PRECICE® Plating System LP4.5-MLT4BRecalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · PRECICE Plate 4.5mm Posteromedial Proximal Tibia Plate, Left, 45mm, Bump · Class 2 · Implant
3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00887517021687), received between Feb 22, 2021 and Mar 30, 2021 — classified as 5 injuries.
3 FDA recall records name this device's identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Oct 29, 2021Open, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Z-0319-2022 · Root cause: Device Design · FDA record ↗
- Feb 12, 2021Open, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1325-2021 · Root cause: Under Investigation by firm · FDA record ↗
- Feb 12, 2021Open, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1323-2021 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available3
- Deformity/ Disfigurement1
- Unspecified Infection1
- Wound Dehiscence1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMar 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION WAS RECEIVED THAT A SUPERFICIAL WOUND INFECTION WAS DIAGNOSED. THE INFECTION IS BELIEVED TO BE THE RESULT OF PROMINENT HARDWARE IN AN AREA OF LIMITED SOFT TISSUE COVERAGE. ADDITIONALLY, THE TENSION OF THE SOFT TISSUE OVER THE PLATE INCREASED AND CONTRIBUTED TO WOUND BREAKDOWN.
Patient problems: Deformity/ Disfigurement; Unspecified Infection
MAUDE report 11589054 ↗InjuryFeb 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED THAT REMOVAL PROCEDURE WAS PERFORMED (B)(6) 2020 DUE TO WOUND BREAKDOWN. AS PER THE REPORTER, THE WOUND WAS IRRIGATED AND DEBRIDED. THE PATIENT WAS SWITCHED TO EXTERNAL FIXATION TO COMPLETE THEIR LENGTHENING TREATMENT.
Patient problems: Wound Dehiscence
MAUDE report 11363723 ↗InjuryFeb 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED THAT AN OSTEOTOMY REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020 FOR PREMATURE CONSOLIDATION ENCOUNTERED 2 CM INTO LENGTHENING. AS PER THE SURGEON, THE EVENT WAS THE RESULT OF SURGEON ERROR, SPECIFICALLY ASSOCIATED WITH INADEQUATE RADIOGRAPHIC ASSESSMENT AND RATE OF CORRECTION AT THE TIME OF FOLLOW UP.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11363720 ↗InjuryFeb 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020. AS PER THE REPORTER, THE PATIENT NEEDED AN OSTEOTOMY IN ORDER TO CONTINUE THE LENGTHENING TREATMENT AFTER RECOVERING FROM COMPARTMENT SYNDROME.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11363718 ↗InjuryFeb 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED THAT A FOLLOW UP NERVE RELEASE/ FASCIOTOMY PROCEDURE WAS PERFORMED ON (B)(6) 2020. AS PER THE REPORTER, THE PATIENT DEVELOPED COMPARTMENT SYNDROME.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11363716 ↗
Other Plate, Fixation, Bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-DRF32Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.