Skip to content

PRECICE® STRYDE System PS10.0-80D335Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Trochanter 10° Bend · Class 2 · Implant

All models of PRECICE STRYDE System (231)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00887517951212), received between Jan 6, 2020 and Aug 31, 2023 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Oct 29, 2021Open, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Open, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break3
  • Defective Device2

Patient problems

  • Failure of Implant4
  • Bone Fracture(s)1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 31, 2023Break; Break

    INFORMATION WAS RECEIVED THAT UPON EXTRACTION OF THE NAIL, THE PROXIMAL END OF THE NAIL BROKE AWAY AT THE SCREW HOLE LEAVING THE DISTAL PORTION OF THE NAIL IN THE PATIENT'S FEMUR. THE SURGEON HAD TO ENTER THE DISTAL SIDE OF THE PATIENT FEMUR, OPENING THE PATIENT UP ABOVE THE KNEE TO PUSH THE DISTAL PORTION OF THE NAIL THROUGH TO COMPLETE THE EXTRACTION PROCESS WHICH RESULTED IN A 5 HOUR SURGERY. IT WAS ADDITIONALLY REPORTED THAT THE PATIENT HAS SUFFERED A FRACTURE OF THE FEMUR DUE TO THE INTERVENTION REQUIRED TO REMOVE THE NAIL FOLLOWING THE BREAKAGE. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Bone Fracture(s)

    MAUDE report 17662512 ↗
  • MalfunctionJul 27, 2021Break; Break

    INFORMATION WAS RECEIVED THAT DURING A PROCEDURE TO REMOVE HARDWARE FROM THE PATIENT'S FEMUR, THE TIP OF THE NAIL BROKE.

    Patient problems: Failure of Implant

    MAUDE report 12233174 ↗
  • MalfunctionNov 25, 2020Break; Break

    SEE CORRECTED INFORMATION IN H10.

    Patient problems: Failure of Implant

    MAUDE report 10902044 ↗
  • MalfunctionNov 20, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A NAIL WAS BENT AFTER EXPLANTATION DURING A REMOVAL PROCEDURE ON (B)(6) 2020. NO PATIENT INJURY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10880684 ↗
  • InjuryJan 6, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT ON (B)(6) 2019 A REVISION PROCEDURE WAS PERFORMED. AS PER THE REPORTER, THE NAIL DID NOT DISTRACT DURING THE LENGTHENING.

    Patient problems: Failure of Implant

    MAUDE report 9555412 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page