Skip to content

ReLine 16018580Recalls & reported problems

Nuvasive, Inc. · RELINE MAS Screw, 8.5x80mm 2C Poly Iliac · Class 2 · Implant

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887517554376), received between May 19, 2023 and Aug 20, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2024Fracture; Fracture

    ON (B)(6) 2024 A PATIENT UNDERWENT A POSTERIOR FIXATION PROCEDURE FROM T7-S2AI. THE PROCEDURE WAS COMPLETED WITHOUT ISSUE. ON (B)(6) 2024 DURING A ROUTINE FOLLOW-UP APPOINTMENT IT WAS DISCOVERED VIA RADIOGRAPH THAT A SCREW AT RIGHT S2AI HAD FRACTURED AT THE NECK OF THE SCREW HEAD. IT IS UNKNOWN IF THE PATIENT WAS EXPERIENCING ANY ADVERSE CONSEQUENCES AS A RESULT OF THE DEVICE FAILURE. ON (B)(6) 2024 A REVISION OCCURRED WHERE THE BROKEN SCREW WAS REMOVED AND REPLACED WITH A 10.5X80MM SCREW. THE PATIENTS CURRENT STATUS IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20000638 ↗
  • MalfunctionMay 19, 2023Fracture; Fracture

    ON (B)(6) 2022 A PATIENT UNDERWENT A TRANSFORAMINAL LUMBAR INTERBODY FUSION PROCEDURE FROM T10 TO S2. THE SURGERY WAS COMPLETED WITHOUT ISSUE. ON (B)(6) 2023 IT WAS DISCOVERED VIA RADIOGRAPH DURING A ROUTINE FOLLOW UP THAT THE SHAFT OF THE LEFT S2AI SCREW FRACTURED WHERE THE HEAD MEATS THE TULIP. THE PATIENT IS REPORTED TO BE ASYMPTOMATIC AND NO FURTHER ACTION WILL BE TAKEN AT THIS TIME. THE PATIENT WILL CONTINUE TO BE MONITORED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16963678 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices