Skip to content

CDI® Blood Parameter Monitoring System 550 550AHCTRecalls & reported problems

Terumo Cardiovascular Systems Corporation · Monitor with one blood parameter module and one H/S Probe. · Class 2

All models of CDI Blood Parameter Monitoring System 550 (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

The Terumo CDI Blood Parameter Monitoring System 550 is a bedside monitor used during cardiopulmonary bypass and ECMO to continuously measure blood gases, potassium, hematocrit and temperature. The 60 reports mostly describe readings that did not match the lab blood gas analyzer, along with a few shutdowns and one overheating and smoking report. No deaths or injuries were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
59
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 59 adverse event reports that list this exact device (DI 00886799001790), received between Mar 14, 2019 and Apr 7, 2026 — classified as 59 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (55 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings55
  • Failure to Power Up2
  • Electrical Shorting1
  • Overheating of Device1
  • Smoking1
  • Thermal Decomposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions43
  • No Consequences Or Impact To Patient20
  • Insufficient Information2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 25 most recent of 59 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 7, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) POTASSIUM (K+) AND CARBON DIOXIDE (CO2) VALUES WERE INACCURATE. THE K+ VALUE DISPLAYED WAS 5.7 WHICH DID NOT MATCH THE BLOOD GAS ANALYZER VALUE OF 4.6. AS MITIGATION, THE TEAM USED THE BLOOD GAS ANALYZER FOR THE REMAINDER OF THE CASE. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24808810 ↗
  • MalfunctionApr 7, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) POTASSIUM (K+) AND CARBON DIOXIDE (CO2) VALUES WERE INACCURATE. THE K+ VALUE DISPLAYED WAS 5.7 WHICH DID NOT MATCH THE BLOOD GAS ANALYZER VALUE OF 3.5. AS MITIGATION, THE TEAM USED THE BLOOD GAS ANALYZER FOR THE REMAINDER OF THE CASE. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24808684 ↗
  • MalfunctionApr 7, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) POTASSIUM (K+) AND CARBON DIOXIDE (CO2) VALUES WERE INACCURATE. THE K+ VALUE DISPLAYED WAS 8.9 WHICH DID NOT MATCH THE BLOOD GAS ANALYZER VALUE OF 4.8. AS MITIGATION, THE TEAM USED THE BLOOD GAS ANALYZER FOR THE REMAINDER OF THE CASE. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24807771 ↗
  • MalfunctionFeb 26, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE TEMPERATURE DISPLAYED ON THE BLOOD PARAMETER MONITOR (BPM) WAS VERY DIFFERENT FROM THE READING ON THE CENTRAL CONTROL MONITOR (CCM). THE BPM TEMPERATURE READING SWITCHED FROM 17 DEGREES CELSIUS TO 31 DEGREES CELSIUS INSTANTLY. IT WAS STATED THAT THE UNSTABLE TEMPERATURE READINGS APPEARED TO AFFECT OTHER VALUES. THE TEAM CONTINUED TO USE THE DEVICE FOR THE REMAINDER OF THE CASE USING BLOOD GAS ANALYZER VALUES. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24456732 ↗
  • MalfunctionFeb 23, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) BEING USED IN CONJUNCTION WITH THE NEONATAL EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) SUDDENLY SHUT DOWN AND THEN AUTOMATICALLY RESTARTED. AFTER RECONNECTING THE UNIT TO THE HEMATOCRIT SATURATION (H/S) PROBE AND PERFORMING CALIBRATION, THE DATA ON THE DISPLAY WAS INACCURATE. THE CARBON DIOXIDE (CO2) WAS 37 BUT THE BPM DISPLAYED 46. THE PARTIAL PRESSURE OXYGEN (PO2) WAS 322 BUT THE BPM DISPLAYED 364. ANOTHER UNEXPECTED SHUTDOWN WAS FOLLOWED BY AN AUTOMATIC RESTART. THE PROCEDURE WAS ATTEMPTED USING THE PREVIOUS CALIBRATION VALUES; HOWEVER, THE GAS DATA SIGNIFICANTLY DEVIATED FROM THE ACTUAL MEASUREMENTS. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24421900 ↗
