Skip to content

PRECICE® STRYDE System PS10.0-80D265Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Trochanter 10° Bend · Class 2 · Implant

All models of PRECICE STRYDE System (231)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
12
Deaths reported
0
Injuries reported
7
Recalls naming it
2

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00887517951182), received between Nov 20, 2020 and Feb 18, 2025 — classified as 7 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2021Open, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Open, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break6
  • Adverse Event Without Identified Device or Use Problem4
  • Corroded4

Patient problems

  • Osteolysis6
  • Pain4
  • Failure of Implant3
  • No Clinical Signs, Symptoms or Conditions3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Inflammation2

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 18, 2025Break; Break

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WILL BE PERFORMED DUE TO A BROKEN NAIL BEING DISCOVERED VIA X-RAY.

    Patient problems: Failure of Implant

    MAUDE report 21403209 ↗
  • MalfunctionFeb 11, 2025Break; Break

    INFORMATION WAS RECEIVED THAT A REMOVAL PROCEDURE WAS PERFORMED DUE TO TWO BROKEN FEMORAL NAILS. THE NAILS WERE DIFFICULT TO REMOVE FROM THE BONE, RESULTING IN REMOVAL OF ONLY THE SCREWS AND PROXIMAL PORTION OF THE NAILS. THE UNIVERSAL REMOVAL KIT WAS USED REMOVAL OF ALL PORTIONS OF THE NAIL WAS UNSUCCESSFUL. DUE TO THIS THE PATIENT'S PROCEDURE WAS PROLONGED.

    Patient problems: Failure of Implant

    MAUDE report 21355203 ↗
  • MalfunctionApr 13, 2022Break; Break

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE NAIL BREAKING. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14104414 ↗
  • MalfunctionNov 19, 2021Corroded; Corroded

    NO ADDITIONAL INFORMATION WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12848916 ↗
  • InjurySep 8, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED VIA REVIEW OF MEDICAL LITERATURE THAT A PATIENT HAD OSTEOLYSIS THE NAIL¿S TELESCOPIC JUNCTION AND REPORTED PAIN DURING DISTRACTION AND CONSOLIDATION. COMPLETE RESOLUTION OF THE OSSEOUS ALTERATIONS WAS ACHIEVED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Osteolysis; Pain

    MAUDE report 12441199 ↗
  • InjurySep 8, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    N/A.

    Patient problems: Osteolysis

    MAUDE report 12440928 ↗
  • MalfunctionMar 30, 2021Corroded; Corroded

    DURING EXPLANT OF THE NAIL AFTER ACHIEVING ITS INTENDED RESULT, IT WAS NOTED THE NAIL WAS CORRODED. NO ADVERSE EVENT OR PATIENT IMPACT WAS REPORTED. IT WAS SUBSEQUENTLY REPORTED THAT AFTER THE NAIL WENT THROUGH THE STERILE CLEANING PROCESS AT THE REPORTING FACILITY, THE REPORTED CORROSION WAS NO LONGER VISIBLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11593080 ↗
  • InjuryMar 29, 2021Break; Corroded; Break; Corroded

    ADDITIONAL INFORMATION WAS RECEIVED VIA REVIEW OF LITERATURE THAT THE PATIENT ALSO EXPERIENCED SWELLING, OSTEOLYSIS, PERIOSTEAL REACTION, AND BONY HYPERTROPHY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Inflammation; Osteolysis; Pain

    MAUDE report 11586720 ↗
  • InjuryMar 29, 2021Break; Corroded; Break; Corroded

    INFORMATION WAS RECEIVED THAT THE PATIENT EXPERIENCED A SUDDEN ONSET OF PAIN AND UPON REMOVAL, THE NAIL WAS FOUND BROKEN AND THERE WERE SIGNS OF CORROSION. RADIOLOGICAL CHANGES WERE ALSO REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Inflammation; Osteolysis; Pain

    MAUDE report 11586717 ↗
  • InjuryDec 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    N/A.

    Patient problems: Osteolysis; Pain

    MAUDE report 11092364 ↗
  • InjuryDec 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE OSTEOLYSIS WAS LYTIC ONLY.

    Patient problems: Osteolysis

    MAUDE report 11092351 ↗
  • MalfunctionNov 20, 2020Break; Break

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON AN UNKNOWN DATE. PER THE REPORTER, THE NAIL WAS BROKEN. THE SURGEON BELIEVES IT MAY HAVE BEEN DUE TO FATIGUE AND SLOW BONE HEALING. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10880737 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →