Skip to content

PRECICE® STRYDE System PS10.0-80B280Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Piriformis Straight · Class 2 · Implant

All models of PRECICE STRYDE System (231)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887517950772), received between Sep 26, 2019 and Mar 15, 2022 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Oct 29, 2021Open, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Open, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Defective Device1
  • Mechanical Problem1

Patient problems

  • Failure of Implant2

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 15, 2022Break; Break

    INFORMATION WAS RECEIVED THAT THE PATIENT HAD COMPLETED LENGTHENING AND CONSOLIDATION, WHEN IT WAS NOTED THAT AFTER EXPLANT, THE NAIL WAS BROKEN THROUGH THE DISTAL PROXIMAL SCREW. NO PATIENT ADVERSE EVENT WAS REPORTED.

    MAUDE report 13765644 ↗
  • MalfunctionNov 16, 2020Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE NAIL FAILED TO DISTRACT INTRA-OPERATIVELY DURING AN IMPLANTATION PROCEDURE ON (B)(6) 2020. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10849303 ↗
  • MalfunctionSep 26, 2019Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT ALLEGED THAT THE STRYDE NAIL'S DISTAL SCREW WAS BACKING OUT AND THE NAIL HAD SUB-OPTIMUM LENGTHENING. AS PER REPORTER, PATIENT CLAIMS TO BE NON-WEIGHT BEARING AND IS USING THE ERC PER SURGEON PRESCRIPTION.

    Patient problems: Failure of Implant

    MAUDE report 9123954 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page