Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00887517951595), received between Oct 5, 2020 and Aug 5, 2024 — classified as 5 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Corroded” (8 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 12, 2021Class IIOpen, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Corroded8
- Material Discolored5
- Physical Resistance/Sticking5
- Adverse Event Without Identified Device or Use Problem2
- Material Erosion2
- Appropriate Device Problem Term/Code Not Available1
- Break1
Patient problems
- Pain4
- Metal Related Pathology3
- Implant Pain2
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Osteolysis2
- Bacterial Infection1
- Erythema1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND FAILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT DIDN'T INDICATE THE REASON FOR REMOVAL. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 19912135 ↗MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND FAILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19912097 ↗MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND FAILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT DIDN'T INDICATE THE REASON FOR REMOVAL. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 19912044 ↗MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS ABLE TO BE DISTRACTED AND RETRACTED AND PASSED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING. THE PATIENT HAD PREVIOUSLY REPORTED EXPERIENCING MID-THIGH PAIN.
Patient problems: Implant Pain
MAUDE report 19911892 ↗MalfunctionJul 26, 2024Corroded; Material Discolored; Material Erosion; Physical Resistance/Sticking; Corroded; Material Discolored; Material Erosion; Physical Resistance/Sticking
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT COULD BE DISTRACTED AND RETRACTED AND PASSED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE MATERIAL WEAR, CORROSION, SCRATCHING, DISCOLORATION, AND GALLING. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING. THE PATIENT HAD PREVIOUSLY REPORTED EXPERIENCING SEVERE PAIN, AND RADIOGRAPHIC OBSERVATIONS IDENTIFIED CORTICAL THICKENING AT THE EXTENDABLE JUNCTION OF THE LEFT NAIL. FURTHERMORE, IT WAS DETERMINED THAT THERE WAS PERIPROSTHETIC FIBROUS TISSUE WITH FOCAL METALLIC DEBRIS. MICROBIOLOGY FINDINGS TAKEN ON (B)(6) 2020 REVEALED THAT THE TISSUE HAD STAPH EPIDERMIDIS AND PAENICLOSTRIDIUM SORDELLII. REPORT 1 OF 2.
Patient problems: Bacterial Infection; Implant Pain; Metal Related Pathology; Skin Infection
MAUDE report 19838645 ↗InjuryMay 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED THAT ON AN UNKNOWN DATE A REVISION PROCEDURE WAS PERFORMED FOR AN UNKNOWN REASON. ADDITIONALLY IT WAS REPORTED THAT THE PATIENT HAD HIGH LEVELS OF CERTAIN METALS IN THEIR BLOOD. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Metal Related Pathology
MAUDE report 14522676 ↗InjuryMay 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED THAT ON AN UNKNOWN DATE A REVISION PROCEDURE WAS PERFORMED FOR AN UNKNOWN REASON. ADDITIONALLY IT WAS REPORTED THAT THE PATIENT HAD HIGH LEVELS OF CERTAIN METALS IN THEIR BLOOD. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Metal Related Pathology
MAUDE report 14522549 ↗InjurySep 2, 2021Corroded; Corroded
INFORMATION WAS RECEIVED THAT THE PATIENT HAD LYTIC-HYPERTROPHIC OSTEOLYSIS AND REPORTED PAIN DURING CONSOLIDATION AND DISTRACTION. ADDITIONALLY, IT WAS REPORTED THAT THE IMPLANT HAD CORROSION.
Patient problems: Osteolysis; Pain
MAUDE report 12417284 ↗MalfunctionJul 13, 2021Break; Corroded; Break; Corroded
INFORMATION WAS RECEIVED THAT THE NAIL WAS BROKEN AND THE PATIENT COMPLAINED OF PAIN. UPON EXPLANT CORROSION WAS SEEN ON THE NAIL.
Patient problems: Pain
MAUDE report 12161386 ↗MalfunctionJun 2, 2021Material Erosion; Material Erosion
NO ADDITIONAL INFORMATION WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11925166 ↗InjuryMar 29, 2021Corroded; Corroded
ADDITIONAL INFORMATION WAS RECEIVED VIA REVIEW OF LITERATURE THAT THE PATIENT EXPERIENCED SEVERE PAIN ON THE RIGHT LOWER LEG.
Patient problems: Pain
MAUDE report 11586723 ↗InjuryOct 5, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020 DUE TO LOCALIZED PAIN IN THE DISTAL TIBIA. X-RAYS WERE TAKEN WHICH SHOWED OSTEOLYSIS AND OSTEOMYELITIS WHERE THE NAIL WAS TELESCOPED.
Patient problems: Erythema; Osteolysis; Pain
MAUDE report 10632295 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page