PRECICE® Bone Transport System BT10-10D380-7Recalls & reported problems
Nuvasive Specialized Orthopedics, Inc. · Antegrade Femur Trochanter 10° Bend, 10.0mm x 380mm, 7/10cm ST · Class 2 · Implant
All models of PRECICE Bone Transport System (561)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887517981288), received between Jun 20, 2023 and Jan 23, 2026 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 12, 2021Class IIOpen, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1324-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Feb 12, 2021Class IIOpen, Classified
Due to complaints of adverse events potentially related to biological safety. The complaints included report of pain and bony abnormalities at the interface between the telescoping nail segments.
Z-1323-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Failure to Advance1
- Fracture1
Patient problems
- Failure of Implant2
- Implant Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 23, 2026Failure to Advance
IT WAS REPORTED THAT AFTER THE INITIAL OR DATE ON (B)(6) 2025 WHERE THE ANTEGRADE IMPLANTATION WAS PERFORMED WITH RETROGRADE TRANSPORT BY USING A CABLE WIRE. THE SLOT WAS MOVED DISTALLY USING FDM. A SURGICAL TOWEL WAS PLACED BETWEEN NAIL AND FDM. THE MANEUVER FUNCTIONED WITHOUT DIFFICULTY. THE IMPLANTATION OF THE NAIL PROCEEDED AS EXPECTED, MARKING THE MAGNET. THEN ON (B)(6) 2025 THERE WAS THE PATIENT ERC4 INTRODUCTION. WE RENEWED CONTACT ON (B)(6) 2025 WHERE THE SURGEON NOTICED THE INCOMPLETE OSTEOTOMY. ON (B)(6) 2025 WE HAD THE REVISION PROCEDURE. THIS INCLUDED THE COMPLETION OF THE OSTEOTOMY. AND FOLLOWING COMPLETION OF THE OSTEOTOMY, THE FAST DISTRACTOR MAX WAS APPLIED IN AN ATTEMPT TO MOVE THE SLOT. THE DEVICE WAS MAINTAINED AT THE MARKED POSITION FOR A MINIMUM OF 5 MINUTES UNDER XRAY CONTROL. NO MOVEMENT OF THE SLOT WAS OBSERVED. SUBSEQUENTLY, THE ERC 4 WAS APPLIED IN AN ADDITIONAL ATTEMPT TO ACTIVE THE SLOT MOVEMENT. AGAIN, NO MOVEMENT WAS DETECTED. THE CABLE WIRE WAS REMOVED. BASED ON INTRAOPERATIVE FINDINGS, NAIL EXCHANGE WAS DEEMED NOT SURGICALLY FEASIBLE. THE BONE TRANSPORT WILL NOW BE PERFORMED USING AN EXTERNAL RING FIXATOR. ADDITIONALLY, REGARDING REVISION SURGERY WE
Patient problems: Failure of Implant
MAUDE report 24167402 ↗InjuryNov 6, 2024Fracture; Fracture
THE INFORMATION RECEIVED INDICATES THAT A REVISION PROCEDURE WAS PERFORMED FOLLOWING THE PATIENT¿S REPORT OF IMPLANT-RELATED PAIN. RADIOGRAPHIC IMAGING CONFIRMED THAT THE IMPLANT HAD FRACTURED. THE PATIENT REPORTED THAT NO TRAUMATIC EVENT PRECEDED THE FRACTURE. THE PATIENT'S MEDICAL HISTORY INCLUDES TREATMENT FOR A LEFT FEMUR FRACTURE UTILIZING AN EXTERNAL FIXATOR IN (B)(6) 2024, FOLLOWED BY MANAGEMENT OF BONE LOSS THROUGH THE TWO-STAGE MASQUELET TECHNIQUE.
Patient problems: Implant Pain
MAUDE report 20623364 ↗MalfunctionJun 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17164693 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page