Nuvasive, Inc. · RELINE-O CoCr Rod, 5.5x500mm Straight · Class 2 · Implant

All models of ReLine (129)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00887517594204), received between Dec 8, 2016 and Jan 7, 2026 — classified as 2 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Fracture” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture10
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Failure of Implant5
  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1
  • Reaction1
  • Seizures1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 7, 2026Fracture

    IT WAS REPORTED THAT ON (B)(6) 2019 THE SURGEON INSERTED COROENT XL PEEK CAGE INTO L1~L5 BY XLIF, AND ANOTHER BRAND SPINAL CAGE INTO L5/S1 BY PLIF. THEN ON (B)(6) 2019 THE SURGEON FIXED SPINE FROM TH10 TO S2 ALAR ILIAC. THE SPINAL ROD WAS FRACTURED AT THE FOLLOWING SITES, BUT WE COULD NOT OBTAIN INFORMATION WHEN THE ROD FRACTURED: TH12/L1; RIGHT SIDE AND LEFT SIDE L4/L5: RIGHT SIDE AND L5/L1 RIGHT SIDE. THEN ON (B)(6) 2025 THE SURGEON PERFORMED REVISION SURGERY. THEN ON (B)(6) 2025 WE OBTAINED INFORMATION THAT THE PATIENT HAD BEEN DISCHARGED FROM THE HOSPITAL. WE DID NOT OBTAIN ANY OTHER PHOTOGRAPHS AND REVISED IMPLANTS FROM THE HOSPITAL. AND WE DO NOT RECEIVE ANY OTHER ADVERSE PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23989554 ↗
  • MalfunctionMay 24, 2024Fracture; Fracture

    ON (B)(6) 2023 A PATIENT UNDERWENT A POSTERIOR FIXATION PROCEDURE FROM T8 TO S2AI. THE SURGERY WAS COMPLETED WITHOUT ISSUE. ON A UNKNOWN DATE, A ROD IN THE CONSTRUCT WAS REPORTED TO HAVE FRACTURED ON THE LEFT SIDE AT S1/SAI. ON (B)(6) 2024 A REVISION SURGERY WAS CONDUCTED TO REMOVE AND REPLACE THE FRACTURED ROD. ADDITIONALLY, AN ILIAC SCREW WAS ADDED ON THE LEFT, AND RODS WERE ADDED, ONE EACH LATERALLY AND MEDIALLY TO REINFORCE THE CONSTRUCT. THE REVISION SURGERY WAS COMPLETED WITHOUT ISSUE WITH NO ADVERSE PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19392220 ↗
  • InjuryJul 29, 2022Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL NINE (9) LEVEL FIXATION PROCEDURE WITH INTERBODY DEVICES PLACED AT L1 TO L5 AND POSTERIOR FIXATION FROM T9 TO ILIAC. SUBSEQUENTLY, RADIOGRAPHS OBTAINED APPROXIMATELY 17 MONTHS AFTER THE INITIAL PROCEDURE REVEALED FRACTURE OF THE RIGHT POSTERIOR FIXATION ROD AT THE L5/S1 SPACE. AN ADDITIONAL PROCEDURE WAS PERFORMED TWO (2) WEEKS LATER TO REINFORCE THE L5/S1 VERTEBRAE WITH ADDITIONAL ROD FIXATION. NO FURTHER PATIENT IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15136586 ↗
  • MalfunctionMay 11, 2021Fracture; Fracture

    ON (B)(6) 2018 A PATIENT UNDERWENT A SPINAL PROCEDURE. ON (B)(6) 2021 THE PATIENT REPORTED HEARING A SOUND WHEN BENDING DOWN. ON (B)(6) 2021 AN X-RAY REVEALED A ROD FRACTURE AT L4/5. ON (B)(6) 2021 A REVISION SURGERY WAS PERFORMED AND BONE FUSION WAS OBSERVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11808538 ↗
  • MalfunctionApr 9, 2021Fracture; Fracture

