Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887517979643), received between Nov 2, 2020 and Nov 2, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Migration” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 12, 2021Class IIOpen, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1324-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Feb 12, 2021Class IIOpen, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Migration1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 2, 2020Migration; Migration

    INFORMATION WAS RECEIVED THAT UPON IMPACTS TO THE INSTRUMENT DURING IMPLANTING PROCEDURE, THE DISTAL CAP CAME LOOSE. THE IMPLANT WAS REMOVED SUCCESSFULLY.

    Patient problems: Failure of Implant

    MAUDE report 10775848 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →