Skip to content

ReLine 11103055Recalls & reported problems

Nuvasive, Inc. · RELINE-O Conn, 5-6/5-6mm Rotating O-O · Class 2 · Implant

All models of ReLine (129)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887517596222), received between Jun 5, 2017 and Nov 25, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Separation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Separation Problem1
  • Unintended Movement1

Patient problems

  • Failure of Implant1
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 25, 2020Separation Problem; Separation Problem

    THE PATIENT UNDERWENT A SPINAL PROCEDURE WITH RELINE PRODUCTS ON UNKNOWN DATE. IN THE SURGERY, THE PHYSICIAN INSTALLED RODS ON THE L3/4 AND ATTACHED O-O ROTATING CONNECTORS ON THE RODS. AFTER THAT, THE PHYSICIAN INSTALLED ANOTHER RODS ON THE O-O ROTATING CONNECTORS AND JOINTED TO L4/SAI WITH ITS RODS. ON (B)(6) 2020, DUE TO A PSEUDARTHROSIS (NONUNION), THE PATIENT UNDERWENT A REVISION SURGERY IN ORDER TO RE-FIX FROM L3 TO SAI WITH SINGLE RODS. INTRAOPERATIVELY, WHEN THE PHYSICIAN TRIED TO REMOVE THE O-O ROTATING CONNECTORS OF L3/4, HE NOTICED THAT THE O-O ROTATING CONNECTOR ON THE LEFT SIDE FRACTURED. HE REMOVED THE BROKEN PARTS, HOWEVER, HE WAS NOT ABLE TO FIND ONE BROKEN PART. HE DECIDED TO COMPLETE THE OPERATION WITHOUT REMOVING IT.

    Patient problems: Nonunion/Delayed-union Bone Fracture

    MAUDE report 10900596 ↗
  • MalfunctionJun 5, 2017Unintended Movement; Unintended Movement

    ON (B)(6) 2017 A PATIENT UNDERWENT A MULTI-LEVEL INTERBODY REPLACEMENT PROCEDURE WITH ADDITIONAL POSTERIOR FIXATION FROM T11 TO S1 WITHOUT ISSUE. DURING A LATER FOLLOW UP VISIT RADIOGRAPHS CONFIRMED TWO LOCK SCREWS BACKING OUT ON THE RIGHT SIDE CONNECTORS. ON (B)(6) 2017 A REVISION TOOK PLACE REPLACING CONNECTORS AND LOCK SCREWS WITHOUT ISSUE. NO PATIENT INJURY HAS BEEN REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 6610960 ↗