Skip to content

Modulus 1775850Recalls & reported problems

Nuvasive, Inc. · Modulus XLIF Inserter · Class 2

All models of Modulus (41)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Nuvasive Modulus XLIF inserter is a surgical tool used to place a fusion cage in the spine. Most of the 37 reports describe the inserter's threaded tip snapping off during implant placement and staying stuck inside the cage, where it could not be retrieved. No deaths and one injury were reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
37
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 37 adverse event reports that list this exact device (DI 00887517913425), received between Jul 25, 2022 and Apr 21, 2026 — classified as 1 injury, 36 malfunctions.

The most frequently reported specific device problem is “Fracture” (37 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 3 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 4 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture37

Patient problems

  • Foreign Body In Patient33
  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 37 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 21, 2026Fracture

    THE TIP OF INSERTER BROKE OFF IN IMPLANT AND TIP WAS NOT ABLE TO BE RETRIEVED. NO ADVERSE EFFECT TO PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 24944020 ↗
  • MalfunctionApr 20, 2026Fracture

    IT WAS REPORTED THAT THE TIP OF THE MODULUS XLIF INSERTER WAS SEVERED. THE TIP WAS BROKEN IN THE CAGE WHILE BEING INSERTED AND IT WAS LEFT SECURELY IN THE CAGE AND COULD NOT BE REMOVED. THE PATIENT IS DOING WELL, NO ADVERSE EFFECT.

    Patient problems: Foreign Body In Patient

    MAUDE report 24932307 ↗
  • MalfunctionApr 16, 2026Fracture

    IT WAS REPORTED THAT THE TIP OF THE INSERTER BROKE OFF IN THE CAGE WHILE BEING INSERTED DURING A XLIF CASE. THE TIP WAS LEFT IN THE PATIENT, AND THE CAGE WAS FURTHER ¿TAMPED¿ IN TO PLACE THE CAGE. THE FRACTURE TIP IS FULLY INSERTED AND STUCK INTO THE CAGE.

    Patient problems: Foreign Body In Patient

    MAUDE report 24903943 ↗
  • MalfunctionFeb 2, 2026Fracture

    IT WAS REPORTED THAT WHILE INSERTING MODULUS IMPLANT, THE INNER THREADED PIECE OF THE INSERTER THAT INTERACTS WITH THE THREADED HOLE IN THE IMPLANT BROKE OFF AND STAYED IN IMPLANT. THEY COULD NOT GET THE IMPLANT OUT SINCE IT WAS MORE THAN HALFWAY IN AND SO WE JUST TAMPED IMPLANT FURTHER INTO DISC SPACE AND LEFT THE BROKEN PIECE IN THE IMPLANT. NOTHING WAS PUT OVER THE TIP; IT¿S THREADED ALL THE WAY INTO THE IMPLANT AND BROKE OFF. NO REVISION NECESSARY AND AS OF NOW NO REPORTED EFFECTS TO PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 24236466 ↗
  • MalfunctionSep 25, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF A MODULUS XLW BROKE OFF INSIDE THE CAGE AND COULD NOT BE RETRIEVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 23146547 ↗
  • MalfunctionSep 10, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF A MODULUS XLW INSERTER BROKE OFF INSIDE THE IMPLANTED CAGE AND WAS UNABLE TO BE RETRIEVED.

