Skip to content

PRECICE® STRYDE System PS13.0-80D265Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Trochanter 10° Bend · Class 2 · Implant

All models of PRECICE STRYDE System (231)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887517951328), received between Aug 5, 2024 and Aug 5, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Corroded” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2021Open, Classified

    The Instructions for Use have been updated to include: (1) Clarification that the device is intended for use only in patients 18 years and older, (2) That patients should weigh 50 lbs. or more while undergoing treatment, and (3) That no more than two devices should be implanted at a time.

    Z-0319-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Feb 12, 2021Open, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Corroded1
  • Material Discolored1
  • Physical Resistance/Sticking1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND FAILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING.

    Patient problems: Insufficient Information

    MAUDE report 19911908 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page