THE DEVICE DURING A RENAL BIOPSY. THE SIDE BUTTON FAILED; THE PHYSICIAN THEN ADJUSTED HIS HAND TO USE THE TRIGGER AT THE BACK OF THE NEEDLE AND THE DEVICE DEPLOYED WITHOUT ANY BUTTONS BEING PUSHED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22047245 ↗MalfunctionNov 20, 2023
WHILE PROVIDER WAS PERFORMING BIOPSY, HE NOTICED THE NEEDLE HAD SOME RESISTANCE. HE REMOVED THE NEEDLE WITH THE SAMPLE AND INSPECTED THE NEEDLE, AND NOTICED SOME BARBS ON THE NEEDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18169134 ↗MalfunctionMar 13, 2023Failure to Obtain Sample; Failure to Obtain Sample
(B)(6) PERFORMED 4 PASSES WITH THE BIOPINCE ULTRA. AFTER 4 ATTEMPTS THERE WAS NO SAMPLE TISSUE PRESENT. DOCTOR TRIED THE DEVICE ON A GEL AND THERE WAS NO CORE CAPTURED EITHER. PATIENT EXPERIENCED SOME BLEEDING.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 16529930 ↗MalfunctionJan 24, 2023Material Too Rigid or Stiff; Material Twisted/Bent; Sharp Edges; Material Twisted/Bent
THE SHANK OF THE NEEDLE IS VERY STIFF. AND WHEN TAKING THE SAMPLE THE NEEDLE CAME OUT LIKE A CARVING KNIFE, WHICH CAUSED MORE THAN NORMAL BLEEDING IN THE PATIENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 16233394 ↗MalfunctionAug 29, 2022Misfire
BPU ¿MISFIRED WHILE IN PATIENTS BODY."
Patient problems: Insufficient Information
MAUDE report 15313094 ↗MalfunctionNov 2, 2021Failure to Fire
DURING BIOPSY, TRIED TO SAMPLE BIOPSY WITH BIOPINCE GUN. IT WOULD NOT FIRE PER MD. FDA SAFETY REPORT ID# (B)(4).
Patient problems: Insufficient Information
MAUDE report 12745280 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K904987. One submission often covers many models, so check the product and lot details in each record.
- Z-2427-2020 · Dec 2, 2019 · TerminatedBioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS - Product Usage: The disposable, automatic BioPince Full Core Biopsy Ins
Other Instrument, Biopsy devices
- MAX-CORE MC1616Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall4,258 reports
- ENCOR® ECP017GSenorx, Inc. · INSTRUMENT, BIOPSY1 recall4,186 reports
- MAX-CORE MC1820Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall2,404 reports
- MAX-CORE MC1816Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall1 death848 reports
- MAX-CORE MC1410Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall399 reports
- MAX-CORE MC1610Bard Peripheral Vascular, Inc. · INSTRUMENT, BIOPSY1 recall394 reports
- ENCOR® ECP0110GVSenorx, Inc. · INSTRUMENT, BIOPSY392 reports
- ENCOR® ECP0110GSenorx, Inc. · INSTRUMENT, BIOPSY390 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page