Skip to content

PRECICE® STRYDE System PS10.0-50SJ235Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Antegrade Tibia 10° Bend · Class 2 · Implant

All models of PRECICE STRYDE System (231)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
12
Deaths reported
0
Injuries reported
5
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00887517951595), received between Oct 5, 2020 and Aug 5, 2024 — classified as 5 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Corroded” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 4 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 12, 2021Class IIOpen, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Corroded8
  • Material Discolored5
  • Physical Resistance/Sticking5
  • Adverse Event Without Identified Device or Use Problem2
  • Material Erosion2
  • Appropriate Device Problem Term/Code Not Available1
  • Break1

Patient problems

  • Pain4
  • Metal Related Pathology3
  • Implant Pain2
  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • Osteolysis2
  • Bacterial Infection1
  • Erythema1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND FAILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT DIDN'T INDICATE THE REASON FOR REMOVAL. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19912135 ↗
  • MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND FAILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19912097 ↗
  • MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS UNABLE TO BE DISTRACTED AND RETRACTED AND FAILED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT DIDN'T INDICATE THE REASON FOR REMOVAL. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 19912044 ↗
  • MalfunctionAug 5, 2024Corroded; Material Discolored; Physical Resistance/Sticking; Corroded; Material Discolored; Physical Resistance/Sticking

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT WAS ABLE TO BE DISTRACTED AND RETRACTED AND PASSED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE CORROSION, SCRATCHING, GALLING, SURFACE DEPOSITS AND DISCOLORATION. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING. THE PATIENT HAD PREVIOUSLY REPORTED EXPERIENCING MID-THIGH PAIN.

    Patient problems: Implant Pain

    MAUDE report 19911892 ↗
  • MalfunctionJul 26, 2024Corroded; Material Discolored; Material Erosion; Physical Resistance/Sticking; Corroded; Material Discolored; Material Erosion; Physical Resistance/Sticking

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED, DETAILING THE EXAMINATION OF THE RETRIEVED IMPLANT. DURING THE ASSESSMENT, IT WAS NOTED THAT THE IMPLANT COULD BE DISTRACTED AND RETRACTED AND PASSED THE FORCE TESTING. ADDITIONALLY, SURFACE DAMAGE GRADING REVEALED NOTABLE MATERIAL WEAR, CORROSION, SCRATCHING, DISCOLORATION, AND GALLING. THE REPORT INDICATED THAT THE ROD WAS REMOVED AS PART OF A PLANNED PROCEDURE FOLLOWING THE ACHIEVEMENT OF THE DESIRED LENGTHENING. THE PATIENT HAD PREVIOUSLY REPORTED EXPERIENCING SEVERE PAIN, AND RADIOGRAPHIC OBSERVATIONS IDENTIFIED CORTICAL THICKENING AT THE EXTENDABLE JUNCTION OF THE LEFT NAIL. FURTHERMORE, IT WAS DETERMINED THAT THERE WAS PERIPROSTHETIC FIBROUS TISSUE WITH FOCAL METALLIC DEBRIS. MICROBIOLOGY FINDINGS TAKEN ON (B)(6) 2020 REVEALED THAT THE TISSUE HAD STAPH EPIDERMIDIS AND PAENICLOSTRIDIUM SORDELLII. REPORT 1 OF 2.

    Patient problems: Bacterial Infection; Implant Pain; Metal Related Pathology; Skin Infection

    MAUDE report 19838645 ↗
  • InjuryMay 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED THAT ON AN UNKNOWN DATE A REVISION PROCEDURE WAS PERFORMED FOR AN UNKNOWN REASON. ADDITIONALLY IT WAS REPORTED THAT THE PATIENT HAD HIGH LEVELS OF CERTAIN METALS IN THEIR BLOOD. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Metal Related Pathology

    MAUDE report 14522676 ↗
  • InjuryMay 27, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED THAT ON AN UNKNOWN DATE A REVISION PROCEDURE WAS PERFORMED FOR AN UNKNOWN REASON. ADDITIONALLY IT WAS REPORTED THAT THE PATIENT HAD HIGH LEVELS OF CERTAIN METALS IN THEIR BLOOD. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Metal Related Pathology

    MAUDE report 14522549 ↗
  • InjurySep 2, 2021Corroded; Corroded

    INFORMATION WAS RECEIVED THAT THE PATIENT HAD LYTIC-HYPERTROPHIC OSTEOLYSIS AND REPORTED PAIN DURING CONSOLIDATION AND DISTRACTION. ADDITIONALLY, IT WAS REPORTED THAT THE IMPLANT HAD CORROSION.

    Patient problems: Osteolysis; Pain

    MAUDE report 12417284 ↗
  • MalfunctionJul 13, 2021Break; Corroded; Break; Corroded

    INFORMATION WAS RECEIVED THAT THE NAIL WAS BROKEN AND THE PATIENT COMPLAINED OF PAIN. UPON EXPLANT CORROSION WAS SEEN ON THE NAIL.

    Patient problems: Pain

    MAUDE report 12161386 ↗
  • MalfunctionJun 2, 2021Material Erosion; Material Erosion

    NO ADDITIONAL INFORMATION WAS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11925166 ↗
  • InjuryMar 29, 2021Corroded; Corroded

    ADDITIONAL INFORMATION WAS RECEIVED VIA REVIEW OF LITERATURE THAT THE PATIENT EXPERIENCED SEVERE PAIN ON THE RIGHT LOWER LEG.

    Patient problems: Pain

    MAUDE report 11586723 ↗
  • InjuryOct 5, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020 DUE TO LOCALIZED PAIN IN THE DISTAL TIBIA. X-RAYS WERE TAKEN WHICH SHOWED OSTEOLYSIS AND OSTEOMYELITIS WHERE THE NAIL WAS TELESCOPED.

    Patient problems: Erythema; Osteolysis; Pain

    MAUDE report 10632295 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page