ReLine 16024530Recalls & reported problems
Nuvasive, Inc. · RELINE MAS Red Screw, 4.5x30mm Poly 2C · Class 2 · Implant
All models of ReLine (129)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887517555885), received between Sep 2, 2022 and Sep 2, 2022 — classified as 3 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth3
Patient problems
- Osteomyelitis3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 2, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
(B)(6) 2021, TH3/4/5/6/10/11/12 WERE FIXED WITH RELINE DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2021 A PATIENT UNDERWENT A SPINAL PROCEDURE AT T3/6 AND T10/12 DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2022 A REVISION SURGERY WAS CONDUCTED DUE TO RECURRENCE OF INFECTION AND LOOSENING OF THE PEDICLE SCREWS ON THE LEFT SIDE. THE RODS WERE CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO THE LEFT TH2, AND THE PEDICLE SCREWS OF TH4, 5, AND 6 WERE INCREASED IN SIZE. A 2-HOLE INLINE CONNECTOR WAS INSTALLED IN TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE ROD ON THE RIGHT SIDE WAS CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO TH2 AND THE PEDICLE SCREW OF TH4 WAS INCREASED IN SIZE. THE PEDICLE SCREWS OF TH5 AND TH6 WERE NOT REMOVED. A 2-HOLE INLINE CONNECTOR WAS PLACED ON THE CAUDAL SIDE OF TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE SCREW HEAD OF THE 2-HOLE INLINE CONNECTOR ON THE RIGHT SIDE WAS NOT STRIPPED. THE PATIENT'S BONE QUALITY WAS IN VERY POOR CONDITION.
Patient problems: Osteomyelitis
MAUDE report 15351872 ↗InjurySep 2, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
ON (B)(6) 2021, TH3/4/5/6/10/11/12 WERE FIXED WITH RELINE DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2021 A PATIENT UNDERWENT A SPINAL PROCEDURE AT T3/6 AND T10/12 DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2022 A REVISION SURGERY WAS CONDUCTED DUE TO RECURRENCE OF INFECTION AND LOOSENING OF THE PEDICLE SCREWS ON THE LEFT SIDE. THE RODS WERE CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO THE LEFT TH2, AND THE PEDICLE SCREWS OF TH4, 5, AND 6 WERE INCREASED IN SIZE. A 2-HOLE INLINE CONNECTOR WAS INSTALLED IN TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE ROD ON THE RIGHT SIDE WAS CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO TH2 AND THE PEDICLE SCREW OF TH4 WAS INCREASED IN SIZE. THE PEDICLE SCREWS OF TH5 AND TH6 WERE NOT REMOVED. A 2-HOLE INLINE CONNECTOR WAS PLACED ON THE CAUDAL SIDE OF TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE SCREW HEAD OF THE 2-HOLE INLINE CONNECTOR ON THE RIGHT SIDE WAS NOT STRIPPED. THE PATIENT'S BONE QUALITY WAS IN VERY POOR CONDITION.
Patient problems: Osteomyelitis
MAUDE report 15351852 ↗InjurySep 2, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
ON (B)(6) 2021, TH3/4/5/6/10/11/12 WERE FIXED WITH RELINE DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2021 A PATIENT UNDERWENT A SPINAL PROCEDURE AT T3/6 AND T10/12 DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2022 A REVISION SURGERY WAS CONDUCTED DUE TO RECURRENCE OF INFECTION AND LOOSENING OF THE PEDICLE SCREWS ON THE LEFT SIDE. THE RODS WERE CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO THE LEFT TH2, AND THE PEDICLE SCREWS OF TH4, 5, AND 6 WERE INCREASED IN SIZE. A 2-HOLE INLINE CONNECTOR WAS INSTALLED IN TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE ROD ON THE RIGHT SIDE WAS CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO TH2 AND THE PEDICLE SCREW OF TH4 WAS INCREASED IN SIZE. THE PEDICLE SCREWS OF TH5 AND TH6 WERE NOT REMOVED. A 2-HOLE INLINE CONNECTOR WAS PLACED ON THE CAUDAL SIDE OF TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE SCREW HEAD OF THE 2-HOLE INLINE CONNECTOR ON THE RIGHT SIDE WAS NOT STRIPPED. THE PATIENT'S BONE QUALITY WAS IN VERY POOR CONDITION.
Patient problems: Osteomyelitis
MAUDE report 15351809 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page