Skip to content

PRECICE® STRYDE System PS10.0-65D250Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Trochanter 10° Bend · Class 2 · Implant

All models of PRECICE STRYDE System (231)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00887517951175), received between Feb 23, 2021 and Aug 5, 2024 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Corroded” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 12, 2021Open, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Corroded8
  • Detachment of Device or Device Component2
  • Naturally Worn2
  • Degraded1
  • Material Discolored1
  • Material Erosion1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Implant Pain1
  • Pain1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND AILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING. THE PATIENT HAD PREVIOUSLY REPORTED EXPERIENCING SIGNIFICANT MID-THIGH PAIN, AND RADIOGRAPHIC OBSERVATIONS IDENTIFIED CORTICAL THICKENING AT THE EXTENDABLE JUNCTION OF THE NAIL.

    Patient problems: Implant Pain

    MAUDE report 19912219 ↗
  • MalfunctionJan 27, 2022Corroded; Detachment of Device or Device Component; Naturally Worn; Corroded; Detachment of Device or Device Component; Naturally Worn

    INFORMATION WAS RECEIVED THAT UPON EXPLANT OF THE NAIL AFTER IT HAD ACHIEVED INTENDED THERAPEUTIC LENGTHENING, IT WAS NOTED THE NAIL EXHIBITED CREVICE CORROSION AND WEAR BETWEEN THE NAIL AND FIXATION SCREWS. ADDITIONALLY, IT WAS FOUND THAT THE CROWN OF THE NAIL HAD BECOME DISPLACED AND APPEARED TO HAVE COLD-WELDED TO THE MALE COMPONENT INSTEAD OF STAYING WITHIN THE FEMALE COMPONENT. NO PATIENT HARM OR ADVERSE IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13367931 ↗
  • MalfunctionJan 27, 2022Corroded; Detachment of Device or Device Component; Naturally Worn; Corroded; Detachment of Device or Device Component; Naturally Worn

    INFORMATION WAS RECEIVED THAT UPON EXPLANT OF THE NAIL AFTER IT HAD ACHIEVED INTENDED THERAPEUTIC LENGTHENING, IT WAS NOTED THE NAIL EXHIBITED CREVICE CORROSION AND WEAR BETWEEN THE NAIL AND FIXATION SCREWS. ADDITIONALLY, IT WAS FOUND THAT THE CROWN OF THE NAIL HAD BECOME DISPLACED AND APPEARED TO HAVE COLD-WELDED TO THE MALE COMPONENT INSTEAD OF STAYING WITHIN THE FEMALE COMPONENT. NO PATIENT HARM OR ADVERSE IMPACT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13367897 ↗
  • MalfunctionNov 22, 2021Corroded; Corroded

    INFORMATION WAS RECEIVED THAT UPON EXPLANT THE NAIL WAS NOTED TO HAVE CREVICE CORROSION AT THE JUNCTION SITE OF THE MALE AND FEMALE COMPONENTS BETWEEN THE SCREWS AND SCREW HOLES OF THE NAIL. CORROSION WAS ALSO NOTED ON THE MALE COMPONENT AND DISTAL SCREW HOLES. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12852264 ↗
  • MalfunctionSep 15, 2021Corroded; Corroded

    NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12478165 ↗
  • MalfunctionAug 23, 2021Corroded; Corroded

    INFORMATION WAS RECEIVED THAT UPON EXPLANT CORROSION WAS SEEN ON THE NAIL. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12356736 ↗
  • MalfunctionAug 3, 2021Corroded; Corroded

    INFORMATION WAS RECEIVED THAT UPON EXPLANT THE NAIL WAS NOTICED TO HAVE DISCOLORATION AROUND THE TELESCOPING AREA. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12267879 ↗
  • MalfunctionJun 2, 2021Material Erosion; Material Erosion

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Pain

    MAUDE report 11922080 ↗
  • MalfunctionMay 6, 2021Corroded; Corroded

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11787559 ↗
  • InjuryFeb 23, 2021Degraded; Degraded

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 11369910 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →