CDI OneView™ Monitoring System CDI754Recalls & reported problems

Terumo Cardiovascular Systems Corporation · CDI OneView™ H/S Probe · Class 2

All models of CDI OneView Monitoring System (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00886799001882), received between Jan 10, 2025 and Jan 10, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K234065); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 3, 2025Class IIOpen, Classified

    During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.

    Z-1492-2025 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 10, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    PER CLINICAL REVIEW: ON (B)(6) 2024 THE TEAM AT THE USER FACILITY EXPERIENCED A PROBLEM WITH THE BLOOD PARAMETER MONITOR (BPM) WHILE ON CARDIOPULMONARY BYPASS, WHEREBY THERE WERE INTERMITTENT VENOUS AND SATURATION VALUES 100 VENOUS SAT POST FIRST IN-VIVO AT 83. THIS COMPLAINT IS RELATED TO AN ISSUE WHERE THE THRESHOLD VALUE IN THE SIGNAL FILTER IMPLEMENTED IN THE HEMATOCRIT SATURATION (H/S) SOFTWARE WAS SET BELOW THE LEVEL OF ELECTRONIC NOISE, RESULTING IN A THRESHOLD TOO RESTRICTIVE FOR NORMAL SIGNAL VARIATION IN A CLINICAL SETTING. THIS WAS CAUSING THE H/S SATURATION OF OXYGEN (SO2) VALUE TO APPEAR AS DASHES INTERMITTENTLY. THE DEVICE WAS NOT CHANGED OUT. THERE WAS NO DELAY, NO BLOOD LOSS, AND THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21124113 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K234065. One submission often covers many models, so check the product and lot details in each record.

  • Z-2018-2025 · May 28, 2025 · Open, Classified
    CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit

Other Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass devices

All Monitor, Blood-Gas, On-Line, Cardiopulmonary Bypass devices →