  • MalfunctionFeb 19, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) BEING USED IN CONJUNCTION WITH THE NEONATAL EXTRACORPOREAL MEMBRANE OXYGENATION (ECMO) SUDDENLY SHUT DOWN AND THEN AUTOMATICALLY RESTARTED. AFTER RECONNECTING THE UNIT TO THE HEMATOCRIT SATURATION (H/S) PROBE AND PERFORMING CALIBRATION, THE DATA ON THE DISPLAY WAS INACCURATE. THE CARBON DIOXIDE (CO2) WAS 37 BUT THE BPM DISPLAYED 46. THE PARTIAL PRESSURE OXYGEN (PO2) WAS 322 BUT THE BPM DISPLAYED 364. ANOTHER UNEXPECTED SHUTDOWN WAS FOLLOWED BY AN AUTOMATIC RESTART. THE PROCEDURE WAS ATTEMPTED USING THE PREVIOUS CALIBRATION VALUES; HOWEVER, THE GAS DATA SIGNIFICANTLY DEVIATED FROM THE ACTUAL MEASUREMENTS. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24389930 ↗
  • MalfunctionNov 18, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) POTASSIUM (K+) READING WAS FALLING TO 1.3 EVEN THOUGH IT WAS JUST CALIBRATED AND THE K+ CODE ENTERED. AS MITIGATION, BLOOD GAS VALUES FROM THE BLOOD GAS ANALYZER WERE RELIED UPON DURING THE CASE. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23587438 ↗
  • MalfunctionOct 27, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE POTASSIUM (K+) AND CARBON DIOXIDE (CO2) DID NOT MATCH THE BLOOD GAS ANALYZER RESULTS AND WERE FLUCTUATING. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23397541 ↗
  • MalfunctionMay 30, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THE BLOOD PARAMETER MONITOR (BPM) WAS NOT READING PROPERLY. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22117602 ↗
  • MalfunctionJan 14, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) HAD INCORRECT AND INCONSISTENT CARBON DIOXIDE (CO2) READINGS. THERE WAS A 20-MINUTE DELAY TO ATTAIN CORRECT CO2 DATA. AS MITIGATION, A BLOOD GAS ANALYZER WAS USED THROUGHOUT THE CASE. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21154090 ↗
  • MalfunctionOct 10, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THERE WERE DIFFERENCES IN THE VALUES FROM THE BLOOD PARAMETER MONITOR (BPM) COMPARED TO THE BLOOD GAS ANALYZER (BGA). THE BGA WOULD SHOW THAT THE PATIENT HAD A SOMEWHAT LOWER PH THAN THE BPM INDICATED. BY THE THIRD SAMPLE, THE VALUES WOULD MATCH. AS MITIGATION, MORE BLOOD SAMPLES WERE PERFORMED TO OBTAIN MORE ACCURATE READINGS. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20422596 ↗
  • MalfunctionOct 10, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    ADDITIONAL INFORMATION WAS RECEIVED THAT THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20422359 ↗
  • MalfunctionOct 10, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THERE WERE DIFFERENCES IN THE VALUES FROM THE BLOOD PARAMETER MONITOR (BPM) COMPARED TO THE BLOOD GAS ANALYZER (BGA). THE BGA WOULD SHOW THAT THE PATIENT HAD A SOMEWHAT LOWER PH THAN THE BPM INDICATED. BY THE THIRD SAMPLE, THE VALUES WOULD MATCH. AS MITIGATION, MORE BLOOD SAMPLES WERE PERFORMED TO OBTAIN MORE ACCURATE READINGS. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20421561 ↗
  • MalfunctionJul 16, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) PARTIAL PRESSURE OF OXYGEN (PO2) WAS READING HIGH DESPITE CALIBRATION AND MULTIPLE BLOOD GAS RECALIBRATIONS. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19756525 ↗
  • MalfunctionApr 24, 2024Overheating of Device; Failure to Power Up; Overheating of Device; Failure to Power Up

    IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE BLOOD PARAMETER MONITOR (BPM) SCREEN WAS BLANK UPON POWER UP. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR FURTHER EVALUATION WHERE IT WAS FOUND THAT THERE WERE INTERNAL COMPONENTS THAT HAD EVIDENCE OF HEAT DAMAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19177614 ↗
  • MalfunctionNov 21, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) VENOUS OXYGEN SATURATION (SVO2) VALUE WAS DIFFERENT THAT THE BLOOD GAS ANALYSIS (BGA) VALUE BY 5-10%. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18181466 ↗
  • MalfunctionAug 4, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) POTASSIUM (K+) LEVELS WERE READING INCORRECTLY. THE BPM WAS RECALIBRATED MID CASE, IT WORKED FOR A SMALL PERIOD OF TIME, BEFORE THE READINGS WERE INCORRECT AGAIN. MANUAL BLOOD GAS ANALYSIS WAS USED FOR THE REMAINDER OF THE CASE. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17459862 ↗
  • MalfunctionMay 22, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR A NON-CLINICAL ACTIVITY, THE BLOOD PARAMETER MONITOR (BPM) MEASURED VALUE WAS INACCURATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16974113 ↗
  • MalfunctionMar 21, 2023Electrical Shorting; Electrical Shorting

    THE PRODUCT SURVEILLANCE TECHNICIAN (PST) REPORTED THAT DURING TESTING OF THE DEVICE AT THE SERVICE CENTER, THE BLOOD PARAMETER MONITOR (BPM) POWER SWITCH CONNECTOR SHORTED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16584467 ↗
  • MalfunctionNov 17, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE BLOOD PARAMETER MONITOR (BPM) MEASURED VALUES WERE VARYING. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT. PER THE CLINICAL REVIEW: ON (B)(6) 2022, THE TEAM EXPERIENCED A PROBLEM WITH THE BPM DURING CPB WHEREBY THE TEMPERATURE MEASURE VALUE HAD VARIATION. THERE WAS NO BLOOD LOSS, NO DELAY, AND NO CHANGE OUT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15813769 ↗
  • MalfunctionNov 4, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) HAD LARGE DISCREPANCIES IN THE HEMOGLOBIN (HB) AND GASES FROM THE HEMATOCRIT (HCT) PROBE. AS MITIGATION, BLOOD GAS ANALYSIS WAS USED TO MONITOR THE PATIENT'S LEVELS. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT. PER CLINICAL REVIEW: THE USER REPORTED THE BPM DURING CPB HAD LARGE DISCREPANCIES IN HB AND VENOUS OXYGEN SATURATION (SVO2) VALUES WHEN COMPARED TO THEIR MEASURED VALUES FROM THE LAB BLOOD GAS ANALYZER. THEY REPORTED DURING USE THE HB VALUE WAS OFF BY 3G, AND AFTER IN VIVO CALIBRATION THE HB VALUE ON THE SCREEN DID NOT CHANGE. THE TEMPERATURE WAS READING HIGH WHEN THE PATIENT'S TEMPERATURE WAS 32 DEGREES. THE SATURATIONS THEN READ 50-60% WHEN THE VENOUS BLOOD GAS SAMPLES SHOWED MIXED VENOUS BLOOD WAS 75-80%. THESE INACCURACIES RESULTED IN, PER THE USER, TRIPLE THE NUMBER OF BLOOD GASES NEEDED TO VERIFY AGAINST THE BPM VALUES. ALTHOUGH, THERE WAS NO DELAY TO THE PROCEDURE, NO INJURY, BLOOD LOSS, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THE USER STATED THIS UNIT WAS PREVIOUSLY SENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15734635 ↗
  • MalfunctionAug 10, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THERE WAS AN INCONSISTENCY IN THE TEMPERATURE READINGS FROM THE BLOOD PARAMETER MONITOR (BPM). THERE WERE NO REPORTED ADVERSE CONSEQUENCES TO THE PATIENT. NO OTHER DETAILS REGARDING THE NATURE OF THIS EVENT WERE PROVIDED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15202993 ↗
  • MalfunctionAug 10, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE BLOOD PARAMETER MONITOR (BPM) HAD A MALFUNCTION. AS MITIGATION, BLOOD GAS ANALYSIS WAS USED TO MONITOR THE LEVELS OF THE PATIENT. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15199395 ↗
  • MalfunctionJun 2, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE SERVICE REPAIR TECHNICIAN (SRT) REPORTED THAT DURING TESTING OF THE DEVICE AT THE SERVICE CENTER, THE BLOOD PARAMETER MONITOR (BPM) FAILED MEASURED TEMPERATURE ACCURACY TESTING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14580613 ↗
  • MalfunctionJun 2, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING SETUP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THERE WERE ISSUES WITH THE ACCURACY OF THE HEMATOCRIT (HCT) PARAMETER PRIOR TO IN VIVO CALIBRATION. AN IN VIVO CALIBRATION WAS DONE AND THE DEVICE WAS CONTINUED TO BE USED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14577724 ↗

Other Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass devices

All Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page