ReLine 16019580Recalls & reported problems
Nuvasive, Inc. · RELINE MAS Screw, 9.5x80mm 2C Polyaxial · Class 2 · Implant
All models of ReLine (129)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887517554734), received between Jun 3, 2021 and Nov 1, 2022 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture4
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Pain2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 1, 2022Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL POSTERIOR FIXATION PROCEDURE. SUBSEQUENTLY, THE PATIENT REPORTED SYMPTOMS OF PAIN APPROXIMATELY FIVE (5) MONTHS LATER AND EXAMINATION IDENTIFIED THAT BOTH PEDICLE SCREWS AT S2 HAD FRACTURED. A REVISION PROCEDURE WAS PERFORMED APPROXIMATELY SIX (6) MONTHS AFTER THE INITIAL PROCEDURE TO REMOVE AND REPLACE THE FRACTURED SCREWS. DURING THE REVISION PROCEDURE, IT WAS IDENTIFIED THAT BOTH SCREWS AT S1 HAD LOOSENED AND FUSION HAD FAILED TO OCCUR AT THE L5/S1 DISC SPACE. THE SCREWS AT S1 WERE REMOVED AND REPLACED. NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS AVAILABLE. REPORT 2 OF 4.
Patient problems: Pain
MAUDE report 15712726 ↗InjuryNov 1, 2022Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL POSTERIOR FIXATION PROCEDURE. SUBSEQUENTLY, THE PATIENT REPORTED SYMPTOMS OF PAIN APPROXIMATELY FIVE (5) MONTHS LATER AND EXAMINATION IDENTIFIED THAT BOTH PEDICLE SCREWS AT S2 HAD FRACTURED. A REVISION PROCEDURE WAS PERFORMED APPROXIMATELY SIX (6) MONTHS AFTER THE INITIAL PROCEDURE TO REMOVE AND REPLACE THE FRACTURED SCREWS. DURING THE REVISION PROCEDURE, IT WAS IDENTIFIED THAT BOTH SCREWS AT S1 HAD LOOSENED AND FUSION HAD FAILED TO OCCUR AT THE L5/S1 DISC SPACE. THE SCREWS AT S1 WERE REMOVED AND REPLACED. NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS AVAILABLE. REPORT 1 OF 4.
Patient problems: Pain
MAUDE report 15712725 ↗MalfunctionSep 28, 2021Fracture; Fracture
ON (B)(6) 2021 A PATIENT'S KYPHOSIS INCREASED AND THEIR CONSTRUCT WAS EXTENDED FROM T10 TO T4 AND A FRACTURED SCREW AT SAI WAS REPLACED. ON (B)(6) 2021 IT WAS REPORTED THAT THE SAI SCREW FRACTURED. NO REVISION SURGERY IS PLANNED AND NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12541392 ↗MalfunctionJun 3, 2021Fracture; Fracture
ON (B)(6) 2020 A POSTERIOR LUMBAR INTERBODY FUSION AT T10/S2AI. ON (B)(6) 2021 A REVISION SURGERY OCCURRED DUE TO A ROD FRACTURE AT LEFT L5/S1 AND A SCREW FRACTURE AT RIGHT S2AI. (2031966-2021-00051) ON (B)(6) 2021 DURING A ROUTINE FOLLOW-UP AN X-RAY REVEALED THAT THE RIGHT S2AI SCREW FRACTURED. BONE FUSION WAS REPORTEDLY COMPLETE AND THE PATIENT'S PHYSICAL ACTIVITY LEVEL WAS NORMAL. THE PHYSICIAN STATED THEY WILL CONTINUE TO MONITOR THE PATIENT. NO REVISION SURGERY PLANNED AND NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11930036 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.