Skip to content

BioPince Ultra Full Core Biopsy Instrument 360-2080-03Recalls & reported problems

Argon Medical Devices, Inc. · BioPince™ Ultra Full Core Biopsy Instrument with Co-Axial Introducer Needle (Biopsy Instrument: 18ga x 20cm)(Introducer Needle: 16ga x 16.8cm) · Class 2

All models of BioPince Ultra Full Core Biopsy Instrument (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00886333224050), received between Jun 16, 2025 and Jun 16, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K904987); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 2, 2024Class IIOpen, Classified

    Holes have been identified in biopsy instrument packaging trays compromising the sterile barrier, which may lead to cross contamination, or the introduction of micro-organisms into the access site, which could lead to an infectious process, bacteremia, or sepsis.

    Z-3192-2024 · Root cause: Packaging change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 16, 2025Break; Break

    CUSTOMER REPORTED "THE INTRODUCER OF THE BPU BROKE OFF INSIDE THE PATIENT WHEN IT WAS BEING ADVANCED. ENDED UP HAVING A SURGEON REMOVE IT."

    Patient problems: Insufficient Information

    MAUDE report 22211685 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K904987. One submission often covers many models, so check the product and lot details in each record.

  • Z-2427-2020 · Dec 2, 2019 · Terminated
    BioPince Full Core Biopsy Instrument 16ga x 15cm.,Use with Optional CO-Axial Needle MCXS1615BP. QTY.: 5/CS - Product Usage: The disposable, automatic BioPince Full Core Biopsy Ins

Other Instrument, Biopsy devices

All Instrument, Biopsy devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page