Skip to content

ReLine 15360500Recalls & reported problems

Nuvasive, Inc. · RELINE-O Ti Rod, 6.0x500mm Straight · Class 2 · Implant

All models of ReLine (129)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00887517510822), received between May 27, 2022 and Feb 16, 2024 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 16, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL EXTREME LATERAL INTERBODY FUSION PROCEDURE AT UNKNOWN LEVEL(S). SEVEN (7) DAYS LATER, THE PATIENT UNDERWENT A SECONDARY PROCEDURE TO ADD POSTERIOR FIXATION FROM T8 TO SAI. APPROXIMATELY FOUR (4) YEARS LATER, DURING A ROUTINE FOLLOW-UP APPOINTMENT, IT WAS DISCOVERED THAT THE FIXATION RODS ON BOTH SIDES OF THE CONSTRUCT AND ONE ROD TO ROD CONNECTOR WERE FRACTURED NEAR THE L3 VERTEBRAL BODY. THE PATIENT WAS NOTED TO BE ASYMPTOMATIC. SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION PROCEDURE TO REPLACE THE FRACTURED IMPLANTS AND REINFORCE THE FIXATION CONSTRUCT WITH BILATERAL DUAL RODS. NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. REPORT 2 OF 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18723130 ↗
  • InjuryFeb 16, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL EXTREME LATERAL INTERBODY FUSION PROCEDURE AT UNKNOWN LEVEL(S). SEVEN (7) DAYS LATER, THE PATIENT UNDERWENT A SECONDARY PROCEDURE TO ADD POSTERIOR FIXATION FROM T8 TO SAI. APPROXIMATELY FOUR (4) YEARS LATER, DURING A ROUTINE FOLLOW-UP APPOINTMENT, IT WAS DISCOVERED THAT THE FIXATION RODS ON BOTH SIDES OF THE CONSTRUCT AND ONE ROD TO ROD CONNECTOR WERE FRACTURED NEAR THE L3 VERTEBRAL BODY. THE PATIENT WAS NOTED TO BE ASYMPTOMATIC. SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION PROCEDURE TO REPLACE THE FRACTURED IMPLANTS AND REINFORCE THE FIXATION CONSTRUCT WITH BILATERAL DUAL RODS. NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. REPORT 1 OF 3.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18723096 ↗
  • InjuryNov 20, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TEN (10) LEVEL POSTERIOR FIXATION PROCEDURE FROM T9 TO S2AI WITH INTERBODY FUSION AT L4/5 AND L5/S1. SUBSEQUENTLY, ON AN UNKNOWN DATE APPROXIMATELY 29 MONTHS POST-OP, THE LEFT FIXATION ROD FRACTURED NEAR L5/S1. A REVISION PROCEDURE WAS PERFORMED TO PLACE A ROD-TO-ROD CONNECTOR AND ADJACENT SEGMENT FIXATION TO REINFORCE THE FRACTURED ROD. NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18173818 ↗
  • InjuryNov 20, 2023Fracture; Fracture

    IN (B)(6) 2020 A PATIENT UNDERWENT A POSTERIOR FIXATION PROCEDURE FROM T9- S2 WITH VERTEBRAL SPACERS BEING PLACED AT L4-S. THE SURGERY WAS COMPLETED WITHOUT ISSUE. ON A UNKNOWN DATE IT WAS DISCOVERED THE PATIENT EXPERIENCED A ROD FRACTURE ON THE LEFT SIDE OF THE CONSTRUCT IN-BETWEEN THE L5-S LEVEL. IN (B)(6) 2022 A REVISION OCCURRED WHERE A CONNECTOR WAS PLACED TO REINFORCE THE FRACTURED ROD. THE REVISION WAS COMPLETED WITHOUT ISSUE. (NUVASIVE COMPLAINT (B)(4) ) ON A UNKNOWN DATE IT WAS DISCOVERED THE PATIENT EXPERIENCED ANOTHER ROD FRACTURE THIS TIME ON THE RIGHT SIDE OF THE CONSTRUCT IN-BETWEEN THE L4-L5 LEVEL. ON (B)(6) 2023 AN ADDITIONAL REVISION OCCURRED WHERE A CONNECTOR WAS PLACED TO REINFORCE THE FRACTURED ROD. THE REVISION WAS COMPLETED WITHOUT ISSUE AND THE PATIENT IS REPORTED TO BE DOING WELL POST REVISION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18173607 ↗
  • MalfunctionMay 27, 2022Fracture; Fracture

    ON A UNKNOWN DATE A PATIENT UNDERWENT A VERTEBRAL INTERBODY REPLACEMENT PROCEDURE FROM T9 TO L2 USING A NUVASIVE VERTEBRAL BODY SPACER AND COMPETITORS PRODUCT FOR THE REMAINING PORTION OF THE CONSTRUCT. ON (B)(6) A REVISION OCCURRED FOR A UNKNOWN REASON TO EXTEND THE CONSTRUCT FROM L4 TO S1 USING NUVASIVE PRODUCTS. ON (B)(6) 2022 AN ADDITIONAL REVISION OCCURRED AS IT WAS DISCOVERED A IMPLANTED ROD FRACTURED AT THE HEAD OF THE RIGHT S1 SCREW. DURING THE REVISION PROCEDURE IT WAS OBSERVED THAT THE SCREWS ON BOTH SIDES OF S1AI AND S2AI WERE FOUND TO HAVE LOOSENING. AS A CONSEQUENCE DURING THE REVISION THEY WERE ALL REPLACED WITH THE SCREWS OF INCREASED SIZE. IT IS UNKNOWN WHETHER THE PATIENT WAS EXPERIENCING ANY ADVERSE CONSEQUENCES AS A RESULT OF THE ROD FRACTURE THE PATIENT IS DOING WELL POST REVISION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14524581 ↗

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →