The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00886874115954), received between Jan 21, 2022 and May 20, 2025 — classified as 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Activation Problem1
- Defective Device1
- Failure to Obtain Sample1
- Material Twisted/Bent1
- Positioning Problem1
- Retraction Problem1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information2
- Foreign Body In Patient1
- Pneumothorax1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2025Unintended System Motion; Material Twisted/Bent; Unintended System Motion; Material Twisted/Bent
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, A BIOPSY NEEDLE HAD A FRICTION ISSUE. THE PROCEDURE WAS COMPLETED. NO REPORTED KNOWN IMPACT OR PATIENT CONSEQUENCE WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22057592 ↗MalfunctionMay 12, 2025Positioning Problem; Positioning Problem
IT WAS REPORTED THAT DURING AN ION ENDOBRONCHIAL LUNG BIOPSY PROCEDURE, THE CUSTOMER STATED THAT THE BIOPSY NEEDLE INSTRUMENT WAS NOTED TO HAVE FRICTION ISSUE. THE DIAGNOSTIC PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22007205 ↗InjuryOct 14, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ION ENDOBRONCHIAL LUNG BIOPSY PROCEDURE, THE PHYSICIAN DID NOT HAVE ACCESS TO A LINEAR ENDO-BRONCHIAL ULTRASOUND (EBUS) SCOPE, SO USED THE ION FOR A MEDIASTINAL STAGING OF A 4R NODE (RIGHT LOWER PARATRACHEAL LYMPH NODE). AFTER THE BIOPSY, IT WAS SEEN WITH THE CAMERA THAT PART OF THE NEEDLE BROKE OFF AND WAS LODGED INTO THE ENDOTRACHEAL (ET) TUBE. AFTER EXTUBATING, THE PART OF THE NEEDLE WAS STILL IN THE TUBE. THERE WAS NO DELAY. THE PHYSICIAN WAS WELL AWARE THAT ION WAS NOT FOR STAGING. THE DIAGNOSTIC PROCEDURE WAS COMPLETED. INTUITIVE SURGICAL, INC. (ISI) OBTAINED THE FOLLOWING ADDITIONAL INFORMATION FROM THE MANAGER OF CLINICAL APPLICATIONS ENGINEERING: THE MEDICAL PERSONNEL STATED THAT THEY THREW THE ET TUBE WITH THE FRAGMENT AWAY AND DID AN AIRWAY SURVEY AFTER AND DID NOT SEE ANY OTHER FRAGMENTS IN THE AIRWAY. THE PATIENT WAS DOING WELL IN RECOVERY.
Patient problems: Foreign Body In Patient
MAUDE report 20447154 ↗MalfunctionNov 14, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING AN ION ENDOBRONCHIAL LUNG BIOPSY PROCEDURE, THE NEEDLE BIOPSY BROKE. ALL PIECES ARE THERE. THE DIAGNOSTIC PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18128343 ↗MalfunctionMay 26, 2023Unintended Movement
NEEDLE SHEATH THROUGH SIDE WALL, SHEATH MOVEMENT.
Patient problems: Insufficient Information
MAUDE report 17009780 ↗MalfunctionMay 1, 2023Failure to Obtain Sample; Activation Problem
THE NEEDLE FOR NEEDLE ASPIRATION BIOPSIES WOULD NOT EXTEND FULLY FROM THE SHEATH AND WERE UNABLE TO BE USED TO RETRIEVE BIOPSIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16843303 ↗MalfunctionOct 28, 2022Retraction Problem
DURING BRONCHOSCOPY WHILE ATTEMPTING TO OBTAIN A NEEDLE ASPIRATION BIOPSY, THE DEVICE MALFUNCTIONED. AFTER THE THIRD PASS, THE NEEDLE BECAME UNABLE TO RETRACT CORRECTLY. THE DEVICE WAS TAKEN OUT OF SERVICE AT THAT TIME AND REPLACED WITH A NEW NEEDLE IN ORDER TO COMPLETE THE PROCESS.
Patient problems: Insufficient Information
MAUDE report 15690247 ↗InjuryJan 21, 2022Defective Device; Defective Device
IT WAS REPORTED THAT DURING AN ION ENDOLUMINAL LUNG BIOPSY PROCEDURE THE 23G FLEXISION BIOPSY NEEDLE SHEATH MOVED SIGNIFICANTLY DURING THE FIRST PASS MADE WITH IT. ON (B)(6) 2022, INTUITIVE SURGICAL INC. (ISI) CONTACTED THE PHYSICIAN OF THIS PROCEDURE AND ADDITIONAL INFORMATION WAS OBTAINED ABOUT THE COMPLAINT: THE PNEUMOTHORAX WAS IDENTIFIED SONOGRAPHICALLY (ULTRASOUND) AFTER THE ION PROCEDURE WAS COMPLETED. THE PHYSICIAN REPORTED THAT THEY BELIEVE THE 23G FLEXISION BIOPSY NEEDLE CONTRIBUTED TO THE PNEUMOTHORAX, BUT THAT THE BIOPSY NEEDLE WAS DISCARDED AND THEREFORE CANNOT BE RETURNED. THE PHYSICIAN CONFIRMED THAT 23G FLEXISION BIOPSY NEEDLE SHEATH MOVEMENT OCCURRED DURING THIS PROCEDURE AND THEY SAID THE MOVEMENT MAY HAVE POSSIBLY CAUSED OR CONTRIBUTED TO THE PNEUMOTHORAX. WHEN ASKED IF THE PHYSICIAN BELIEVES THIS PNEUMOTHORAX WOULD HAVE LIKELY OCCURRED VIA ANOTHER BIOPSY MODALITY, THE PHYSICIAN ANSWERED: ¿CAN¿T ANSWER THIS.¿ THE PHYSICIAN REPORTED THAT THE PNEUMOTHORAX WAS MODERATE IN SIZE, AND WAS PROBABLY A ONETIME LEAK THAT, ONCE IT WAS DRAINED COMPLETELY, DID NOT RECOLLECT. THE PATIENT EXPERIENCED CHEST PAIN AND HYPOXIC RESPIRATORY FAILURE. WHEN ASKED IF THE PATIENT WAS EVER
Patient problems: Pneumothorax
MAUDE report 13322743 ↗
Other Bronchoscope (flexible or rigid) devices
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-XP190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1,146 reports
- Ion 380748Intuitive Surgical, Inc. · Bronchoscope (flexible or rigid)1 recall97 deaths1,278 reports
- VISERA ELITE XENON LIGHT SOURCE OLYMPUS CLV-S190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)1,155 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)5 recalls3 deaths621 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
- BRONCHOVIDEOSCOPE BF-1T150Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls212 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page