Skip to content

ReLine 16024530Recalls & reported problems

Nuvasive, Inc. · RELINE MAS Red Screw, 4.5x30mm Poly 2C · Class 2 · Implant

All models of ReLine (129)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887517555885), received between Sep 2, 2022 and Sep 2, 2022 — classified as 3 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth3

Patient problems

  • Osteomyelitis3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjurySep 2, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    (B)(6) 2021, TH3/4/5/6/10/11/12 WERE FIXED WITH RELINE DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2021 A PATIENT UNDERWENT A SPINAL PROCEDURE AT T3/6 AND T10/12 DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2022 A REVISION SURGERY WAS CONDUCTED DUE TO RECURRENCE OF INFECTION AND LOOSENING OF THE PEDICLE SCREWS ON THE LEFT SIDE. THE RODS WERE CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO THE LEFT TH2, AND THE PEDICLE SCREWS OF TH4, 5, AND 6 WERE INCREASED IN SIZE. A 2-HOLE INLINE CONNECTOR WAS INSTALLED IN TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE ROD ON THE RIGHT SIDE WAS CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO TH2 AND THE PEDICLE SCREW OF TH4 WAS INCREASED IN SIZE. THE PEDICLE SCREWS OF TH5 AND TH6 WERE NOT REMOVED. A 2-HOLE INLINE CONNECTOR WAS PLACED ON THE CAUDAL SIDE OF TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE SCREW HEAD OF THE 2-HOLE INLINE CONNECTOR ON THE RIGHT SIDE WAS NOT STRIPPED. THE PATIENT'S BONE QUALITY WAS IN VERY POOR CONDITION.

    Patient problems: Osteomyelitis

    MAUDE report 15351872 ↗
  • InjurySep 2, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    ON (B)(6) 2021, TH3/4/5/6/10/11/12 WERE FIXED WITH RELINE DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2021 A PATIENT UNDERWENT A SPINAL PROCEDURE AT T3/6 AND T10/12 DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2022 A REVISION SURGERY WAS CONDUCTED DUE TO RECURRENCE OF INFECTION AND LOOSENING OF THE PEDICLE SCREWS ON THE LEFT SIDE. THE RODS WERE CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO THE LEFT TH2, AND THE PEDICLE SCREWS OF TH4, 5, AND 6 WERE INCREASED IN SIZE. A 2-HOLE INLINE CONNECTOR WAS INSTALLED IN TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE ROD ON THE RIGHT SIDE WAS CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO TH2 AND THE PEDICLE SCREW OF TH4 WAS INCREASED IN SIZE. THE PEDICLE SCREWS OF TH5 AND TH6 WERE NOT REMOVED. A 2-HOLE INLINE CONNECTOR WAS PLACED ON THE CAUDAL SIDE OF TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE SCREW HEAD OF THE 2-HOLE INLINE CONNECTOR ON THE RIGHT SIDE WAS NOT STRIPPED. THE PATIENT'S BONE QUALITY WAS IN VERY POOR CONDITION.

    Patient problems: Osteomyelitis

    MAUDE report 15351852 ↗
  • InjurySep 2, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    ON (B)(6) 2021, TH3/4/5/6/10/11/12 WERE FIXED WITH RELINE DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2021 A PATIENT UNDERWENT A SPINAL PROCEDURE AT T3/6 AND T10/12 DUE TO INFECTIOUS SPONDYLITIS. ON (B)(6) 2022 A REVISION SURGERY WAS CONDUCTED DUE TO RECURRENCE OF INFECTION AND LOOSENING OF THE PEDICLE SCREWS ON THE LEFT SIDE. THE RODS WERE CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO THE LEFT TH2, AND THE PEDICLE SCREWS OF TH4, 5, AND 6 WERE INCREASED IN SIZE. A 2-HOLE INLINE CONNECTOR WAS INSTALLED IN TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE ROD ON THE RIGHT SIDE WAS CUT ON THE CAUDAL SIDE OF TH6. A PEDICLE SCREW WAS ADDED TO TH2 AND THE PEDICLE SCREW OF TH4 WAS INCREASED IN SIZE. THE PEDICLE SCREWS OF TH5 AND TH6 WERE NOT REMOVED. A 2-HOLE INLINE CONNECTOR WAS PLACED ON THE CAUDAL SIDE OF TH6 AND CONNECTED TO THE ROD ON THE CAUDAL SIDE. THE SCREW HEAD OF THE 2-HOLE INLINE CONNECTOR ON THE RIGHT SIDE WAS NOT STRIPPED. THE PATIENT'S BONE QUALITY WAS IN VERY POOR CONDITION.

    Patient problems: Osteomyelitis

    MAUDE report 15351809 ↗

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page