    Patient problems: Foreign Body In Patient

    MAUDE report 23016045 ↗
  • MalfunctionFeb 7, 2025Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A ANTERIOR LONGITUDINAL INTERBODY FUSION PROCEDURE FROM L2 TO S1. DURING THE PROCEDURE THE THREADED TIP OF THE INSERTER FRACTURED OFF DURING IMPLANT INSERTION AT THE L3-L4 LEVEL. THE FRACTURED PORTION WAS RETAINED IN THE IMPLANT SO THE SURGEON OPTED TO LEAVE THE PORTION IN SITU. NO ADDITIONAL PATIENT OR PROCEDURE IMPACT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21333251 ↗
  • MalfunctionDec 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING INTERBODY IMPLANT INSERTION THE DISTAL, THREADED TIP, OF THE DEVICE FRACTURED OFF INTO THE ENGAGED IMPLANT AND COULD NOT BE RETRIEVED. THE FRACTURED PORTION WAS RETAINED IN THE IMPLANT SO THE SURGEON OPTED TO LEAVE THE PORTION IN SITU. NO FURTHER PATIENT IMPACT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21039802 ↗
  • MalfunctionDec 13, 2024Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT UNDERWENT A POSTERIOR SPINAL PROCEDURE AND WHILE DURING INTERBODY IMPLANT INSERTION THE DISTAL, THREADED TIP, OF THE DEVICE FRACTURED OFF INTO THE ENGAGED IMPLANT AND COULD NOT BE RETRIEVED. THE FRACTURED PORTION WAS RETAINED IN THE IMPLANT SO THE SURGEON OPTED TO LEAVE THE PORTION IN SITU. NO FURTHER PATIENT OR CASE IMPACT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20930590 ↗
  • MalfunctionDec 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT WHILE INSERTING THE INTERBODY IMPLANT DURING AN EXTREME LATERAL INTERBODY FUSION CASE, THE DISTAL, THREADED, TIP OF THE DEVICE FRACTURED OFF INTO THE ENGAGED IMPLANT AND COULD NOT BE RETRIEVED. THE FRACTURED PORTION WAS RETAINED IN THE IMPLANT SO THE SURGEON OPTED TO LEAVE THE PORTION IN SITU. NO FURTHER PATIENT IMPACT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20857164 ↗
  • MalfunctionDec 4, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING INTERBODY IMPLANT INSERTION THE DISTAL, THREADED TIP, OF THE DEVICE FRACTURED OFF INTO THE ENGAGED IMPLANT AND COULD NOT BE RETRIEVED. THE FRACTURED PORTION WAS RETAINED IN THE IMPLANT SO THE SURGEON OPTED TO LEAVE THE PORTION IN SITU. NO FURTHER PATIENT IMPACT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20849419 ↗
  • MalfunctionSep 19, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING INTERBODY IMPLANT INSERTION THE DISTAL, THREADED TIP, OF THE DEVICE FRACTURED OFF INTO THE ENGAGED IMPLANT AND COULD NOT BE RETRIEVED. THE FRACTURED PORTION WAS RETAINED IN THE IMPLANT SO THE SURGEON OPTED TO LEAVE THE PORTION IN SITU. NO FURTHER PATIENT IMPACT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 20263727 ↗
  • MalfunctionSep 13, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING INTERBODY IMPLANT INSERTION THE DISTAL, THREADED TIP, OF THE DEVICE FRACTURED OFF INTO THE ENGAGED IMPLANT AND COULD NOT BE RETRIEVED. THE FRACTURED PORTION WAS RETAINED IN THE IMPLANT SO THE SURGEON OPTED TO LEAVE THE PORTION IN SITU. NO FURTHER PATIENT IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20216024 ↗
  • MalfunctionMay 22, 2024Fracture; Fracture

    ON 13 MAY 2024 A PATIENT UNDERWENT AN EXTREME LATERAL INTERBODY FUSION PROCEDURE FROM L3-L4. DURING INTERBODY IMPLANT INSERTION THE DISTAL, THREADED TIP, OF THE DEVICE FRACTURED OFF INTO THE ENGAGED IMPLANT AND COULD NOT BE RETRIEVED. THE FRACTURED PORTION WAS RETAINED IN THE IMPLANT SO THE SURGEON OPTED TO LEAVE THE PORTION IN SITU. NO FURTHER PATIENT IMPACT WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 19376626 ↗
  • MalfunctionMay 15, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A LUMBAR PROCEDURE THE DISTAL, THREADED, TIP OF A INTERBODY INSERTER FRACTURED OFF WITHIN THE INTERBODY AND WAS RETAINED IN THE IMPLANT. THE TIP WAS LEFT IN-SITU, EMBEDDED IN THE IMPLANT. NO FURTHER PATIENT IMPACT WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 19324135 ↗
  • MalfunctionFeb 29, 2024Fracture; Fracture

    ON (B)(6) 2024 DURING AN EXTREME LATERAL INTERBODY FUSION AT L3/5 DURING THE CAGE INSERTION AT L3/4 THE INSERTER THREADS SNAPPED. THE THREADS COULD NOT BE RETRIEVED AND WERE LEFT INSIDE THE CAGE IN-SITU. NO PATIENT INJURIES WERE REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 18812728 ↗
  • MalfunctionFeb 9, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL FOUR (4) LEVEL XLIF PROCEDURE. DURING INSERTION OF THE MODULUS IMPLANT AT THE 3RD LEVEL, L2/3, THE DISTAL THREADED TIP OF THE INSERTER FRACTURED OFF INSIDE THE MODULUS CAGE AND WAS UNABLE TO BE RETRIEVED. THE IMPLANT WAS SLIGHTLY ANTERIOR TO THE INTENDED PLACEMENT; HOWEVER, THE SURGEON WAS CONCERNED THAT REPOSITIONING THE IMPLANT MAY RISK PATIENT SAFETY. AS A RESULT, THE IMPLANT WAS LEFT IN THE ANTERIOR POSITION AND THE PROCEDURE WAS COMPLETED WITH NO FURTHER IMPACT TO THE PATIENT. THE SURGEON PLANS TO MONITOR THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 18681546 ↗
  • MalfunctionFeb 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT WHILE MALLETING THE INTERBODY INTO THE DISC SPACE VIA THE INTERBODY INSERTER, THE DISTAL ENGAGEMENT FEATURE OF THE INSERTER FRACTURED OFF INSIDE THE INTERBODY AND COULD NOT BE RECOVERED. THEREFORE, THE FRACTURED PIECE REMAINS IN THE PATIENT, LODGED WITHIN THE INTERBODY. THERE WAS NO PATIENT HARM OR ADVERSE CONSEQUENCE RELATED TO THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 18636785 ↗
  • MalfunctionJan 12, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE DISTAL THREADED TIP OF THE INSERTER FRACTURED OFF INSIDE THE INTERBODY WHILE MALLEATING THE INTERBODY FOR PLACEMENT AT L2/3. PER SURGEON DECISION, THE FRACTURED TIP WAS NOT RETRIEVED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE PATIENT, USER, OR PROCEDURE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 18508614 ↗
  • MalfunctionOct 31, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING INSERTION OF THE CAGE INTO THE DISC SPACE AT L3/4, THE DISTAL TIP OF THE INSERTER FRACTURED OFF INTO THE CAGE AND COULD NOT BE RECOVERED. THEREFORE, THE INSERTER TIP REMAINS LODGED INSIDE THE CAGE. THERE WAS NO PATIENT HARM OR ADVERSE CONSEQUENCE RELATED TO THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 18044405 ↗
  • MalfunctionSep 7, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING INSERTION OF THE CAGE INTO THE DISC SPACE AT L2/3, THE INSERTER TIP FRACTURED OFF INSIDE THE CAGE AND COULD NOT BE RETRIEVED. THEREFORE, THE INSERTER TIP REMAINS LODGED INSIDE THE IMPLANTED CAGE. THERE WAS NO PATIENT HARM OR ADVERSE CONSEQUENCE RELATED TO THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Foreign Body In Patient

    MAUDE report 17702471 ↗
  • MalfunctionSep 7, 2023Fracture; Fracture

    THREADS FOR INSERTER BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17700992 ↗
  • MalfunctionAug 15, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING INSERTION OF THE CAGE INTO THE DISC SPACE, THE DISTAL TIP OF THE INSERTER FRACTURED OFF INTO THE CAGE AND COULD NOT BE RECOVERED. THEREFORE, THE INSERTER TIP REMAINS LODGED INSIDE THE IMPLANTED CAGE. THERE WAS NO PATIENT HARM OR ADVERSE CONSEQUENCE RELATED TO THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17553523 ↗
  • MalfunctionAug 2, 2023Fracture; Fracture

    ON (B)(6) 2023 A PATIENT UNDERWENT A EXTREME LATERAL INTERBODY FUSION PROCEDURE ON UNKNOWN LEVEL OF THE SPINE. DURING THE PROCEDURE THE DISTAL TIP OF INSERTER FRACTURED OFF INTO THE CAGE AND WAS LEFT IN THE PATIENT AS IT WAS DETERMINED INTRAOPERATIVELY THAT IT COULD CAUSE MORE TRAUMA TO THE PATIENT TO EXPLANT THE CAGE OPPOSED TO LEAVING IT COLD WELDED IN THE ENGAGED SCREW. NO PATIENT INJURY OR CONSEQUENCE WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 17446007 ↗
  • MalfunctionJun 30, 2023Fracture; Fracture

    DURING A SPINAL PROCEDURE THE TIP WAS FRACTURED OFF INSIDE THE CAGE AND LEFT IN-SITU. NO PATIENT INJURY OR CONSEQUENCE WAS REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 17238906 ↗

Other Intervertebral fusion device with bone graft, thoracic devices

All Intervertebral fusion device with bone graft, thoracic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page