PRECICE® STRYDE System PS10.0-65D250Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Trochanter 10° Bend · Class 2 · Implant
All models of PRECICE STRYDE System (231)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00887517951175), received between Feb 23, 2021 and Aug 5, 2024 — classified as 1 injury, 9 malfunctions.
The most frequently reported specific device problem is “Corroded” (8 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Feb 12, 2021Open, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Corroded8
- Detachment of Device or Device Component2
- Naturally Worn2
- Degraded1
- Material Discolored1
- Material Erosion1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Implant Pain1
- Pain1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND AILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING. THE PATIENT HAD PREVIOUSLY REPORTED EXPERIENCING SIGNIFICANT MID-THIGH PAIN, AND RADIOGRAPHIC OBSERVATIONS IDENTIFIED CORTICAL THICKENING AT THE EXTENDABLE JUNCTION OF THE NAIL.
Patient problems: Implant Pain
MAUDE report 19912219 ↗MalfunctionJan 27, 2022Corroded; Detachment of Device or Device Component; Naturally Worn; Corroded; Detachment of Device or Device Component; Naturally Worn
INFORMATION WAS RECEIVED THAT UPON EXPLANT OF THE NAIL AFTER IT HAD ACHIEVED INTENDED THERAPEUTIC LENGTHENING, IT WAS NOTED THE NAIL EXHIBITED CREVICE CORROSION AND WEAR BETWEEN THE NAIL AND FIXATION SCREWS. ADDITIONALLY, IT WAS FOUND THAT THE CROWN OF THE NAIL HAD BECOME DISPLACED AND APPEARED TO HAVE COLD-WELDED TO THE MALE COMPONENT INSTEAD OF STAYING WITHIN THE FEMALE COMPONENT. NO PATIENT HARM OR ADVERSE IMPACT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13367931 ↗MalfunctionJan 27, 2022Corroded; Detachment of Device or Device Component; Naturally Worn; Corroded; Detachment of Device or Device Component; Naturally Worn
INFORMATION WAS RECEIVED THAT UPON EXPLANT OF THE NAIL AFTER IT HAD ACHIEVED INTENDED THERAPEUTIC LENGTHENING, IT WAS NOTED THE NAIL EXHIBITED CREVICE CORROSION AND WEAR BETWEEN THE NAIL AND FIXATION SCREWS. ADDITIONALLY, IT WAS FOUND THAT THE CROWN OF THE NAIL HAD BECOME DISPLACED AND APPEARED TO HAVE COLD-WELDED TO THE MALE COMPONENT INSTEAD OF STAYING WITHIN THE FEMALE COMPONENT. NO PATIENT HARM OR ADVERSE IMPACT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13367897 ↗MalfunctionNov 22, 2021Corroded; Corroded
INFORMATION WAS RECEIVED THAT UPON EXPLANT THE NAIL WAS NOTED TO HAVE CREVICE CORROSION AT THE JUNCTION SITE OF THE MALE AND FEMALE COMPONENTS BETWEEN THE SCREWS AND SCREW HOLES OF THE NAIL. CORROSION WAS ALSO NOTED ON THE MALE COMPONENT AND DISTAL SCREW HOLES. NO PATIENT ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12852264 ↗MalfunctionSep 15, 2021Corroded; Corroded
NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12478165 ↗MalfunctionAug 23, 2021Corroded; Corroded
INFORMATION WAS RECEIVED THAT UPON EXPLANT CORROSION WAS SEEN ON THE NAIL. NO PATIENT ADVERSE EVENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12356736 ↗MalfunctionAug 3, 2021Corroded; Corroded
INFORMATION WAS RECEIVED THAT UPON EXPLANT THE NAIL WAS NOTICED TO HAVE DISCOLORATION AROUND THE TELESCOPING AREA. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12267879 ↗MalfunctionJun 2, 2021Material Erosion; Material Erosion
MalfunctionMay 6, 2021Corroded; Corroded
NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11787559 ↗InjuryFeb 23, 2021Degraded; Degraded
NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11369910 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page