    A PATIENT UNDERWENT A SPINAL PROCEDURE TO ADDRESS KYPHOSIS FROM T10 TO S2A1 WITH A SPACER CAGE BEING PLACED BETWEEN L5/S. ON AN UNKNOWN DATE THE PATIENT HEARD A SOUND AND FOLLOWED UP ON (B)(6) 2021 WHERE RADIOGRAPHS CONFIRMED THAT THE LEFT ROD HAD FRACTURED AROUND L5/S1 AND A SCREW HEAD HAD FRACTURED AT S2A1. THE PATIENT UNDERWENT REOPERATION ON (B)(6) 2021 WHERE THE FRACTURED SCREW AND ROD WERE REPLACED AND AN EXTRA ROD WAS ADDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11642824 ↗
  • MalfunctionFeb 6, 2021Fracture; Loosening of Implant Not Related to Bone-Ingrowth; Fracture; Loosening of Implant Not Related to Bone-Ingrowth

    ON (B)(6) 2020, THE PATIENT UNDERWENT A SPINAL PROCEDURE WITH RELINE PRODUCTS AT L2/S2AI. ON (B)(6) 2020, IT WAS DISCOVERED DURING ROUTINE FOLLOW UP X-RAY THAT RODS AT L4/5 FRACTURED. ON (B)(6) 2020, A REVISION SURGERY WAS PERFORMED. IN THE REVISION SURGERY, SACRAL SCREWS HAD LOOSENED, SO THE SCREWS WERE REMOVED AND INSERTED ONE SIZE LARGER SCREWS INSTEAD, AND THE FRACTURED RODS WERE REPLACED TO NEW TI RODS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 11290242 ↗
  • MalfunctionMay 29, 2020Fracture; Fracture

    ON UNKNOWN DATE, AN X-RAY REVEALED THAT AN IMPLANTED RELINE ROD BROKE AT NEAR L5 LEVEL. AS PER REPORTER THE ROD WAS PLANNED TO BE REPLACED ON (B)(6) 2020.

    Patient problems: Failure of Implant

    MAUDE report 10099837 ↗
  • InjuryDec 19, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RECEIVED INFORMATION STATING PATIENT WAS EXPERIENCING SEVERE ALLERGIC REACTIONS POST OPERATIVE. INQUIRY RECEIVED, BREAKDOWN OF THE METALS IN EACH OF THE PIECES UTILIZED IN THE PATIENT'S PROCEDURE.

    Patient problems: Reaction; Seizures

    MAUDE report 9497776 ↗
  • MalfunctionAug 8, 2018Break; Break

    ON (B)(6) 2018, PATIENT UNDERWENT A SPINAL FUSION PROCEDURE AT T8 TO PELVIS LEVELS. A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2018 DUE TO BROKEN RODS. NO PATIENT INJURY HAS BEEN REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 7762908 ↗
  • MalfunctionDec 11, 2017Fracture; Fracture

    ON (B)(6) 2017 A PATIENT UNDERWENT A POSTERIOR SPINAL FUSION PROCEDURE AT L4-S1 LEVELS. LATE (B)(6), THE PATIENT EXPERIENCED PAIN, RADIOGRAPHS SHOWED A FRACTURED ROD. ON (B)(6) 2017 A REVISION PROCEDURE TOOK PLACE REPLACING THE ROD AND REINFORCING THE CONSTRUCT. NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 7106619 ↗
  • MalfunctionJun 6, 2017Fracture; Fracture

    ON (B)(6) 2016, A (B)(6) YEAR-OLD MALE PATIENT UNDERWENT POSTERIOR SPINAL FUSION AT THE T10-S2 LEVELS. THE ROD REPORTEDLY FRACTURED ON BOTH SIDES AT THE L4/5 LEVEL. ON (B)(6) 2017 REVISION SURGERY OCCURRED TO REPLACE THE FRACTURED ROD. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 6615991 ↗
  • MalfunctionDec 8, 2016Fracture; Fracture

    ON (B)(6) 2016 A PATIENT UNDERWENT A POSTERIOR FIXATION PROCEDURE. ON (B)(6) 2016 A REVISION PROCEDURE TOOK PLACE TO REPLACE A FRACTURED ROD AT L4/5. NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 6155797 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →