PRECICE® STRYDE System PS13.0-80B280Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Piriformis Straight · Class 2 · Implant
All models of PRECICE STRYDE System (231)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887517950918), received between Mar 3, 2020 and Aug 31, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Defective Device” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Oct 29, 2021Class IIOpen, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 12, 2021Class IIOpen, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Defective Device1
- Mechanical Problem1
Patient problems
- Failure of Implant2
- Limb Pain2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 31, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED VIA THE PATIENT THAT THE THEY HAVE BEEN EXPERIENCING SEVERE PAIN AND ARE RESTRICTED DUE TO THE SEVERITY OF THE PAIN. THE PATIENT IS BARELY ABLE TO PICK UP AN OBJECT FROM FLOOR LEVEL OR ENDURE SITTING FOR LONG PERIODS OF TIME AND WALKING HURTS AS THE PAIN IS CONSTANT. ADDITIONALLY, THEY REPORTED THAT THEY HAVE OCCASIONALLY TAKEN PAIN MEDICATION WHICH EASED THE SHARPNESS BUT NOT ENOUGH NOR LONG ENOUGH. THE PATIENT HAS TO PUSH THEMSELVES UP USING THEIR ARMS WHEN STANDING UP AND HAS TROUBLE DRIVING. AT TIMES THEIR KNEE WILL TRY TO BUCKLE AND THEY HAVE A LOT OF SWELLING IN THE LOWER LEG. ONCE THE BONE HEALED THE PAIN REPORTEDLY GOT WORSE. THE NAIL REMAINS IMPLANTED AT THIS TIME. A REVISION IS PLANNED FOR AN UNKNOWN DATE. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Limb Pain
MAUDE report 17662446 ↗InjuryAug 31, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED VIA THE PATIENT THAT THE THEY HAVE BEEN EXPERIENCING SEVERE PAIN AND ARE RESTRICTED DUE TO THE SEVERITY OF THE PAIN. THE PATIENT IS BARELY ABLE TO PICK UP AN OBJECT FROM FLOOR LEVEL OR ENDURE SITTING FOR LONG PERIODS OF TIME AND WALKING HURTS AS THE PAIN IS CONSTANT. ADDITIONALLY, THEY REPORTED THAT THEY HAVE OCCASIONALLY TAKEN PAIN MEDICATION WHICH EASED THE SHARPNESS BUT NOT ENOUGH NOR LONG ENOUGH. THE PATIENT HAS TO PUSH THEMSELVES UP USING THEIR ARMS WHEN STANDING UP AND HAS TROUBLE DRIVING. AT TIMES THEIR KNEE WILL TRY TO BUCKLE AND THEY HAVE A LOT OF SWELLING IN THE LOWER LEG. ONCE THE BONE HEALED THE PAIN REPORTEDLY GOT WORSE. THE NAIL REMAINS IMPLANTED AT THIS TIME. A REVISION IS PLANNED FOR AN UNKNOWN DATE. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Limb Pain
MAUDE report 17662425 ↗InjuryNov 16, 2020Mechanical Problem; Mechanical Problem
NO ADDITIONAL EVENT INFORMATION WAS PROVIDED.
Patient problems: Failure of Implant
MAUDE report 10849229 ↗InjuryMar 3, 2020Defective Device; Defective Device
INFORMATION WAS RECEIVED THAT A REVISION SURGERY WILL BE PERFORMED ON (B)(6) 2020. AS PER REPORTER, X-RAY IMAGES REVEALED THAT THE NAIL ON THE LEFT SIDE WAS NOT LENGTHENING.
Patient problems: Failure of Implant
MAUDE report 9783286 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page