Skip to content

PRECICE® Plating System LP4.5-PRF4BRecalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · PRECICE Plate 4.5mm Proximal Femur Plate, Right, 45mm, Bump · Class 2 · Implant

All models of PRECICE Plating System (67)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
3

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00887517021366), received between Jun 22, 2021 and Nov 30, 2021 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Corroded” (6 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Oct 29, 2021Class IIOpen, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Class IIOpen, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1325-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Feb 12, 2021Class IIOpen, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Corroded6
  • Material Discolored1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Deformity/ Disfigurement1
  • Impaired Healing1
  • Insufficient Information1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 30, 2021Corroded; Corroded

    DURING ROUTINE REMOVAL OF THE PLATE AFTER THE LENGTHENING GOAL HAD BEEN ACHIEVED, CORROSION WAS DISCOVERED ON THE PLATE USED ON THE LEFT TIBIA. THE CORROSION WAS NOTED AROUND PLATE'S MALE/FEMALE JUNCTION AND SELECT SCREW SITES. AT THE TIME OF THE REMOVAL, IT WAS ALSO NOTED THAT THERE WAS A DELAYED UNION OF THE REGENERATE IN THE LEFT TIBIA; HOWEVER, NO OSTEOLYSIS WAS FOUND. A LOW PROFILE STATIC LOCKING PLATE WAS IMPLANTED TO PROVIDE EXTENDED SURGICAL SITE STABILITY AND PRESERVE NEW BONE LENGTH. NO PATIENT HARM OR ADVERSE IMPACT IS ASSOCIATED WITH THIS EVENT.

    Patient problems: Impaired Healing; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12899367 ↗
  • MalfunctionNov 30, 2021Corroded; Corroded

    DURING ROUTINE REMOVAL OF THE PLATE AFTER THE LENGTHENING GOAL HAD BEEN ACHIEVED, CORROSION WAS DISCOVERED ON THE PLATE USED ON THE RIGHT TIBIA. THE CORROSION WAS NOTED AROUND PLATE'S MALE/FEMALE JUNCTION AND SELECT SCREW SITES. NO PATIENT HARM OR ADVERSE OUTCOME IS ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12899292 ↗
  • MalfunctionSep 20, 2021Corroded; Material Twisted/Bent; Corroded; Material Twisted/Bent

    IT WAS REPORTED THAT UPON PLANNED REMOVAL OF THE PRECICE PLATE AFTER LENGTHENING GOALS WERE SUCCESSFULLY ACHIEVED AND THERE WAS RADIOGRAPHIC EVIDENCE OF QUALITY REGENERATE BONE/CONSOLIDATION, IT WAS NOTED THE PLATE HAD VISIBLE SIGNS OF CORROSION AROUND THE SCREW HOLES IN THE DISTRACTION ROD SIDE OF THE PLATE. NO ADVERSE PATIENT IMPACT WAS NOTED AND THE SURGEON WILL CONTINUE WITH THE PLANNED POST-OP FOLLOW UP VISITS/PATIENT MONITORING PROTOCOL. IT WAS ALSO REPORTED THAT THE PATIENT HAD BEEN PARTIALLY WEIGHT-BEARING IN THE LATE CONSOLIDATION PHASE OF TREATMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12497601 ↗
  • InjuryJul 27, 2021Corroded; Corroded

    NO ADDITIONAL INFORMATION WAS RECEIVED.

    Patient problems: Deformity/ Disfigurement

    MAUDE report 12233285 ↗
  • MalfunctionJul 7, 2021Corroded; Corroded

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12132628 ↗
  • MalfunctionJun 22, 2021Corroded; Material Discolored; Corroded; Material Discolored

    INFORMATION WAS RECEIVED THAT DURING IMPLANT REMOVAL PROCEDURE, IT WAS DISCOVERED THAT THE PLATE HAD VISIBLE DISCOLORATION AND CORROSION. AT THE TIME OF REMOVAL, A LOW PROFILE NAIL WAS INSERTED FOR EXTRA STABILITY. THE DESIRED LENGTH WAS ACHIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12044921 ↗

FDA clearance and classification

Cleared through 510(k) K192181 on Nov 7, 2019 (decision: Substantially Equivalent). Submitted by Nuvasive Specialized Orthopedics, Inc. as “PRECICE Plating System”. The FDA review took 87 days. FDA record ↗

The FDA regulates product code HRS (plate, fixation, bone) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the PRECICE® Plating System LP4.5-PRF4B

Has the PRECICE® Plating System LP4.5-PRF4B been recalled?

Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-0319-2022 (Class II), began on Oct 29, 2021: The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the PRECICE® Plating System LP4.5-PRF4B?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jun 22, 2021 and Nov 30, 2021. In those reports, the most common specific device problems are “Corroded” (6), “Material Discolored” (1) and “Material Twisted/Bent” (1); the most common patient problems are “Deformity/ Disfigurement” (1) and “Impaired Healing” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PRECICE® Plating System LP4.5-PRF4B?

None of the reports are classified as deaths, 1 as an injury and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the PRECICE® Plating System LP4.5-PRF4B FDA approved?

It was cleared through 510(k) K192181 on Nov 7, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the PRECICE® Plating System LP4.5-PRF4B MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the PRECICE® Plating System LP4.5-PRF4B, and is it still sold?

The FDA's device database lists Nuvasive Specialized Orthopedics, Inc. as the company responsible for it (brand PRECICE® Plating System, model LP4.5-PRF4B). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Plate, Fixation, Bone devices

All Plate, Fixation, Bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page