PRECICE® STRYDE System PS10.0-80D280Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Trochanter 10° Bend · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887517951199), received between Mar 30, 2020 and Mar 29, 2021 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Corroded” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Oct 29, 2021Open, Classified
The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.
Z-0319-2022 · Root cause: Device Design · FDA record ↗
- Feb 12, 2021Open, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1323-2021 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Corroded1
- Defective Device1
- Leak/Splash1
- Mechanical Problem1
Patient problems
- Failure of Implant2
- Inflammation2
- Osteolysis2
- Pain2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 29, 2021Corroded; Corroded
INFORMATION WAS RECEIVED THAT A REVISION WAS PERFORMED IN (B)(6) 2021 DUE TO X-RAY IMAGES REVEALING THAT THE PATIENT'S BONE WAS BREAKING DOWN. UPON EXPLANT, IT WAS NOTED THAT THE NAIL HAD CORROSION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Inflammation; Osteolysis; Pain
MAUDE report 11586621 ↗InjuryFeb 23, 2021Leak/Splash; Leak/Splash
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2021. AS PER THE REPORTER, THE PATIENT COMPLAINED OF INCREASING PAIN IN THE FEMUR. X-RAYS CONFIRMED INFLAMMATION AND OSTEOLYSIS WHICH IS BELIEVED TO BE ATTRIBUTED TO LEAKAGE/SPILLAGE FROM THE NAIL.
Patient problems: Inflammation; Osteolysis; Pain
MAUDE report 11369913 ↗InjurySep 28, 2020Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A REVISION PROCEDURE WAS COMPLETED ON (B)(6) 2020 AS THE NAIL WAS NOT LENGTHENING POST-OPERATIVELY. NO PATIENT INJURY WAS REPORTED.
Patient problems: Failure of Implant
MAUDE report 10597699 ↗MalfunctionMar 30, 2020Defective Device; Defective Device
INFORMATION WAS RECEIVED THAT AN OSTEOTOMY PROCEDURE WAS PERFORMED ON (B)(6) 2020. AS PER THE REPORTER, THE NAIL WAS NOT LENGTHENING IN THE LEFT LEG.
Patient problems: Failure of Implant
MAUDE report 9901015 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page