Skip to content

Single-Loop Snare Retrieval Kit 380911120Recalls & reported problems

Argon Medical Devices, Inc. · Single-Loop Snare Retrieval Kit · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00886333224609), received between Jan 18, 2022 and May 29, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Fragmentation1

Patient problems

  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 29, 2025Break; Break

    CUSTOMER REPORTED, "RADIOPAQUE TIP BROKE OFF OF THE INTERMEDIATE CATHETER FOR THE SNARE. TIP LODGE IN THE HEART AND WAS UNABLE TO BE RETRIEVED. NO IDENTIFIED CONTRIBUTING FACTORS. DR. DENIED ANY USE OUTSIDE OF NORMAL PRACTICES. SLIGHT CURVED SHAPE WAS ADDED TO THE CATHETER FOR DIRECTION ABILITY BUT THIS IS NOT OUTSIDE OF NORMAL PRACTICES. NO IDENTIFIED CHANGES IN PRACTICES. USER WAS PROPERLY TRAINED. NO NOTED PATIENT PATHOLOGIES OR CONDITIONS THAT WOULD CONTRIBUTE TO TIP BREAKING OFF. PATIENT IS STABLE BUT WILL REQUIRE ADDITIONAL MONITORING. PER DR., PATIENT WILL HAVE ADDITIONAL X-RAY TODAY, IN ONE WEEK, AND IN 3 MONTHS TO MONITOR FOREIGN OBJECT."

    Patient problems: Foreign Body In Patient

    MAUDE report 22107664 ↗
  • MalfunctionJan 18, 2022Material Fragmentation

    AFTER THEY SNARED THE FILTER WITH THE RETRIEVAL KIT SNARE, THE DOCTOR SINCHED THE FILTER INTO THE 9FR SHEATH. THE DOCTOR THEN SHEATHED DOWN OVER THE FILTER WITH THE 11FR SHEATH AND RETRIEVED THE FILTER. THEN THE DOCTOR PUSHED SOME CONTRAST IN THE 11FR SHEATH FOR A FINAL CAVAGRAM, AND IT LOOKED GOOD. THEN WHEN HE PULLED THE 11FR SHEATH OUT IT HAD A STRING OF THE INNER SHEATH ATTACHED TO THE END OF THE 11FR SHEATH. IT WAS REMOVED FROM THE PATIENT WITHOUT INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13280433 ↗