PRECICE® Bone Transport System BT10-80D320-7Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Trochanter 10° Bend, 10.0mm x 320mm, 7/8cm ST · Class 2 · Implant
All models of PRECICE Bone Transport System (561)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887517980205), received between Dec 22, 2025 and Dec 22, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Failure to Advance” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 12, 2021Class IIOpen, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1324-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Feb 12, 2021Class IIOpen, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Advance1
Patient problems
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2025Failure to Advance; Failure to Advance
ON (B)(6) 2025, A PATIENT UNDERWENT ANKLE ARTHRODESIS, AND A BONE TRANSPORT FEMUR NAIL, PREVIOUSLY PLANNED BY THE SURGEON, WAS IMPLANTED. DURING POSTOPERATIVE FOLLOW-UP, IT WAS OBSERVED THAT THE IMPLANT WAS NOT FUNCTIONING. A NEW SURGERY WAS SCHEDULED FOR (B)(6) 2025. THE IMPLANT WAS CHECKED BOTH INSIDE AND OUTSIDE THE PATIENT USING A FAST DISTRACTOR AND WITH ERC (ELECTROMAGNETIC RESONANCE IMAGING), AND THE MECHANISM WAS FOUND TO BE NON-FUNCTIONAL. THEREFORE, IT WAS DECIDED TO REPLACE IT WITH A NEW ONE, WHICH ULTIMATELY DID FUNCTION. THE DELAY IN REPORTING THE INCIDENT WAS DUE TO THE HOSPITAL FIRST DECONTAMINATING THE NAIL, AND THEN SEVERAL INTERNAL COMMUNICATION ERRORS (AN INTERNAL NON-CONFORMITY REPORT HAS BEEN FILED REGARDING THIS ISSUE, AND TRAINING WILL BE PROVIDED).
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 23865365 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page