Skip to content

PRECICE® Bone Transport System BT10-60X280-6Recalls & reported problems

Nuvasive Specialized Orthopedics, Inc. · Retrograde Femur Straight, 10.0mm x 280mm, 6/6cm ST · Class 2 · Implant

All models of PRECICE Bone Transport System (561)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Feb 12, 2021Class IIOpen, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Feb 12, 2021Class IIOpen, Classified

    Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.

    Z-1324-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page